Ultrasound-guided internal jugular vein cannulation: Can an artery be missed?

Gyaninder Pal Singh1, Vikas Chauhan1, Indu Kapoor1, Abhyuday Kumar2.   

Abstract

Entities:  

Year:  2018        PMID: 29643643      PMCID: PMC5885434          DOI: 10.4103/0970-9185.227391

Source DB:  PubMed          Journal:  J Anaesthesiol Clin Pharmacol        ISSN: 0970-9185


× No keyword cloud information.
Internal jugular vein (IJV) cannulation using ultrasound guidance has become a common practice as it provides improved safety profile.[1] Here we report a case of arterial cannulation occurring despite delineation of the carotid artery (CA) using ultrasound guidance. A 10-year-old female patient was posted for correction of kyphoscoliosis. In the operating room, the patient was prepared for the placement of central venous catheter into the right IJV. A 7.5 MHz linear probe of an ultrasound machine (Sonosite Inc. Bothell USA), was placed in short axis view. Both right IJV and CA were visualized, with CA being medial to IJV [Figure 1]. Under real-time ultrasound, IJV was punctured and a 6.0 Fr central venous catheter (Multicath 2, Vygon, France) was inserted. On transducing catheter, the monitor showed an arterial waveform and pressure values. To confirm, the probe was placed longitudinally, and it was discovered that catheter was not present in the IJV. Placing the probe in oblique axis, revealed two pulsating vessels (arteries), on each side of IJV [Figure 2a] and the catheter visualized entering into the artery lateral to the IJV [Figure 2b]. The effort to obtain a single view of all three vessels in longitudinal axis was not successful. In the short axis, 2 ml 0.9% normal saline was injected into the catheter and fluid was seen entering in the CA [Figure 3]. Thus, the catheter was removed, and pressure was applied for an adequate period. No evidence of hematoma was seen on ultrasound. During the second attempt, the ultrasound probe was placed obliquely, midway between the long and short axis (medial oblique axis) and IJV was cannulated successfully. It was further confirmed by short and long axis views.
Figure 1

Short axis view on ultrasound showing the right carotid artery medially (arrow A) and the right internal jugular vein laterally (arrow B)

Figure 2

(a) Oblique axis view showing two arteries (white arrows) one each on medial and lateral aspect of internal jugular vein. (b) Central venous catheter seen entering into the artery lateral to internal jugular vein. Black arrows showing the path of central venous catheter

Figure 3

(a) Short axis view showing the carotid artery (medial) and internal jugular vein (lateral) and path of catheter (white arrow). Catheter entered the artery on lateral aspect passing through the internal jugular vein (not visible in short axis view). (b) Saline injected through the catheter seen entering the carotid artery (black arrow)

Short axis view on ultrasound showing the right carotid artery medially (arrow A) and the right internal jugular vein laterally (arrow B) (a) Oblique axis view showing two arteries (white arrows) one each on medial and lateral aspect of internal jugular vein. (b) Central venous catheter seen entering into the artery lateral to internal jugular vein. Black arrows showing the path of central venous catheter (a) Short axis view showing the carotid artery (medial) and internal jugular vein (lateral) and path of catheter (white arrow). Catheter entered the artery on lateral aspect passing through the internal jugular vein (not visible in short axis view). (b) Saline injected through the catheter seen entering the carotid artery (black arrow) Despite the use of ultrasound, arterial puncture has been described in up to 4% of IJV cannulation.[2] In our case, the saline injected was seen entering into the CA, which suggested that the catheterized artery to be in continuation of CA. Literature shows that CA can have various anatomical variations.[3] Two main ultrasound approaches (long- and short-axis) are used for IJV cannulation. These two techniques have inherent advantages and risks, and neither technique best prevents CA puncture reliably.[4] A continuous scan in long axis can depict intravascular guidewire and catheter placement. However, a medial oblique approach allows for optimal imaging of the IJV and CA side by side and to follow the needle throughout from skin insertion to vessel penetration.[5] The oblique rotation helped us identifying the aberrant cannulated artery. We hypothesize that the cannulated artery was either a loop of ICA with its distal part lying lateral to IJV and not visualized on short axis view or an aberrant branch of ICA. For IJV cannulation, ultrasound requires technical expertise involving different approach techniques. Careful maneuvering of the probe and not depending on any single approach can help in detecting any anatomical variation of CA, thereby decreasing the risk of arterial puncture.

Financial support and sponsorship

Nil.

Conflicts of interest

There are no conflicts of interest.
  5 in total

1.  Bilateral aberrant cervical internal carotid arteries.

Authors:  D J Shanley
Journal:  Neuroradiology       Date:  1992       Impact factor: 2.804

Review 2.  The "medial-oblique" approach to ultrasound-guided central venous cannulation--maximize the view, minimize the risk.

Authors:  Ralph Dilisio; Alexander J C Mittnacht
Journal:  J Cardiothorac Vasc Anesth       Date:  2012-06-09       Impact factor: 2.628

3.  Implantation of 3951 long-term central venous catheters: performances, risk analysis, and patient comfort after ultrasound-guidance introduction.

Authors:  Adriano Peris; Giovanni Zagli; Manuela Bonizzoli; Giovanni Cianchi; Marco Ciapetti; Rosario Spina; Valentina Anichini; Francesco Lapi; Stefano Batacchi
Journal:  Anesth Analg       Date:  2010-09-09       Impact factor: 5.108

4.  Long versus short axis ultrasound guided approach for internal jugular vein cannulation: a prospective randomised controlled trial.

Authors:  Suresh Chittoodan; Dorothy Breen; Brian D O'Donnell; Gabrielle Iohom
Journal:  Med Ultrason       Date:  2011-03       Impact factor: 1.611

5.  An unseen danger: frequency of posterior vessel wall penetration by needles during attempts to place internal jugular vein central catheters using ultrasound guidance.

Authors:  Michael Blaivas; Srikar Adhikari
Journal:  Crit Care Med       Date:  2009-08       Impact factor: 7.598

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.