Implementation of Size-Dependent Local Diagnostic Reference Levels for CT Angiography.

Hub Boere1, Nienke G Eijsvoogel1,2, Anna M Sailer1,3, Joachim E Wildberger1,2, Michiel W de Haan1, Marco Das1,4, Cecile R L P N Jeukens1.   

Abstract

OBJECTIVE: Diagnostic reference levels (DRLs) are established for standard-sized patients; however, patient dose in CT depends on patient size. The purpose of this study was to introduce a method for setting size-dependent local diagnostic reference levels (LDRLs) and to evaluate these LDRLs in comparison with size-independent LDRLs and with respect to image quality.
MATERIALS AND METHODS: One hundred eighty-four aortic CT angiography (CTA) examinations performed on either a second-generation or third-generation dual-source CT scanner were included; we refer to the second-generation dual-source CT scanner as "CT1" and the third-generation dual-source CT scanner as "CT2." The volume CT dose index (CTDIvol) and patient diameter (i.e., the water-equivalent diameter) were retrieved by dose-monitoring software. Size-dependent DRLs based on a linear regression of the CTDIvol versus patient size were set by scanner type. Size-independent DRLs were set by the 5th and 95th percentiles of the CTDIvol values. Objective image quality was assessed using the signal-to-noise ratio (SNR), and subjective image quality was assessed using a 4-point Likert scale.
RESULTS: The CTDIvol depended on patient size and scanner type (R2 = 0.72 and 0.78, respectively; slope = 0.05 and 0.02 mGy/mm; p < 0.001). Of the outliers identified by size-independent DRLs, 30% (CT1) and 67% (CT2) were adequately dosed when considering patient size. Alternatively, 30% (CT1) and 70% (CT2) of the outliers found with size-dependent DRLs were not identified using size-independent DRLs. A negative correlation was found between SNR and CTDIvol (R2 = 0.36 for CT1 and 0.45 for CT2). However, all outliers had a subjective image quality score of sufficient or better.
CONCLUSION: We introduce a method for setting size-dependent LDRLs in CTA. Size-dependent LDRLs are relevant for assessing the appropriateness of the radiation dose for an individual patient on a specific CT scanner.

Entities:  

Keywords:  CT angiography; image quality; local diagnostic reference levels; radiation dose; size-dependent diagnostic reference levels; volume CT dose index (CTDIvol)

Mesh:

Substances:

Year:  2018        PMID: 29570369     DOI: 10.2214/AJR.17.18566

Source DB:  PubMed          Journal:  AJR Am J Roentgenol        ISSN: 0361-803X            Impact factor:   3.959


  3 in total

1.  Local clinical diagnostic reference levels for chest and abdomen CT examinations in adults as a function of body mass index and clinical indication: a prospective multicenter study.

Authors:  Hugues Brat; Federica Zanca; Stéphane Montandon; Damien Racine; Benoit Rizk; Eric Meicher; Dominique Fournier
Journal:  Eur Radiol       Date:  2019-05-29       Impact factor: 5.315

2.  Potential of a machine-learning model for dose optimization in CT quality assurance.

Authors:  Axel Meineke; Christian Rubbert; Lino M Sawicki; Christoph Thomas; Yan Klosterkemper; Elisabeth Appel; Julian Caspers; Oliver T Bethge; Patric Kröpil; Gerald Antoch; Johannes Boos
Journal:  Eur Radiol       Date:  2019-02-19       Impact factor: 5.315

3.  Probability of receiving a high cumulative radiation dose and primary clinical indication of CT examinations: a 5-year observational cohort study.

Authors:  Cécile R L P N Jeukens; Hub Boere; Bart A J M Wagemans; Patty J Nelemans; Estelle C Nijssen; Rebecca Smith-Bindman; Joachim E Wildberger; Anna M Sailer
Journal:  BMJ Open       Date:  2021-01-17       Impact factor: 2.692

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.