Literature DB >> 29580829

Sweat test for cystic fibrosis: Wearable sweat sensor vs. standard laboratory test.

Dong-Hoon Choi1, Abigail Thaxton2, In Cheol Jeong3, Kain Kim1, Patrick R Sosnay2, Garry R Cutting4, Peter C Searson5.   

Abstract

BACKGROUND: Sweat chloride testing for diagnosis of cystic fibrosis (CF) involves sweat induction, collection and handling, and measurement in an analytical lab. We have developed a wearable sensor with an integrated salt bridge for real-time measurement of sweat chloride concentration. Here, in a proof-of-concept study, we compare the performance of the sensor to current clinical practice in CF patients and healthy subjects.
METHOD: Sweat was induced on both forearms of 10 individuals with CF and 10 healthy subjects using pilocarpine iontophoresis. A Macroduct sweat collection device was attached to one arm and sweat was collected for 30 min and then sent for laboratory analysis. A sensor was attached to the other arm and the chloride ion concentration monitored in real time for 30 min using a Bluetooth transceiver and smart phone app.
RESULTS: Stable sweat chloride measurements were obtained within 15 min following sweat induction using the wearable sensor. We define the detection time as the time at which the standard deviation of the real-time chloride ion concentration remained below 2 mEq/L for 5 min. The sweat volume for sensor measurements at the detection time was 13.1 ± 11.4 μL (SD), in many cases lower than the minimum sweat volume of 15 μL for conventional testing. The mean difference between sweat chloride concentrations measured by the sensor and the conventional laboratory practice was 6.2 ± 9.5 mEq/L (SD), close to the arm-to-arm variation of about 3 mEq/L. The Pearson correlation coefficient between the two measurements was 0.97 highlighting the excellent agreement between the two methods.
CONCLUSION: A wearable sensor can be used to make real-time measurements of sweat chloride within 15 min following sweat induction, requiring a small sweat volume, and with excellent agreement to standard methods.
Copyright © 2018 European Cystic Fibrosis Society. Published by Elsevier B.V. All rights reserved.

Entities:  

Keywords:  Cystic fibrosis; Personalized medicine; Sweat rate measurement; Sweat test; Wearable sweat chloride sensor

Mesh:

Substances:

Year:  2018        PMID: 29580829     DOI: 10.1016/j.jcf.2018.03.005

Source DB:  PubMed          Journal:  J Cyst Fibros        ISSN: 1569-1993            Impact factor:   5.482


  14 in total

1.  Wearable flexible sweat sensors for healthcare monitoring: a review.

Authors:  Michael Chung; Giuseppino Fortunato; Norbert Radacsi
Journal:  J R Soc Interface       Date:  2019-10-09       Impact factor: 4.118

2.  Silver-manganese nanocomposite modified screen-printed carbon electrode in the fabrication of an electrochemical, disposable biosensor strip for cystic fibrosis.

Authors:  P Arun Kumar; Aarathi Pradeep; Bipin Kumar G Nair; T G Satheesh Babu; Punathil Vasu Suneesh
Journal:  Mikrochim Acta       Date:  2022-08-11       Impact factor: 6.408

Review 3.  Influencing factors and drug application of iontophoresis in transdermal drug delivery: an overview of recent progress.

Authors:  Yu Wang; Lijuan Zeng; Wenting Song; Jianping Liu
Journal:  Drug Deliv Transl Res       Date:  2021-01-23       Impact factor: 4.617

4.  State of Sweat: Emerging Wearable Systems for Real-Time, Noninvasive Sweat Sensing and Analytics.

Authors:  Roozbeh Ghaffari; Da Som Yang; Joohee Kim; Amer Mansour; John A Wright; Jeffrey B Model; Donald E Wright; John A Rogers; Tyler R Ray
Journal:  ACS Sens       Date:  2021-08-05       Impact factor: 9.618

5.  The Dynamic Response of Sweat Chloride to Changes in Exercise Load Measured by a Wearable Sweat Sensor.

Authors:  Dong-Hoon Choi; Grant B Kitchen; Kerry J Stewart; Peter C Searson
Journal:  Sci Rep       Date:  2020-05-07       Impact factor: 4.379

6.  Study of the Dielectric Properties of Artificial Sweat Mixtures at Microwave Frequencies.

Authors:  Angie R Eldamak; Sarah Thorson; Elise C Fear
Journal:  Biosensors (Basel)       Date:  2020-06-09

7.  Out-of-clinic measurement of sweat chloride using a wearable sensor during low-intensity exercise.

Authors:  Dong-Hoon Choi; Grant B Kitchen; Mark T Jennings; Garry R Cutting; Peter C Searson
Journal:  NPJ Digit Med       Date:  2020-03-27

8.  PHAIR: a biosensor for pH measurement in air-liquid interface cell culture.

Authors:  Mohammadhossein Dabaghi; Neda Saraei; Gang Xu; Abiram Chandiramohan; Jonas Yeung; Jenny P Nguyen; Milica Vukmirovic; Ponnambalam Ravi Selvaganapathy; Jeremy A Hirota
Journal:  Sci Rep       Date:  2021-02-10       Impact factor: 4.379

9.  Soft, skin-interfaced sweat stickers for cystic fibrosis diagnosis and management.

Authors:  Tyler R Ray; Maja Ivanovic; Paul M Curtis; Daniel Franklin; Kerem Guventurk; William J Jeang; Joseph Chafetz; Hannah Gaertner; Grace Young; Steve Rebollo; Jeffrey B Model; Stephen P Lee; John Ciraldo; Jonathan T Reeder; Aurélie Hourlier-Fargette; Amay J Bandodkar; Jungil Choi; Alexander J Aranyosi; Roozbeh Ghaffari; Susanna A McColley; Shannon Haymond; John A Rogers
Journal:  Sci Transl Med       Date:  2021-03-31       Impact factor: 17.956

10.  Passively Addressable Ultra-Low Volume Sweat Chloride Sensor.

Authors:  Antra Ganguly; Shalini Prasad
Journal:  Sensors (Basel)       Date:  2019-10-22       Impact factor: 3.576

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.