In vitro methods for studying the accuracy of velocity determination and spatial resolution of a color Doppler flow mapping system.

T Tamura, A Yoganathan, D J Sahn.   

Abstract

We tested a color Doppler flow mapping system (Aloka 880, 3.5 mHz) to image the steady flow of a suspension of cornstarch (0.1% to 1.75%) in water, flowing at known velocities through a series of Tygon tubing with internal diameters of 1.6 to 4.8 mm and latex rubber tubing (7.9 to 15.9 mm) in a water tank. Flow through the tubes was imaged at 2 to 12 cm and 2 to 18 cm distances from the transducer, by the use of the near and far focus settings of the instrument. First, the color flow diameters were compared to the known internal diameters of the tubing oriented at a 30- to 70-degree flow angle to the direction of interrogation to test axial resolution as well as lateral resolution. For the near transmit focus setting, the color diameters increasingly overestimated the actual flow diameters as the distance from the transducer increased progressively beyond 6 cm (i.e., overestimation of up to 25 mm or 500% for the 4.8 mm tubing and 250% for the 9.5 mm tubing at 16 cm from the transducer). For the far transmit focus setting and for distances within 10 cm, the color diameters did not differ significantly (i.e., more than 1 mm) from the actual diameter. The lowest velocity color encoded by the instrument was 3 cm/sec for a pulse repetition frequency of 4 kHz. At a flow velocity of about 50 cm/sec, 6 cm distance from the transducer and mid gain settings, flow through the 1.6 mm tube barely color coded, while flow through 2.4, 3.2, and 4.8 mm tubes was easily imaged.(ABSTRACT TRUNCATED AT 250 WORDS)

Entities:  

Mesh:

Year:  1987        PMID: 2955684     DOI: 10.1016/0002-8703(87)90319-x

Source DB:  PubMed          Journal:  Am Heart J        ISSN: 0002-8703            Impact factor:   4.749


  2 in total

Review 1.  [Perioperative echocardiography: basic principles].

Authors:  M Nowak; P Rosenberger; T W Felbinger; A E Götz; S K Shernan; K Unertl; H K Eltzschig
Journal:  Anaesthesist       Date:  2006-03       Impact factor: 1.041

2.  Velocity measurements with a new ultrasonic Doppler method independent of angle of incidence.

Authors:  Shigeru Akamatsu; Yuji Kondo; Shuji Dohi
Journal:  J Anesth       Date:  1996-06       Impact factor: 2.078

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.