Literature DB >> 2957338

Doppler microwave. A clinical assessment of its efficacy as an arterial pulse sensing technique.

M A Papp, C Hughes, J C Lin, J M Pouget.   

Abstract

A new Doppler microwave device designed for noninvasive arterial pulse detection by assessment of arterial wall motion was clinically evaluated. Carotid pulse tracings obtained with the microwave device were compared with simultaneously obtained intra-aortic pressure pulse waves. Ease and quality of the recording were also evaluated in 25 consecutive patients undergoing echocardiography. In 92% of the patients tested, good quality carotid pulse tracings could be obtained. T1/2 values and systolic time intervals were calculated from these recordings and correlated well with the known disease states of these patients, which ranged from aortic stenosis and aortic insufficiency to chronic heart failure. Finally, microwave tracings were faithful reproductions of invasive recordings of arterial pressures when compared on the basis of left ventricular ejection times and pulse contour.

Entities:  

Mesh:

Year:  1987        PMID: 2957338

Source DB:  PubMed          Journal:  Invest Radiol        ISSN: 0020-9996            Impact factor:   6.016


  1 in total

1.  A novel method for artery detection in laparoscopic surgery.

Authors:  Hamed Akbari; Yukio Kosugi; Kazunori Kihara
Journal:  Surg Endosc       Date:  2007-12-20       Impact factor: 4.584

  1 in total