Analytical interference of HBOC-201 (Hemopure, a synthetic hemoglobin-based oxygen carrier) on four common clinical chemistry platforms.

Erik A Korte1, Nicole Pozzi2, Nina Wardrip3, M Tayyeb Ayyoubi1, Saeed A Jortani4.   

Abstract

BACKGROUND: There are 13 million blood transfusions each year in the US. Limitations in the donor pool, storage capabilities, mass casualties, access in remote locations and reactivity of donors all limit the availability of transfusable blood products to patients. HBOC-201 (Hemopure®) is a second-generation glutaraldehyde-polymer of bovine hemoglobin, which can serve as an "oxygen bridge" to maintain oxygen carrying capacity while transfusion products are unavailable. Hemopure presents the advantages of extended shelf life, ambient storage, and limited reactive potential, but its extracellular location can also cause significant interference in modern laboratory analyzers similar to severe hemolysis.
METHODS: Observed error in 26 commonly measured analytes was determined on 4 different analytical platforms in plasma from a patient therapeutically transfused Hemopure as well as donor blood spiked with Hemopure at a level equivalent to the therapeutic loading dose (10% v/v).
RESULTS: Significant negative error ratios >50% of the total allowable error (>0.5tAE) were reported in 23/104 assays (22.1%), positive bias of >0.5tAE in 26/104 assays (25.0%), and acceptable bias between -0.5tAE and 0.5tAE error ratio was reported in 44/104 (42.3%). Analysis failed in the presence of Hemopure in 11/104 (10.6%). Observed error is further subdivided by platform, wavelength, dilution and reaction method.
CONCLUSION: Administration of Hemopure (or other hemoglobin-based oxygen carriers) presents a challenge to laboratorians tasked with analyzing patient specimens. We provide laboratorians with a reference to evaluate patient samples, select optimal analytical platforms for specific analytes, and predict possible bias beyond the 4 analytical platforms included in this study.
Copyright © 2018 Elsevier B.V. All rights reserved.

Entities:  

Keywords:  Absorbance; Anemia; Bias; Extracellular hemoglobin; Hemoglobin-based oxygen carrier; Hemopure; Interference; Transfusion

Mesh:

Substances:

Year:  2018        PMID: 29580857     DOI: 10.1016/j.cca.2018.03.017

Source DB:  PubMed          Journal:  Clin Chim Acta        ISSN: 0009-8981            Impact factor:   3.786


  2 in total

1.  Management of thymoma-associated pure red cell aplasia: A novel use of blood substitute HBOC-201 in a Jehovah's Witness.

Authors:  Maria M Rubinstein; Cheryl Goss; Scott T Avecilla; Gregory P Dubé; Gregory J Riely; Jodi V Mones
Journal:  Clin Case Rep       Date:  2019-12-26

2.  Sublingual Microcirculatory Characteristics of a Case of Profound Chemotherapy-Induced Anemia Treated With a Hemoglobin-Based Oxygen Carrier.

Authors:  Jason Stankiewicz; Maniraj Jeyaraju; Sanjay Maheshwari; Andrew R Deitchman; Michael T McCurdy
Journal:  Cureus       Date:  2021-05-15
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.