Residual moisture and waterborne pathogens inside flexible endoscopes: Evidence from a multisite study of endoscope drying effectiveness.

Cori L Ofstead1, Otis L Heymann2, Mariah R Quick2, John E Eiland2, Harry P Wetzler2.   

Abstract

BACKGROUND: Endoscopy-associated infection transmission is frequently linked to inadequate reprocessing. Residual organic material and moisture may foster biofilm development inside endoscopes. This study evaluated the effectiveness of endoscope drying and storage methods and assessed associations between retained moisture and contamination.
METHODS: Endoscope reprocessing, drying, and storage practices were assessed at 3 hospitals. Researchers performed visual examinations and tests to detect fluid and contamination on patient-ready endoscopes.
RESULTS: Fluid was detected in 22 of 45 (49%) endoscopes. Prevalence of moisture varied significantly by site (5%; 83%; 85%; P < .001). High adenosine triphosphate levels were found in 22% of endoscopes, and microbial growth was detected in 71% of endoscopes. Stenotrophomonas maltophilia, Citrobacter freundii, and Lecanicillium lecanii/Verticillium dahliae were found. Retained fluid was associated with significantly higher adenosine triphosphate levels (P < .01). Reprocessing and drying practices conformed with guidelines at 1 site and were substandard at 2 sites. Damaged endoscopes were in use at all sites.
CONCLUSIONS: Inadequate reprocessing and insufficient drying contributed to retained fluid and contamination found during this multisite study. More effective methods of endoscope reprocessing, drying, and maintenance are needed to prevent the retention of fluid, organic material, and bioburden that could cause patient illness or injury.
Copyright © 2018 Association for Professionals in Infection Control and Epidemiology, Inc. Published by Elsevier Inc. All rights reserved.

Entities:  

Keywords:  Adenosine triphosphate; Contamination; Endoscope; Infection; Reprocessing; Stenotrophomonas maltophilia

Mesh:

Substances:

Year:  2018        PMID: 29609854     DOI: 10.1016/j.ajic.2018.03.002

Source DB:  PubMed          Journal:  Am J Infect Control        ISSN: 0196-6553            Impact factor:   2.918


  13 in total

Review 1.  Recent advances in understanding the epidemiology of healthcare-associated infections.

Authors:  Pranavi Sreeramoju
Journal:  F1000Res       Date:  2019-01-25

Review 2.  Emerging next-generation robotic colonoscopy systems towards painless colonoscopy.

Authors:  Chung-Kwong Yeung; Jo Lk Cheung; Biji Sreedhar
Journal:  J Dig Dis       Date:  2019-04       Impact factor: 2.325

Review 3.  Duodenoscope-associated infection prevention: A call for evidence-based decision making.

Authors:  Cori L Ofstead; Brandy L Buro; Krystina M Hopkins; John E Eiland; Harry P Wetzler; David R Lichtenstein
Journal:  Endosc Int Open       Date:  2020-11-17

4.  Feasibility study of a single-use balloon-assisted robotic colonoscope in healthy volunteers.

Authors:  Chi-Chung Foo; Wai-Keung Leung; Thomas Ka-Luen Lui; Jo Lai-Ken Cheung; Kwok-Wai Lam; Biji Sreedhar; Chung-Kwong Yeung
Journal:  Endosc Int Open       Date:  2021-03-17

5.  Advances in the Microbiology of Stenotrophomonas maltophilia.

Authors:  Joanna S Brooke
Journal:  Clin Microbiol Rev       Date:  2021-05-26       Impact factor: 50.129

Review 6.  SpyGlass application for duodenoscope working channel inspection: Impact on the microbiological surveillance.

Authors:  Tao-Chieh Liu; Chen-Ling Peng; Hsiu-Po Wang; Hsin-Hung Huang; Wei-Kuo Chang
Journal:  World J Gastroenterol       Date:  2020-07-14       Impact factor: 5.742

7.  Potential impact of contaminated bronchoscopes on novel coronavirus disease (COVID-19) patients.

Authors:  Cori L Ofstead; Krystina M Hopkins; Matthew J Binnicker; Gregory A Poland
Journal:  Infect Control Hosp Epidemiol       Date:  2020-04-02       Impact factor: 3.254

8.  Comparative Study of Microbiological Monitoring Results from Three Types of Sampling Methods after Gastrointestinal Endoscope Reprocessing.

Authors:  Su Ma; Lili Feng; Ziyi Jiang; Xian Gao; Xisha Long; Shaonan Zhuang; Wenxia Ding; Taiyao Chen; Zhaoshen Li; Lingjuan Zhang; Huijun Xi; Hongzhi Zhang
Journal:  Biomed Res Int       Date:  2019-12-03       Impact factor: 3.411

Review 9.  Recent Advances in Endoscope Disinfection: Where Do We Stand in the COVID era?

Authors:  Tiffany Chua; Nasir Halim; Sofiya Reicher
Journal:  Tech Innov Gastrointest Endosc       Date:  2020-10-19

10.  Establishment of a quality control circle to reduce biofilm formation in flexible endoscopes by improvement of qualified cleaning rate.

Authors:  Yingxia Luo; Qixuan Yang; Bingkun Li; Yao Yao
Journal:  J Int Med Res       Date:  2020-09       Impact factor: 1.671

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.