| Literature DB >> 29691080 |
Teemu Karjalainen1, Konsta Pamilo1, Aleksi Reito2.
Abstract
We describe 2 cases of articulation-related failures resulting in revision surgery after a Motec total wrist arthroplasty: one with an adverse reaction to metal debris and the other with an adverse reaction to polyether ether ketone. In the first patient, blood cobalt and chrome levels were elevated and magnetic resonance imaging showed clear signs of a pseudotumor. The other patient had an extensive release of polyether ether ketone particles into the surrounding synovia due to adverse wear conditions in the cup, leading to the formation of a fluid-filled cyst sac with a black lining and diffuse lymphocyte-dominated inflammation in the synovia. We recommend regular follow-up including x-rays, monitoring of cobalt and chrome ion levels, and a low threshold for cross-sectional imaging in patients who have undergone total wrist arthroplasty with a Motec joint prosthesis. Wear-related problems can also develop in implants in which polyether ether ketone is the bulk material.Entities:
Mesh:
Substances:
Year: 2018 PMID: 29691080 DOI: 10.1016/j.jhsa.2018.03.010
Source DB: PubMed Journal: J Hand Surg Am ISSN: 0363-5023 Impact factor: 2.230