Pre-treatment verification of lung SBRT VMAT plans with delivery errors: Toward a better understanding of the gamma index analysis.

Thahabah Alharthi1, Elise M Pogson2, Sankar Arumugam3, Lois Holloway4, David Thwaites5.   

Abstract

PURPOSE: To study the sensitivity of the ArcCHECK in detecting delivery errors for lung stereotactic body radiotherapy (SBRT) using the Volumetric Modulated Arc Therapy (VMAT) technique and to evaluate the sensitivity of eight global and local gamma tolerances with different cut-off percentages.
METHODS: Baseline VMAT plans were generated for 15 lung SBRT patients. We delivered the smallest errors(gantry, collimator, and multileaf collimator MLC) which had ≥ ±2% dose difference in the modified treatment plans compared to the baseline plan (the clinical significance of those errors were assessed in our previous study. A total of 100 plan in which 15 baseline plans were measured using the ArcCheck detector along with ion chamber measurements. The sensitivity of the global and local gamma-index method using criteria of 1%/1 mm, 2%/1 mm, 2%/2 mm, and 3%/3 mm was investigated.
RESULTS: The gamma (γ) pass rates for these plans exhibited considerable spread. The majority of simulated errors were not detected. Broadly similar detection levels were achieved with the different gamma criteria and cut-offs. Combining ion chamber measurements with ArcCHECK did not improve error detection.
CONCLUSIONS: Commonly adopted gamma criteria are not sensitive enough to validate lung SBRT VMAT plans at the 2% dose difference level. The error detection levels are fairly consistent despite changes in gamma criteria and cut-offs. The choice of gamma criteria was not significant and there was no clear benefit in tightening the gamma criteria.
Copyright © 2018 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.

Entities:  

Keywords:  ArcCHECK; Delivery error; Gamma index; Lung SBRT VMAT

Mesh:

Year:  2018        PMID: 29685425     DOI: 10.1016/j.ejmp.2018.04.005

Source DB:  PubMed          Journal:  Phys Med        ISSN: 1120-1797            Impact factor:   2.685


  3 in total

1.  Evaluation of the ability of three commercially available dosimeters to detect systematic delivery errors in step-and-shoot IMRT plans.

Authors:  Alison Gray; Omemh Bawazeer; Sankar Arumugam; Philip Vial; Joseph Descallar; David Thwaites; Lois Holloway
Journal:  Rep Pract Oncol Radiother       Date:  2021-09-30

2.  Assessment of Statistical Process Control Based DVH Action Levels for Systematic Multi-Leaf Collimator Errors in Cervical Cancer RapidArc Plans.

Authors:  Hanyin Zhang; Wenli Lu; Haixia Cui; Ying Li; Xin Yi
Journal:  Front Oncol       Date:  2022-05-18       Impact factor: 5.738

3.  Study on the ability of 3D gamma analysis and bio-mathematical model in detecting dose changes caused by dose-calculation-grid-size (DCGS).

Authors:  Han Bai; Sijin Zhu; Xingrao Wu; Xuhong Liu; Feihu Chen; Jiawen Yan
Journal:  Radiat Oncol       Date:  2020-07-06       Impact factor: 3.481

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.