Hub Boere1, Nienke G Eijsvoogel1,2, Anna M Sailer1,3, Joachim E Wildberger1,2, Michiel W de Haan1, Marco Das1,4, Cecile R L P N Jeukens1. 1. 1 Department of Radiology and Nuclear Medicine, Maastricht University Medical Center, P. Debyelaan 25, 6229 HX, Maastricht, The Netherlands. 2. 2 CARIM, School for Cardiovascular Diseases, Maastricht University Medical Center, Maastricht, The Netherlands. 3. 3 Department of Radiology, Stanford University School of Medicine, Stanford, CA. 4. 4 Klinik für Diagnostische und Interventionelle Radiologie, Duisburg, Germany.
Abstract
OBJECTIVE: Diagnostic reference levels (DRLs) are established for standard-sized patients; however, patient dose in CT depends on patient size. The purpose of this study was to introduce a method for setting size-dependent local diagnostic reference levels (LDRLs) and to evaluate these LDRLs in comparison with size-independent LDRLs and with respect to image quality. MATERIALS AND METHODS: One hundred eighty-four aortic CT angiography (CTA) examinations performed on either a second-generation or third-generation dual-source CT scanner were included; we refer to the second-generation dual-source CT scanner as "CT1" and the third-generation dual-source CT scanner as "CT2." The volume CT dose index (CTDIvol) and patient diameter (i.e., the water-equivalent diameter) were retrieved by dose-monitoring software. Size-dependent DRLs based on a linear regression of the CTDIvol versus patient size were set by scanner type. Size-independent DRLs were set by the 5th and 95th percentiles of the CTDIvol values. Objective image quality was assessed using the signal-to-noise ratio (SNR), and subjective image quality was assessed using a 4-point Likert scale. RESULTS: The CTDIvol depended on patient size and scanner type (R2 = 0.72 and 0.78, respectively; slope = 0.05 and 0.02 mGy/mm; p < 0.001). Of the outliers identified by size-independent DRLs, 30% (CT1) and 67% (CT2) were adequately dosed when considering patient size. Alternatively, 30% (CT1) and 70% (CT2) of the outliers found with size-dependent DRLs were not identified using size-independent DRLs. A negative correlation was found between SNR and CTDIvol (R2 = 0.36 for CT1 and 0.45 for CT2). However, all outliers had a subjective image quality score of sufficient or better. CONCLUSION: We introduce a method for setting size-dependent LDRLs in CTA. Size-dependent LDRLs are relevant for assessing the appropriateness of the radiation dose for an individual patient on a specific CT scanner.
OBJECTIVE: Diagnostic reference levels (DRLs) are established for standard-sized patients; however, patient dose in CT depends on patient size. The purpose of this study was to introduce a method for setting size-dependent local diagnostic reference levels (LDRLs) and to evaluate these LDRLs in comparison with size-independent LDRLs and with respect to image quality. MATERIALS AND METHODS: One hundred eighty-four aortic CT angiography (CTA) examinations performed on either a second-generation or third-generation dual-source CT scanner were included; we refer to the second-generation dual-source CT scanner as "CT1" and the third-generation dual-source CT scanner as "CT2." The volume CT dose index (CTDIvol) and patient diameter (i.e., the water-equivalent diameter) were retrieved by dose-monitoring software. Size-dependent DRLs based on a linear regression of the CTDIvol versus patient size were set by scanner type. Size-independent DRLs were set by the 5th and 95th percentiles of the CTDIvol values. Objective image quality was assessed using the signal-to-noise ratio (SNR), and subjective image quality was assessed using a 4-point Likert scale. RESULTS: The CTDIvol depended on patient size and scanner type (R2 = 0.72 and 0.78, respectively; slope = 0.05 and 0.02 mGy/mm; p < 0.001). Of the outliers identified by size-independent DRLs, 30% (CT1) and 67% (CT2) were adequately dosed when considering patient size. Alternatively, 30% (CT1) and 70% (CT2) of the outliers found with size-dependent DRLs were not identified using size-independent DRLs. A negative correlation was found between SNR and CTDIvol (R2 = 0.36 for CT1 and 0.45 for CT2). However, all outliers had a subjective image quality score of sufficient or better. CONCLUSION: We introduce a method for setting size-dependent LDRLs in CTA. Size-dependent LDRLs are relevant for assessing the appropriateness of the radiation dose for an individual patient on a specific CT scanner.
Entities:
Keywords:
CT angiography; image quality; local diagnostic reference levels; radiation dose; size-dependent diagnostic reference levels; volume CT dose index (CTDIvol)
Authors: Axel Meineke; Christian Rubbert; Lino M Sawicki; Christoph Thomas; Yan Klosterkemper; Elisabeth Appel; Julian Caspers; Oliver T Bethge; Patric Kröpil; Gerald Antoch; Johannes Boos Journal: Eur Radiol Date: 2019-02-19 Impact factor: 5.315
Authors: Cécile R L P N Jeukens; Hub Boere; Bart A J M Wagemans; Patty J Nelemans; Estelle C Nijssen; Rebecca Smith-Bindman; Joachim E Wildberger; Anna M Sailer Journal: BMJ Open Date: 2021-01-17 Impact factor: 2.692