Improved spatial localization in magnetic resonance spectroscopic imaging with two-dimensional PSF-Choice encoding.

Shelley HuaLei Zhang1, Stephan E Maier2, Lawrence P Panych3.   

Abstract

PURPOSE: Magnetic resonance spectroscopic imaging (MRSI), under low-spatial resolution settings, often suffers signal contamination from neighboring voxels due to ringing artifacts. Spatial localization can be improved by constraining the point-spread-function (PSF). Here the effectiveness of the two-dimensional PSF-Choice technique in providing improved spatial localization for MRSI is demonstrated. THEORY AND METHODS: The PSF-Choice technique constrains the PSF to a desired shape by manipulating the weighting of RF excitation pulse throughout phase-encode steps. Based on a Point REsolved SpectroScopy (PRESS)-type sequence, PSF-Choice encoding was implemented along two dimensions to excite a two-dimensional Gaussian profile, by replacing the usual RF excitation pulse with a train of pulses that is modified at each phase-encoding step. The method was proven mathematically, and demonstrated experimentally in phantoms containing prostate relevant metabolic compounds of choline, creatine and citrate.
RESULTS: Using a dedicated prostate-mimicking spectroscopy phantom surrounded by oil, it was found that there is significantly less signal contamination from oil for PSF-Choice encoding compared with standard phase encoding. In particular, with standard phase encoding, there was a significant difference (p = 0.014) between ratios of Choline + Creatine to Citrate for voxels well within the phantom compared to voxels within the phantom but near the boundary with oil. The ratios in boundary voxels were also significantly different (p = 0.035) from reference values obtained using the prostate phantom with no oil present. In contrast, no significant differences were found in comparisons of these ratios when encoding with PSF-Choice.
CONCLUSION: The PSF-Choice scheme applied along two dimensions produces MR spectroscopic images with substantially reduced truncation artifacts and spectral contamination.
Copyright © 2018 Elsevier Inc. All rights reserved.

Entities:  

Keywords:  Fourier encoding; Gibbs ringing; PSF-Choice encoding; Point spread function; Prostate; RF encoding; Spectroscopic imaging; Voxel bleeding

Mesh:

Substances:

Year:  2018        PMID: 29547794      PMCID: PMC5895504          DOI: 10.1016/j.jmr.2018.03.002

Source DB:  PubMed          Journal:  J Magn Reson        ISSN: 1090-7807            Impact factor:   2.229


  15 in total

1.  Fast acquisition-weighted three-dimensional proton MR spectroscopic imaging of the human prostate.

Authors:  Tom W J Scheenen; Dennis W J Klomp; Stefan A Röll; Jurgen J Fütterer; Jelle O Barentsz; Arend Heerschap
Journal:  Magn Reson Med       Date:  2004-07       Impact factor: 4.668

2.  Investigation of the PSF-choice method for reduced lipid contamination in prostate MR spectroscopic imaging.

Authors:  Lawrence P Panych; Joseph R Roebuck; Nan-kuei Chen; Yi Tang; Bruno Madore; Clare M Tempany; Robert V Mulkern
Journal:  Magn Reson Med       Date:  2012-11       Impact factor: 4.668

3.  Gibbs-ringing artifact removal based on local subvoxel-shifts.

Authors:  Elias Kellner; Bibek Dhital; Valerij G Kiselev; Marco Reisert
Journal:  Magn Reson Med       Date:  2015-11-24       Impact factor: 4.668

4.  A fast spectroscopic imaging method using a blipped phase encode gradient.

Authors:  P Webb; D Spielman; A Macovski
Journal:  Magn Reson Med       Date:  1989-12       Impact factor: 4.668

5.  PSF-choice: a novel MRI method for shaping point-spread functions in phase-encoding dimensions.

Authors:  Lawrence P Panych; Lei Zhao; Robert V Mulkern
Journal:  Magn Reson Med       Date:  2005-07       Impact factor: 4.668

Review 6.  MR spectroscopic imaging: principles and recent advances.

Authors:  Stefan Posse; Ricardo Otazo; Stephen R Dager; Jeffry Alger
Journal:  J Magn Reson Imaging       Date:  2012-11-27       Impact factor: 4.813

7.  Selective Fourier transform localization.

Authors:  H R Brooker; T H Mareci; J T Mao
Journal:  Magn Reson Med       Date:  1987-11       Impact factor: 4.668

8.  High speed 1H spectroscopic imaging in human brain by echo planar spatial-spectral encoding.

Authors:  S Posse; G Tedeschi; R Risinger; R Ogg; D Le Bihan
Journal:  Magn Reson Med       Date:  1995-01       Impact factor: 4.668

9.  Prostate cancer metabolite quantification relative to water in 1H-MRSI in vivo at 3 Tesla.

Authors:  Mary A McLean; Tristan Barrett; Vincent J Gnanapragasam; Andrew N Priest; Ilse Joubert; David J Lomas; David E Neal; John R Griffiths; Evis Sala
Journal:  Magn Reson Med       Date:  2010-11-16       Impact factor: 4.668

10.  Prostate cancer aggressiveness: in vivo assessment of MR spectroscopy and diffusion-weighted imaging at 3 T.

Authors:  Thiele Kobus; Pieter C Vos; Thomas Hambrock; Maarten De Rooij; Christina A Hulsbergen-Van de Kaa; Jelle O Barentsz; Arend Heerschap; Tom W J Scheenen
Journal:  Radiology       Date:  2012-07-27       Impact factor: 11.105

View more
  2 in total

1.  Fast, regional three-dimensional hybrid (1D-Hadamard 2D-rosette) proton MR spectroscopic imaging in the human temporal lobes.

Authors:  Assaf Tal; Tiejun Zhao; Claudiu Schirda; Hoby P Hetherington; Jullie W Pan; Oded Gonen
Journal:  NMR Biomed       Date:  2021-03-23       Impact factor: 4.478

2.  Advanced magnetic resonance spectroscopic neuroimaging: Experts' consensus recommendations.

Authors:  Andrew A Maudsley; Ovidiu C Andronesi; Peter B Barker; Alberto Bizzi; Wolfgang Bogner; Anke Henning; Sarah J Nelson; Stefan Posse; Dikoma C Shungu; Brian J Soher
Journal:  NMR Biomed       Date:  2020-04-29       Impact factor: 4.044

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.