Literature DB >> 29703274

Romer Labs RapidChek®Listeria monocytogenes Test System for the Detection of L. monocytogenes on Selected Foods and Environmental Surfaces.

Gregory Juck1, Verapaz Gonzalez1, Ann-Christine Olsson Allen1, Meredith Sutzko1, Kody Seward1, Mark T Muldoon1.   

Abstract

The Romer Labs RapidChek® Listeria monocytogenes test system (Performance Tested Method 011805) was validated against the U.S. Department of Agriculture-Food Safety and Inspection Service Microbiology Laboratory Guidebook (USDA-FSIS/MLG), U.S. Food and Drug Association Bacteriological Analytical Manual (FDA/BAM), and AOAC Official Methods of Analysis (AOAC/OMA) cultural reference methods for the detection of L. monocytogenes on selected foods including hot dogs, frozen cooked breaded chicken, frozen cooked shrimp, cured ham, and ice cream, and environmental surfaces including stainless steel and plastic in an unpaired study design. The RapidChek method uses a proprietary enrichment media system, a 44-48 h enrichment at 30 ± 1°C, and detects L. monocytogenes on an immunochromatographic lateral flow device within 10 min. Different L. monocytogenes strains were used to spike each of the matrixes. Samples were confirmed based on the reference method confirmations and an alternate confirmation method. A total of 140 low-level spiked samples were tested by the RapidChek method after enrichment for 44-48 h in parallel with the cultural reference method. There were 88 RapidChek presumptive positives. One of the presumptive positives was not confirmed culturally. Additionally, one of the culturally confirmed samples did not exhibit a presumptive positive. No difference between the alternate confirmation method and reference confirmation method was observed. The respective cultural reference methods (USDA-FSIS/MLG, FDA/BAM, and AOAC/OMA) produced a total of 63 confirmed positive results. Nonspiked samples from all foods were reported as negative for L. monocytogenes by all methods. Probability of detection analysis demonstrated no significant differences in the number of positive samples detected by the RapidChek method and the respective cultural reference method.

Entities:  

Mesh:

Year:  2018        PMID: 29703274     DOI: 10.5740/jaoacint.18-0035

Source DB:  PubMed          Journal:  J AOAC Int        ISSN: 1060-3271            Impact factor:   1.913


  1 in total

1.  Comparison of three neutralizing broths for environmental sampling of low levels of Listeria monocytogenes desiccated on stainless steel surfaces and exposed to quaternary ammonium compounds.

Authors:  Fengmin Li; Zhihan Xian; Hee Jin Kwon; Jiyoon Yoo; Laurel Burall; Stuart J Chirtel; Thomas S Hammack; Yi Chen
Journal:  BMC Microbiol       Date:  2020-11-02       Impact factor: 3.605

  1 in total