Accuracy of Dual-Energy Virtual Monochromatic CT Numbers: Comparison between the Single-Source Projection-Based and Dual-Source Image-Based Methods.

Takashi Ueguchi1, Ryota Ogihara2, Sachiko Yamada2.   

Abstract

RATIONALE AND
OBJECTIVES: To investigate the accuracy of dual-energy virtual monochromatic computed tomography (CT) numbers obtained by two typical hardware and software implementations: the single-source projection-based method and the dual-source image-based method.
MATERIALS AND METHODS: A phantom with different tissue equivalent inserts was scanned with both single-source and dual-source scanners. A fast kVp-switching feature was used on the single-source scanner, whereas a tin filter was used on the dual-source scanner. Virtual monochromatic CT images of the phantom at energy levels of 60, 100, and 140 keV were obtained by both projection-based (on the single-source scanner) and image-based (on the dual-source scanner) methods. The accuracy of virtual monochromatic CT numbers for all inserts was assessed by comparing measured values to their corresponding true values. Linear regression analysis was performed to evaluate the dependency of measured CT numbers on tissue attenuation, method, and their interaction.
RESULTS: Root mean square values of systematic error over all inserts at 60, 100, and 140 keV were approximately 53, 21, and 29 Hounsfield unit (HU) with the single-source projection-based method, and 46, 7, and 6 HU with the dual-source image-based method, respectively. Linear regression analysis revealed that the interaction between the attenuation and the method had a statistically significant effect on the measured CT numbers at 100 and 140 keV.
CONCLUSIONS: There were attenuation-, method-, and energy level-dependent systematic errors in the measured virtual monochromatic CT numbers. CT number reproducibility was comparable between the two scanners, and CT numbers had better accuracy with the dual-source image-based method at 100 and 140 keV.
Copyright © 2018 The Association of University Radiologists. Published by Elsevier Inc. All rights reserved.

Keywords:  CT number; Dual-energy CT; beam hardening; phantom study; virtual monochromatic CT

Mesh:

Year:  2018        PMID: 29573936     DOI: 10.1016/j.acra.2018.02.022

Source DB:  PubMed          Journal:  Acad Radiol        ISSN: 1076-6332            Impact factor:   3.173


  3 in total

1.  Quantitative evaluation of disease severity in connective tissue disease-associated interstitial lung disease by dual-energy computed tomography.

Authors:  Ling Chen; Min Zhu; Haiyan Lu; Ting Yang; Wanjiang Li; Yali Zhang; Qibing Xie; Zhenlin Li; Huajing Wan; Fengming Luo
Journal:  Respir Res       Date:  2022-03-05

2.  Impact of beam-hardening corrections on proton relative stopping power estimates from single- and dual-energy CT.

Authors:  Michael S Chacko; Dee Wu; Hardev S Grewal; Jagadeesh R Sonnad
Journal:  J Appl Clin Med Phys       Date:  2022-07-11       Impact factor: 2.243

3.  Evaluating the Image Quality of Monoenergetic Images From Dual-Energy Computed Tomography With Low-Concentration and Low-Flow-Rate Contrast Media for the Arterials Supply to the Nipple-Areola Complex in Breast Cancer Compared With Conventional Computed Tomography Angiography.

Authors:  Xiangfei Zeng; Xiaoxia Wang; Huifang Chen; Daihong Liu; Jinfang Shi; Jing Zhang; Xiaoqin Li; Liyin Zhang; Yi Yang; Jiuquan Zhang
Journal:  J Comput Assist Tomogr       Date:  2020 Nov/Dec       Impact factor: 2.081

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.