Commissioning and validation of commercial deformable image registration software for adaptive contouring.

Swamidas V Jamema1, Reena Phurailatpam2, Siji N Paul2, Kishore Joshi2, D D Deshpande3.   

Abstract

PURPOSE: To report the commissioning and validation of deformable image registration(DIR) software for adaptive contouring.
METHODS: DIR (SmartAdapt®v13.6) was validated using two methods namely contour propagation accuracy and landmark tracking, using physical phantoms and clinical images of various disease sites. Five in-house made phantoms with various known deformations and a set of 10 virtual phantoms were used. Displacement in lateral, anterio-posterior (AP) and superior-inferior (SI) direction were evaluated for various organs and compared with the ground truth. Four clinical sites namely, brain (n = 5), HN (n = 9), cervix (n = 18) and prostate (n = 23) were used. Organs were manually delineated by a radiation oncologist, compared with the deformable image registration (DIR) generated contours. 3D slicer v4.5.0.1 was used to analyze Dice Similarity Co-efficient (DSC), shift in centre of mass (COM) and Hausdorff distances Hf95%/avg.
RESULTS: Mean (SD) DSC, Hf95% (mm), Hfavg (mm) and COM of all the phantoms 1-5 were 0.84 (0.2) mm, 5.1 (7.4) mm, 1.6 (2.2) mm, and 1.6 (0.2) mm respectively. Phantom-5 had the largest deformation as compared to phantoms 1-4, and hence had suboptimal indices. The virtual phantom resulted in consistent results for all the ROIs investigated. Contours propagated for brain patients were better with a high DSC score (0.91 (0.04)) as compared to other sites (HN: 0.84, prostate: 0.81 and cervix 0.77). A similar trend was seen in other indices too. The accuracy of propagated contours is limited for complex deformations that include large volume and shape change of bladder and rectum respectively. Visual validation of the propagated contours is recommended for clinical implementation.
CONCLUSION: The DIR algorithm was commissioned and validated for adaptive contouring.
Copyright © 2018 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.

Entities:  

Keywords:  Commissioning; Deformable image registration; Validation

Mesh:

Year:  2018        PMID: 29609810     DOI: 10.1016/j.ejmp.2018.01.013

Source DB:  PubMed          Journal:  Phys Med        ISSN: 1120-1797            Impact factor:   2.685


  2 in total

Review 1.  Registration methods in radiotherapy.

Authors:  Paweł Czajkowski; Tomasz Piotrowski
Journal:  Rep Pract Oncol Radiother       Date:  2018-10-10

2.  Quantifying the accuracy of deformable image registration for cone-beam computed tomography with a physical phantom.

Authors:  Richard Y Wu; Amy Y Liu; Tyler D Williamson; Jinzhong Yang; Paul G Wisdom; Xiaorong R Zhu; Steven J Frank; Clifton D Fuller; Gary B Gunn; Song Gao
Journal:  J Appl Clin Med Phys       Date:  2019-09-21       Impact factor: 2.102

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.