Role of an Intensity-Transformation Function in Enhancement of Bone Scintigraphy Images.

Anil Kumar Pandey1, Vishali Dhiman2, Akshima Sharma2, Sreedharan Thankarajan ArunRaj2, Vivek Baghel2, Chetan Patel2, Param Dev Sharma3, Chandra Sekhar Bal2, Rakesh Kumar1.   

Abstract

Bone scintigraphy images might exceed the dynamic range (the ratio between the highest and the lowest displayable brightness) of the monitor. In such a case, a high-intensity area accompanied by loss of detail in other structures in the displayed image make the clinical interpretation challenging. We have investigated the role of an intensity-transformation (IT) function in enhancement of these types of images.
Methods: Forty high-dynamic-range bone scintigraphy images were processed using an IT function. The IT function has 2 parameters: threshold and slope. With the threshold kept equal to the mean count of the image, the slope was varied from 1 to 20. A software program developed in-house was used to process the images. Twenty output images corresponding to one input image were visually inspected by 2 experienced nuclear medicine physicians to select images of diagnostic quality, and from their selection was determined the standardized slope that produced the maximum number of diagnostic images. The 2 physicians also scored the quality of the input and output images (at the standardized slope) on a scale of 1-5. The Student t test was used to determine the significance of differences in mean score between the input and output images at an α significance level of 0.05.
Results: Application of the IT function with standardized parameters significantly improved the quality of high-dynamic-range bone scintigraphy images (P < 0.001, with α = 0.05). A slope of 8 maximized the number of diagnostic images.
Conclusion: The IT function has a significant role in enhancing high-dynamic-range bone scintigraphy images.
© 2018 by the Society of Nuclear Medicine and Molecular Imaging.

Keywords:  high-dynamic-range bone scintigraphy image; image enhancement; intensity-transformation function

Mesh:

Year:  2018        PMID: 29599398     DOI: 10.2967/jnmt.117.202929

Source DB:  PubMed          Journal:  J Nucl Med Technol        ISSN: 0091-4916


  2 in total

1.  Improving technetium-99m methylene diphosphonate bone scan images using histogram specification technique.

Authors:  Anil Kumar Pandey; Param Dev Sharma; Akshima Sharma; Ashish Negi; Girish Kumar Parida; Harish Goyal; Chandra Sekhar Bal; Rakesh Kumar
Journal:  World J Nucl Med       Date:  2020-07-22

2.  Methylene Diphosphonate Bone Scan Scintigraphic Image Enhancement using Gamma Correction and Optimizing the Value of Gamma.

Authors:  Anshul Sharma; Anil Kumar Pandey; Deepak Khichi; Rakesh Kumar
Journal:  Indian J Nucl Med       Date:  2019-12-31
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.