Quality assurance in CT: implementation of the updated national diagnostic reference levels using an automated CT dose monitoring system.

E Appel1, P Kröpil2, O T Bethge2, J Aissa2, C Thomas2, G Antoch2, J Boos2.   

Abstract

AIM: To evaluate the implementation of the updated computed tomography (CT) diagnostic reference levels (DRLs) from the German Federal Office for Radiation Protection into clinical routine using an automatic CT dose monitoring system. METHODS AND MATERIALS: CT radiation exposure was analysed before and after implementing the updated national DRLs into routine clinical work in 2016. After the implementation process, institutional CT protocols were mapped to the anatomical regions for which DRLs were provided. Systematically, protocols that exceeded the thresholds were optimised and analysed in detail. The CT radiation output parameters analysed were volumetric CT dose index (CTDIvol) and dose-length product (DLP). Three radiologists evaluated subjective image quality using a three-point Likert scale.
RESULTS: The study included 94,258 CT series (from 27,103 CT examinations) in adult patients performed in 2016. When averaged over all body regions with available DRL, institutional CTDIvol/DLP values were always below the DRLs (65.2±32.9%/67.3±41.5% initially; 59.4±32%/60.5±39.9% after optimisation). Values exceeding the national DRLs were found for pelvis (n=268; CTDIvol 107.7±65.7%/DLP 106.3±79.3%), lumbar spine (n=91; 160.8±74.7%/175.2±104.1%), and facial bones (n=527; 108±39%/152.7±75.7%). After optimisation, CTDIvol and DLP were 87.9±73%/87.8±80.8% for the pelvis, 67.8±33.2%/74.5±50.6% for the lumbar spine and 95.1±45.8%/133.3±74.6% for the viscerocranium.
CONCLUSION: An automatic CT dose monitoring system enabled not only comprehensive monitoring of a DRL implementation process but can also help to optimise radiation exposure.
Copyright © 2018 The Royal College of Radiologists. Published by Elsevier Ltd. All rights reserved.

Entities:  

Mesh:

Year:  2018        PMID: 29567269     DOI: 10.1016/j.crad.2018.02.012

Source DB:  PubMed          Journal:  Clin Radiol        ISSN: 0009-9260            Impact factor:   2.350


  3 in total

1.  Local clinical diagnostic reference levels for chest and abdomen CT examinations in adults as a function of body mass index and clinical indication: a prospective multicenter study.

Authors:  Hugues Brat; Federica Zanca; Stéphane Montandon; Damien Racine; Benoit Rizk; Eric Meicher; Dominique Fournier
Journal:  Eur Radiol       Date:  2019-05-29       Impact factor: 5.315

2.  Potential of a machine-learning model for dose optimization in CT quality assurance.

Authors:  Axel Meineke; Christian Rubbert; Lino M Sawicki; Christoph Thomas; Yan Klosterkemper; Elisabeth Appel; Julian Caspers; Oliver T Bethge; Patric Kröpil; Gerald Antoch; Johannes Boos
Journal:  Eur Radiol       Date:  2019-02-19       Impact factor: 5.315

3.  Utilization of artificial intelligence approach for prediction of DLP values for abdominal CT scans: A high accuracy estimation for risk assessment.

Authors:  H O Tekin; Faisal Almisned; T T Erguzel; Mohamed M Abuzaid; W Elshami; Antoaneta Ene; Shams A M Issa; Hesham M H Zakaly
Journal:  Front Public Health       Date:  2022-07-28
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.