Literature DB >> 29663665

Investigation of nano-sized debris released from CoCrMo secondary interfaces in total hip replacements: Digestion of the flakes.

Alina M Crainic1, Mauro Callisti1,2, Martin R Palmer3, Richard B Cook1.   

Abstract

The in vivo release of wear debris and corrosion products from the metallic interfaces of total hip replacements is associated with a wide spectrum of adverse body reactions and systemic manifestations. The origin of debris and the electrochemical conditions at the sites of material loss both play a role in determining the physicochemical characteristics of the particles, and thus influence their in vivo reactivity. Debris retrieved from revised CoCrMo tapers and cement-stem interfaces consists of heterogeneous flakes that comprise mechanically mixed metal particles, corrosion products and organic material. Detailed investigation of the size and composition of the metal debris contained within these composites requires the digestion of the flakes to release the small metal particles. Here, we compare alkaline and enzymatic digestion methods that both aim to fragment the flakes and reveal their smallest building blocks. The characterization of debris cleaned with both methods revealed crystalline Cr oxide nanoparticles and clusters. Comparison between the treatments showed that the alkaline method is more efficient in fragmenting the flakes and provided cleaner and generally smaller nanoparticles than exhibited in debris released with the enzymatic treatment. The provision of cleaner nanoparticles from the alkaline method also allows the physicochemical properties of the particles to be more clearly identified.
© 2018 Wiley Periodicals, Inc. J Biomed Mater Res Part B: Appl Biomater, 107B: 424-434, 2019. © 2018 Wiley Periodicals, Inc.

Entities:  

Keywords:  digestion of wear and corrosion flakes; electron microscopy; hip replacements; nanoparticle characterization; taper and stem debris

Mesh:

Substances:

Year:  2018        PMID: 29663665     DOI: 10.1002/jbm.b.34134

Source DB:  PubMed          Journal:  J Biomed Mater Res B Appl Biomater        ISSN: 1552-4973            Impact factor:   3.368


  2 in total

1.  In Vitro Evidence for Cell-Accelerated Corrosion Within Modular Junctions of Total Hip Replacements.

Authors:  Divya Rani Bijukumar; Shruti Salunkhe; Dalton Morris; Abhijith Segu; Deborah J Hall; Robin Pourzal; Mathew T Mathew
Journal:  J Orthop Res       Date:  2019-09-01       Impact factor: 3.494

2.  Simultaneous Characterization of Implant Wear and Tribocorrosion Debris within Its Corresponding Tissue Response Using Infrared Chemical Imaging.

Authors:  Songyun Liu; Deborah J Hall; Craig J Della Valle; Michael J Walsh; Joshua J Jacobs; Robin Pourzal
Journal:  Biotribology (Oxf)       Date:  2021-02-19