| Literature DB >> 29554960 |
Sarah Hindenberg1, Melanie Keßler2, Sabine Zielinsky2, Judith Langenstein3, Andreas Moritz2, Natali Bauer2.
Abstract
BACKGROUND: Species-specific point-of-care tests (POCT) permit a rapid analysis of canine C-reactive protein (CRP), enabling veterinarians to include CRP in clinical decisions. Aim of the study was to evaluate a novel POCT for canine CRP (Point Strip™ Canine CRP Assay) run on a small in-house-analyzer (Point Reader™ V) using lithium heparin plasma and to compare assay performance to an already established canine CRP assay (Gentian Canine CRP Immunoassay) run on two different bench top analyzers serving as reference methods (ABX Pentra 400, AU 5800). Linearity was assessed by stepwise dilution of plasma samples with high CRP concentrations. Limit of quantification (LoQ) was determined by repeated measurements of samples with low CRP concentrations. Coefficient of variation (CV) at low (10-50 mg/l), moderate (50-100 mg/l), and high (100-200 mg/l) CRP concentrations was investigated as well as possible interferences. Method comparison study was performed using 45 samples of healthy and diseased dogs. Quality criteria were fulfilled if the total observed error (TEobs = 2CV% + bias%) was below the minimal total allowable error of 44.4% (TE min). Additionally, a reference range (n = 60 healthy dogs) was established.Entities:
Keywords: Acute phase protein; Bias; C-reactive protein; Canine; Heparin plasma; Interference; Method validation; Point-of-care analyzer; Point-of-care assay; Total allowable error
Mesh:
Substances:
Year: 2018 PMID: 29554960 PMCID: PMC5859731 DOI: 10.1186/s12917-018-1415-2
Source DB: PubMed Journal: BMC Vet Res ISSN: 1746-6148 Impact factor: 2.741
Quality specifications
| Acceptance limits | Quality parameters | CV (%) | Bias (%) | TE (%) |
|---|---|---|---|---|
| - minimally acceptable | 18.2 | 14.3 | 44.4 | |
| - desired | 12.2 | 9.5 | 29.6 | |
| - optimal | 6.1 | 4.8 | 14.8 | |
Quality specifications are derived from biological variation as published in the addendum of the Total allowable error guidelines of the American Society of Veterinary Clinical Pathology [32]
Abbreviations: CV coefficient of variation, TE total allowable error
Fig. 1Linearity of two diluted lithium heparin plasma samples with high CRP concentrations. a Linearity under dilution for measurement of a canine serum sample originally containing 244.8 mg/l CRP as assessed with the Pentra 400 bench top analyzer. b Linearity under dilution for measurement of a canine serum sample originally containing 361.3 mg/l CRP as assessed with the Pentra 400 bench top analyzer. A serial dilution was performed in both cases to achieve 8 different CRP concentrations, i.e, 1.0, 0.8, 0.6, 0.4, 0.2, 0.1, 0.05, 0.025 parts of the original concentration
Linearity and recovery
| Dilution Factor | Expected concentration [mg/l] | Mean measured concentration [mg/l] | Recovery [%] | Biasobs [%] | %Biasobs < TEdes (29.6%) | %Biasobs < TEmin (44.4%) |
|---|---|---|---|---|---|---|
| A | ||||||
| 0.025 | 6.1 | 6.5 | 106.8 | 6.8 | Yes | Yes |
| 0.05 | 12.2 | 11.9 | 97.0 | 3.1 | Yes | Yes |
| 0.1 | 24.5 | 19.8 | 81.0 | 19.0 | Yes | Yes |
| 0.2 | 49.0 | 36.5 |
| 25.5 | Yes | Yes |
| 0.4 | 97.9 | 59.8 |
|
|
| Yes |
| 0.6 | 146.9 | 79.8 |
|
|
|
|
| 0.8 | 195.8 | 88.0 |
|
|
|
|
| 1 | 244.8 | 122.3 |
|
|
|
|
| B | ||||||
| 0.025 | 9.0 | 6.2 |
|
|
|
|
| 0.05 | 18.1 | 12.3 |
|
|
|
|
| 0.1 | 36.1 | 20.4 |
|
|
|
|
| 0.2 | 72.2 | 36.7 |
|
|
|
|
| 0.4 | 144.4 | 72.5 |
|
|
|
|
| 0.6 | 216.6 | 85.8 |
|
|
|
|
| 0.8 | 288.8 | 112.6 |
|
|
|
|
| 1 | 361.3 | 132.4 |
|
|
|
|
Linearity and recovery rates of CRP measurements performed with a canine serum sample containing A) 244.8 mg/l or B) 361.3 mg/l CRP respectively as determined on the ABX Pentra 400 analyzer
Quality parameters exceeding the quality criteria reported previously [32, 36], are marked in bold letters
Abbreviations: Bias observed bias, TE total allowable error, TE desirable TE, TE minimally acceptable TE
Intra- and inter-assay CVs obtained from replicate CRP measurements with the POCT
| Three CRP concentration ranges / analyzer | Doga | Intra-assay CV | Inter-assay CV | ||||
|---|---|---|---|---|---|---|---|
| Mean | SD | CV | Mean | SD | CV | ||
| ≥ 10 - < 50 mg/l | 1 | 21.3 | 4.7 |
| 23.2 | 6.3 |
|
| 2 | 43.1 | 2.3 | 5.4 | 32.3 | 2.0 | 6.2 | |
| 3 | 23.1 | 3.7 |
| 15.4 | 0.8 | 5.0 | |
| 4 | 29.9 | 4.0 |
| 32.2 | 2.8 | 8.8 | |
| ≥ 50 - < 100 mg/l | 5 | 55.7 | 3.0 | 5.5 | 71.0 | 7.6 | 10.7 |
| 6 | 79.8 | 4.7 | 5.9 | 74.4 | 5.5 | 7.3 | |
| ≥ 100 - < 200 mg/l | 7 | 113.9 | 7.4 | 6.5 | 102.4 | 7.2 | 7.0 |
| 8 | 113.9 | 9.1 | 8.0 | 110.5 | 11.1 | 10.0 | |
aSamples from different dogs were used for assessment of intra- and inter- assay CV
CV > CVdes (i.e., 12.2%) and CV > CVmin (i.e., 18.2%) is shown in bold letters
Abbreviations: SD standard deviation, CV coefficient of variation, CV desirable CV, CV minimally acceptable CV
Determination of the lower limit of quantification
| Three low CRP concentrations | Dog | Intra-assay CV | |||
|---|---|---|---|---|---|
| Mean | SD | CV | % CV < CVdes | ||
| < 1 mg/l | 1 | < 5* | n.d. | n.d. | n.d. |
| ~ 10 mg/L | 2 | 11.5 | 0.7 | 6.2 | Yes |
| ~ 20 mg/l | 3 | 16.8 | 1.8 | 11.0 | Yes |
Intra-assay CVs for the CRP analysis obtained from replicate measurements at low CRP values with the POCT
*values < 5 mg/l are generally reported as “< 5 mg/l”, therefore no further calculations were possible
Abbreviations: SD standard deviation, CV coefficient of variation, CV desirable CV, n.d. not done
Observed interference effects of bilirubin, hemoglobin, lipid on triplicate CRP measurement with the POCT
| Interferent | Mean CRPcontrol [mg/l] ± SD | Mean CRPtest [mg/l] ± SD | Mean bias [mg/l] | %Biasobs | %Biasobs < TEdes (29.6%) | %Biasobs < TEopt (14.8%) |
|---|---|---|---|---|---|---|
| Bilirubin | 37.6 ± 5.6 | 39.8 ± 0.4 | 2.2 | 5.8 | Yes | Yes |
| Hemoglobin | 29.5 ± 1.7 | 30.1 ± 4.1 | 0.6 | 2.0 | Yes | Yes |
| Soy bean emulsion | 35.1 ± 1.4 | 39.1 ± 0.8 | 4.0 | 11.3 | Yes | Yes |
Test samples (CRPtest) spiked with the interfering substances were compared to control samples (CRPcontrol) spiked with equal volume of the diluent used in the test sample. %Biasobs (observed bias) for the interfering substance was acceptable if %Biasobs < desired total allowable error (TEdes) and excellent if %Biasobs < optimal total allowable error (TEopt) [32]
Results of method comparison: Intra-assay precision
| Intra-assay CV median (range) at three CRP concentration ranges / analyzer | Point Reader™ V | ABX Pentra 400 | AU 5800 |
|---|---|---|---|
| ≥ 10 - < 50 mg/l | 3.53% | 0.62% | 0.76% |
| ≥ 50 - < 100 mg/l | 3.50% | 0.43% | 1.68% |
| ≥ 100 - < 200 mg/l | 5.62% | 0,30% | 1.50% |
Intra-assay CVs for the CRP analysis obtained from duplicate measurements performed on two automated bench top analyzers (ABX Pentra, AU 5800) and the point-of-care analyzer
Abbreviations: CV coefficient of variation
Fig. 2Results of method comparison between the POCT and two automated analyzers. The point-of-care analyzer was compared to a CRP assay run on two automated large bench top analyzers (ABX Pentra 400, AU 5800). A1–3: Passing- Bablok regression analysis with 95% confidence interval (CI) of CRP measurements performed with the three analyzers. A1: 95% CI of slope and intercept of the regression line were 0.74 to 0.90 and − 4.52 to 4.10, respectively. A2: 95% CI of slope and intercept were 0.86 to 0.95 and − 2.15 to 2.25, respectively. A3: There was a 95% CI of slope and intercept of 0.85 to 0.95 and − 2.75 to 2.02, respectively. B1–3: Bland-Altman difference plot demonstrating mean %Biasobs with its 95% confidence interval and its 1.96fold standard deviation (SD) indicative of its limits of agreement. The minimally acceptable total allowable error (TEmin) of 44.4% [32] represents the quality limit. C1–3: Bland-Altman difference plot demonstrating the absolute bias between CRP measurements obtained with three analyzers. For remainder of key, refer to Fig. 1b
Fig. 3Results of method comparison: Differences of medians and ranges between the analyzers. Box- and - whisker diagram demonstrating median and range of the CRP measurements obtained with the three analyzers. The horizontal line in the boxes is consistent with the median, the whiskers indicate the range and the box represents the 25th -75th percentile
Results of method comparison: Total observed error
| TEobs at three CRP concentration ranges / analyzer | ABX Pentra 400 vs POCT | AU 5800 vs POCT | ABX Pentra 400 vs AU 5800 |
|---|---|---|---|
| ≥ 10 - < 50 mg/l | 27.3% | 17.3% | 10.7% |
| ≥ 50 - < 100 mg/l | 26.7% | 16.7% | 12.6% |
| ≥ 100 - < 200 mg/l |
| 21.9% | 12.2% |
Total observed error (TEobs) for the CRP analysis performed on two automated bench top analyzers (ABX Pentra 400, AU 5800) and the point-of-care analyzer TEobs > TEdes (i.e., 29.6%) but < TEmin (i.e., 44.4%) is shown in bold letters
Abbreviations: vs versus, TE desirable total allowable error, TE minimally acceptable total error