Mixing studies during hepatic artery infusion in an in vitro model.

R J Lutz1, D L Miller.   

Abstract

A glass model of the hepatic artery network was used to study the effect of infusion rate on the degree of mixing from an end hole catheter placed in the gastroduodenal artery. Red dye solutions were infused at rates ranging from 1 ml/hour to 20 ml/minute. Effluent samples from each of 16 branch arteries were collected and dye concentrations were analyzed by means of a spectrophotometer. Low infusion rates, e.g., up to 5 ml/minute, showed streaming of the dye solutions and a nonhomogeneous dye distribution in the distal branches. At 20 ml/minute, dye distribution was much more uniform. These experiments are designed to simulate intrahepatic infusion of chemotherapeutic drug solutions. Theoretical considerations suggesting a pharmacokinetic advantage of intraarterial delivery implicitly assume uniform distribution of drug solutions to all perfused tissue. The in vitro data in this study suggests that this assumption may not be operative under certain infusion conditions. Slow infusion can lead to streaming and nonuniform distribution of infused drug solutions, which may in part explain the variability in tumor response in different tissue regions and also some observed toxicities, such as bile duct stricturing and fibrosis after intrahepatic infusions.

Entities:  

Mesh:

Substances:

Year:  1988        PMID: 2970291     DOI: 10.1002/1097-0142(19880915)62:6<1066::aid-cncr2820620608>3.0.co;2-f

Source DB:  PubMed          Journal:  Cancer        ISSN: 0008-543X            Impact factor:   6.860


  4 in total

1.  Study of catheter designs and drug mixing processes using 2D steady numerical simulations.

Authors:  R Mongrain; K Kandarpa; A Garon; O F Bertrand; M Bertrand
Journal:  Med Biol Eng Comput       Date:  1999-01       Impact factor: 2.602

2.  Mixing during intravertebral arterial infusions in an in vitro model.

Authors:  Robert J Lutz; Kathy Warren; Frank Balis; Nicholas Patronas; Robert L Dedrick
Journal:  J Neurooncol       Date:  2002-06       Impact factor: 4.130

3.  Low infusion rate digital subtraction angiography to predict regional perfusion in hepatic arterial chemotherapy.

Authors:  J C Andrews; D M Williams; B Shapiro; W D Ensminger
Journal:  Cardiovasc Intervent Radiol       Date:  1989 Sep-Oct       Impact factor: 2.740

4.  Intra-arterial infusion of N-isopropyl-p[123I]iodoamphetamine for assessing effective blood supply to pulmonary and hepatic neoplasms.

Authors:  C Miyazaki
Journal:  Ann Nucl Med       Date:  1994-05       Impact factor: 2.668

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.