Literature DB >> 29557863

Effect of Postoperative Mechanical Axis Alignment on Survival and Functional Outcomes of Modern Total Knee Arthroplasties with Cement: A Concise Follow-up at 20 Years.

Matthew P Abdel1, Matthieu Ollivier1, Sebastien Parratte1, Robert T Trousdale1, Daniel J Berry1, Mark W Pagnano1.   

Abstract

We previously compared the 15-year survivorship of total knee arthroplasty (TKA) implants that were mechanically aligned (0° ± 3° relative to the mechanical axis) compared with those that were outside that range and considered outliers. The original publication included 398 TKAs (292 in the aligned group and 106 in the outlier group) performed from 1985 to 1990. At the time of follow-up in the previous study, 138 patients (155 TKAs) had died and 59 knees had been revised. Since that publication, 49 additional patients (87 knees) have died. At 20 years, 57 (19.5%) of the 292 knees in the mechanically aligned group had been revised compared with 16 (15.1%) of the 106 knees in the outlier group (p = 0.97). Postoperative alignment within 0° ± 3° of the mechanical axis did not provide a functional advantage at 1, 5, 10, 15, and/or 20 years postoperatively as demonstrated by the Knee Society scores being similar between the groups (p ≥ 0.2 at all intervals). At 20 years, we once again did not find that neutral mechanical alignment provided better implant survivorship than that found in the outlier group. LEVEL OF EVIDENCE: Therapeutic Level III. See Instructions for Authors for a complete description of levels of evidence.

Entities:  

Mesh:

Substances:

Year:  2018        PMID: 29557863     DOI: 10.2106/JBJS.16.01587

Source DB:  PubMed          Journal:  J Bone Joint Surg Am        ISSN: 0021-9355            Impact factor:   5.284


  18 in total

1.  Linked Anatomic Kinematic Arthroplasty: A Unique Approach to Balancing in Total Knee Arthroplasty.

Authors:  Seth Stake; Safa C Fassihi; Alex Gu; James Burns; George Thomas; Evan H Argintar; Kamran Sadr
Journal:  J Orthop       Date:  2020-01-14

2.  Preliminary experience with an image-free handheld robot for total knee arthroplasty: 77 cases compared with a matched control group.

Authors:  P Bollars; A Boeckxstaens; J Mievis; S Kalaai; M G M Schotanus; D Janssen
Journal:  Eur J Orthop Surg Traumatol       Date:  2020-01-16

3.  Linked kinematic knee balancing in unicompartmental knee arthroplasty.

Authors:  Ryan Mortman; Casey Gioia; Seth Stake; Peter Z Berger; Alex Gu; Safa C Fassihi; Kamran N Sadr; Evan H Argintar
Journal:  J Orthop       Date:  2021-02-20

4.  CORR Insights®: Does Robotic-assisted TKA Result in Better Outcome Scores or Long-Term Survivorship Than Conventional TKA? A Randomized, Controlled Trial.

Authors:  Lawrence D Dorr
Journal:  Clin Orthop Relat Res       Date:  2020-02       Impact factor: 4.755

5.  Custom Implants in TKA Provide No Substantial Benefit in Terms of Outcome Scores, Reoperation Risk, or Mean Alignment: A Systematic Review.

Authors:  Eran Beit Ner; Saad Dosani; Leela C Biant; Gwenllian Fflur Tawy
Journal:  Clin Orthop Relat Res       Date:  2021-06-01       Impact factor: 4.755

6.  Trends in computer navigation and robotic assistance for total knee arthroplasty in the United States: an analysis of patient and hospital factors.

Authors:  Joseph K Antonios; Shane Korber; Lakshmanan Sivasundaram; Cory Mayfield; Hyunwoo Paco Kang; Daniel A Oakes; Nathanael D Heckmann
Journal:  Arthroplast Today       Date:  2019-03-12

7.  A Retrospective Study Comparing a Patient-specific Design Total Knee Arthroplasty With an Off-the-Shelf Design: Unexpected Catastrophic Failure Seen in the Early Patient-specific Design.

Authors:  Carlos J Meheux; Kwan J Park; Terry A Clyburn
Journal:  J Am Acad Orthop Surg Glob Res Rev       Date:  2019-11-04

8.  Factors associated with pain intensity and magnitude of limitations among people with hip and knee arthritis.

Authors:  Benjamin Kopp; Kenneth Furlough; Tyler Goldberg; David Ring; Karl Koenig
Journal:  J Orthop       Date:  2021-05-21

9.  Influence of surgical factors on patient satisfaction after bi-cruciate stabilized total knee arthroplasty: retrospective examination using multiple regression analysis.

Authors:  Hiroshi Inui; Shuji Taketomi; Ryota Yamagami; Kenichi Kono; Kohei Kawaguchi; Kosuke Uehara; Sakae Tanaka
Journal:  BMC Musculoskelet Disord       Date:  2021-02-23       Impact factor: 2.362

Review 10.  Comparison of Kinematic Alignment and Mechanical Alignment in Total Knee Arthroplasty: A Meta-analysis of Randomized Controlled Clinical Trials.

Authors:  Zhi-Xiang Gao; Neng-Ji Long; Shao-Yun Zhang; Wei Yu; Yi-Xin Dai; Cong Xiao
Journal:  Orthop Surg       Date:  2020-10-25       Impact factor: 2.071