Literature DB >> 29683103

In vivo experiences with magnetic resonance imaging scans in Vibrant Soundbridge type 503 implantees.

I Todt1, P Mittmann1, A Ernst1, S Mutze2, G Rademacher2.   

Abstract

OBJECTIVE: To observe the effects of magnetic resonance imaging scans in Vibrant Soundbridge 503 implantees at 1.5T in vivo.
METHODS: In a prospective case study of five Vibrant Soundbridge 503 implantees, 1.5T magnetic resonance imaging scans were performed with and without a headband. The degree of pain was evaluated using a visual analogue scale. Scan-related pure tone audiogram and audio processor fitting changes were assessed.
RESULTS: In all patients, magnetic resonance imaging scans were performed without any degree of pain or change in pure tone audiogram or audio processor fitting, even without a headband.
CONCLUSION: In this series, 1.5T magnetic resonance imaging scans were performed with the Vibrant Soundbridge 503 without complications. Limitations persist in terms of magnetic artefacts.

Entities:  

Keywords:  Magnetic Resonance Imaging; Middle Ear Implants; Postoperative Complications

Mesh:

Year:  2018        PMID: 29683103     DOI: 10.1017/S0022215118000518

Source DB:  PubMed          Journal:  J Laryngol Otol        ISSN: 0022-2151            Impact factor:   1.469


  3 in total

Review 1.  [MRI in patients with auditory implants equipped with implanted magnets-an update : Overview and procedural management].

Authors:  S Nospes; M A Brockmann; A Läßig
Journal:  Radiologe       Date:  2019-01       Impact factor: 0.635

2.  Two-phase survey on the frequency of use and safety of MRI for hearing implant recipients.

Authors:  Paul van de Heyning; Griet Mertens; Vedat Topsakal; Ruben de Brito; Wilhelm Wimmer; Marco D Caversaccio; Stefan Dazert; Stefan Volkenstein; Mario Zernotti; Lorne S Parnes; Hinrich Staecker; Iain A Bruce; Gunesh Rajan; Marcus Atlas; Peter Friedland; Piotr H Skarzynski; Serafima Sugarova; Vladislav Kuzovkov; Abdulrahman Hagr; Robert Mlynski; Joachim Schmutzhard; Shin-Ichi Usami; Luis Lassaletta; Javier Gavilán; Benoit Godey; Christopher H Raine; Rudolf Hagen; Georg M Sprinzl; Kevin Brown; Wolf-Dieter Baumgartner; Eva Karltorp
Journal:  Eur Arch Otorhinolaryngol       Date:  2021-03-31       Impact factor: 2.503

3.  Active Middle Ear Implant in a Patient with Neurofibromatosis Type 1 and Multiple Calvarial Defects: A Case Report.

Authors:  Kei Kajihara; Akira Ganaha; Keiji Matsuda; Takeshi Nakamura; Tetsuya Tono
Journal:  J Int Adv Otol       Date:  2022-03       Impact factor: 1.316

  3 in total