| Literature DB >> 29695126 |
Hong Heng See1,2, Nor Akma Mamat3,4, Peter C Hauser5.
Abstract
A method for on-line matrix elimination to enable selective quantification of ultraviolet absorbing analytes by a flow-injection analysis procedure is described. Selectivity is achieved by electric field driven extraction across a polymer inclusion membrane. The method was demonstrated on the example of the determination of naproxen from spiked human urine. Membranes of 10 μm thickness were employed which consisted of 7.5 mg cellulose triacetate as base polymer, 5 mg of o-nitrophenyl octyl ether as plasticizer and 7.5 mg of Aliquat 336 as cationic carrier. Ten μL of sample was introduced into a continuous stream of background solution consisting of 100 µM aqueous NaClO₄ with a flow rate of 2 μL/min while applying a voltage of 150 V to the extraction cell. The target ion was electrokinetically transported across the membrane and enriched in 1.5 μL of a stagnant acceptor solution. This was subsequently pumped past a flow-through UV detector for quantification. The method showed a linear range from 5 to 200 µM with a correlation coefficient of 0.9978 and a reproducibility of typically 7% (n = 8). The detection limit of the method for naproxen was 2 µM.Entities:
Keywords: electromembrane extraction; flow-injection analysis; matrix elimination; miniaturization
Mesh:
Year: 2018 PMID: 29695126 PMCID: PMC6100346 DOI: 10.3390/molecules23051000
Source DB: PubMed Journal: Molecules ISSN: 1420-3049 Impact factor: 4.411
Figure 1Schematic drawing of the electromembrane extraction (EME)-cell and the associated manifold including the UV-absorption detector. Not drawn to scale.
Typical operating sequence.
| Step | Operation | Position of Injector Valve | Donor Flow Rate (μL/min) | Acceptor Flow Rate (μL/min) |
|---|---|---|---|---|
| 1 | Flushing of manifold | Load | 2 | 2 |
| 2 | Turning off acceptor flow | Load | 2 | 0 |
| 3 | Injecting sample into injection valve | Load | 2 | 0 |
| 4 | Introducing sample into extraction cell and turning on high voltage supply | Inject | 2 | 0 |
| 5 | Turning off high voltage supply | Inject | 2 | 0 |
| 6 | Turning on acceptor pump and purging acceptor solution through UV-detector | Inject | 2 | 1–4 |
| 7 | Flushing of manifold | Load | 2 | 2 |
Figure 2Absorbance spectra of naproxen (10 mg/100 mL methanol) and urine (diluted 1:3 (v/v) with methanol and then centrifuged).
Figure 3Effect of the content of Aliquat 336 in membranes based on 7.5 mg cellulose triacetate (CTA) and 5 mg o-nitrophenyl octyl ether (NPOE) on the extraction efficiency of naproxen. Donor background: 100 μM NaClO4 at 2 μL/min; acceptor phase: 100 μM NaClO4 in 1.5 µL; standard: 10 μL of 50 μM naproxen in 100 μM NaClO4, applied voltage: 150 V. Each data point represents the mean value for five measurements.
Figure 4Effect of donor background flow rate on the extraction efficiency of naproxen. Membrane: CTA 7.5 mg, NPOE 5 mg, Aliquat 336 7.5 mg. Other conditions as for Figure 3. Each data point represents the mean value for five measurements.
Standard deviations and recoveries of naproxen from spiked urine samples.
| Amount Added (μM) | Amount Found (μM) | Relative Recovery (%) | RSD% ( |
|---|---|---|---|
| 200 | 197.0 | 98.5 | 5.5 |
| 50 | 49.5 | 99.0 | 4.8 |