| Literature DB >> 29657596 |
Tatsuhiko Hirao1, Junichi Nitta1, Akira Sato1, Yoshihide Takahashi2, Masahiko Goya2, Kenzo Hirao2.
Abstract
A 46-year-old female with a body mass index of 38.9 kg/m2 and no organic heart disease underwent a subcutaneous implantable cardioverter-defibrillator implantation for secondary prevention of sudden cardiac death in the setting of idiopathic ventricular fibrillation. Defibrillation threshold (DFT) testing during implantation detected high shock impedance and high DFT. Fluoroscopy revealed subcoil fat between the lead and the sternum, which we suspected was the reason for the high shock impedance and high DFT. We repositioned the lead to a site just above the sternum and the shock impedance and DFT improved to within the respective normal ranges.Entities:
Keywords: fat layer; high defibrillation threshold; obesity; subcutaneous implantable cardiac defibrillator; ventricular fibrillation
Year: 2018 PMID: 29657596 PMCID: PMC5891406 DOI: 10.1002/joa3.12033
Source DB: PubMed Journal: J Arrhythm ISSN: 1880-4276
Figure 1Positioning of the lead in fluoroscopy In the right anterior oblique (RAO) 70‐degree view, there was a space between the lead and sternum that included fat tissue (left); this was the reason for the high DFT. We repositioned the lead to a site closer to the sternum (right). After repositioning, the DFT testing results improved. RAO, right anterior oblique; DFT, defibrillation threshold
Figure 2Positioning of the lead and generator in X‐ray After the implantation, the positioning of lead and generator was confirmed to be normal