Optimisation of radiation dose and image quality in mobile neonatal chest radiography.

V I Hinojos-Armendáriz1, S J Mejía-Rosales2, M C Franco-Cabrera3.   

Abstract

PURPOSE: To optimise the radiation dose and image quality for chest radiography in the neonatal intensive care unit (NICU) by increasing the mean beam energy.
METHODS: Two techniques for the acquisition of NICU AP chest X-ray images were compared for image quality and radiation dose. 73 images were acquired using a standard technique (56 kV, 3.2 mAs and no additional filtration) and 90 images with a new technique (62 kV, 2 mAs and 2 mm Al filtration). The entrance surface air kerma (ESAK) was measured using a phantom and compared between the techniques and against established diagnostic reference levels (DRL). Images were evaluated using seven image quality criteria independently by three radiologists. Images quality and radiation dose were compared statistically between the standard and new techniques.
RESULTS: The maximum ESAK for the new technique was 40.20 μGy, 43.7% of the ESAK of the standard technique. Statistical evaluation demonstrated no significant differences in image quality between the two acquisition techniques.
CONCLUSIONS: Based on the techniques and acquisition factors investigated within this study, it is possible to lower the radiation dose without any significant effects on image quality by adding filtration (2 mm Al) and increasing the tube potential. Such steps are relatively simple to undertake and as such, other departments should consider testing and implementing this dose reduction strategy within clinical practice where appropriate.
Copyright © 2017 The College of Radiographers. Published by Elsevier Ltd. All rights reserved.

Entities:  

Keywords:  Chest radiography; Neonatal radiography; Optimisation; Paediatric radiology; Radiation protection

Mesh:

Year:  2017        PMID: 29605105     DOI: 10.1016/j.radi.2017.09.004

Source DB:  PubMed          Journal:  Radiography (Lond)        ISSN: 1078-8174


  4 in total

1.  Neonatal digital chest radiography- should we be using additional copper filtration?

Authors:  Jenna Ruth Tugwell-Allsup; Rhys Wyn Morris; Kate Thomas; Richard Hibbs; Andrew England
Journal:  Br J Radiol       Date:  2021-12-14       Impact factor: 3.039

2.  An anthropomorphic phantom representing a prematurely born neonate for digital x-ray imaging using 3D printing: Proof of concept and comparison of image quality from different systems.

Authors:  Nikolaus Irnstorfer; Ewald Unger; Azadeh Hojreh; Peter Homolka
Journal:  Sci Rep       Date:  2019-10-07       Impact factor: 4.379

3.  Paediatric image-guided radiation therapy: determining and evaluating appropriate kilovoltage planar exposure factors for the Varian on-board imager.

Authors:  John Ryan; David Willis
Journal:  J Med Radiat Sci       Date:  2019-09-03

4.  Influence of the use of various imaging units and projections on the radiation dose received by children during chest digital radiography.

Authors:  Hongrong Xu; Kaiping Huang; Bo Liu; Jinhua Cai; Huan Zheng; Helin Zheng; Qiurui Yang; Changhong Yao
Journal:  PLoS One       Date:  2021-08-05       Impact factor: 3.240

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.