| Literature DB >> 2970441 |
P K Paulsen1, H Nygaard, J M Hasenkam, J Gormsen, H Stødkilde-Jørgensen, O Albrechtsen.
Abstract
To analyze velocity spectral energy distribution in humans, blood velocities were recorded by means of hot-film anemometry at 41 predetermined measurement points in the cross-sectional area of the ascending aorta approximately 6 cm downstream of the aortic valves. Measurements were made in 8 patients with normal aortic valves, in 4 after insertion of a St. Jude Medical (SJM) aortic valve and in 3 after insertion of a Starr-Edwards Silastic Ball (SSB) aortic valve. Data analysis based on Fast Fourier Transform demonstrated that turbulence energy was lower in patients with normal aortic valves than in patients after insertion of an artificial valve in the aortic position and probably more pronounced after SSB valves than after SJM valves. The spatial distribution of the turbulence energy above 100 Hz was more irregular than corresponding laminar velocities previously presented. The VER100 (Velocity Energy Ratio at 100 Hz, i.e. the velocity energy above 100 Hz divided by the total velocity energy) proved useful for evaluating differences in flow disturbances downstream of different aortic valves. The mean VER100 in the three categories of patients were respectively 0.3, 1.4, and 2.1%.Entities:
Mesh:
Year: 1988 PMID: 2970441
Source DB: PubMed Journal: Int J Artif Organs ISSN: 0391-3988 Impact factor: 1.595