Notes from the Field: False-Negative Hepatitis B Surface Antigen Test Results in a Hemodialysis Patient - Nebraska, 2017.

Blake Hendrickson, Saleem Kamili, Tim Timmons, Peter C Iwen, Caitlin Pedati, Thomas Safranek.   

Abstract

Entities:  

Mesh:

Substances:

Year:  2018        PMID: 29543792      PMCID: PMC5857195          DOI: 10.15585/mmwr.mm6710a6

Source DB:  PubMed          Journal:  MMWR Morb Mortal Wkly Rep        ISSN: 0149-2195            Impact factor:   17.586


× No keyword cloud information.
In March 2017, the Nebraska Department of Health and Human Services (NDHHS) was contacted by a hemodialysis clinic regarding a patient who had tested negative for hepatitis B virus (HBV) surface antigen (HBsAg) after vaccination in 2010 and who later tested positive for HBsAg. A public health investigation subsequently determined that the false-negative results were caused by a surface antigen mutation. Notably, several commercial HBsAg testing kits cannot detect this mutant virus, making it a challenging pathogen for public health surveillance and intervention efforts (). When the patient started dialysis in 2010, there was no evidence that the patient had ever been tested for HBV. The patient’s vaccination status was also unknown, and 4 doses of HBV vaccine were administered at 0, 1, 2, and 6 months, as recommended for hemodialysis patients (). Postvaccination testing in 2010 was positive for hepatitis B surface antibody (anti-HBs = 82 mIU/mL), indicating immunity to HBV. Postvaccination HBsAg testing performed with a Food and Drug Administration (FDA)–approved assay at Laboratory A was negative. As recommended for patients with demonstrated immunity, only anti-HBs was monitored routinely from 2010 to 2016 to ensure continued protection (anti-HBs >10 mIU/mL) (). In 2016, the patient was hospitalized for acute shortness of breath, and an HBsAg test performed as part of a routine evaluation at laboratory B was positive, indicating a current infection with HBV. The positive result was confirmed by additional testing at commercial and public health laboratories (Table). Insufficient data were available to determine whether the patient acquired HBV infection before or after vaccination. Specific precautions for HBV-positive patients in a dialysis center include receiving dialysis in a separate room and being assigned separate staff members and equipment (). These control measures were not implemented during 2010–2016 because the index patient had not had a positive HBsAg test result. After the positive HBsAg results were reported, an epidemiologic investigation was initiated by NDHHS and the Lincoln-Lancaster County Health Department to determine the cause of the false-negative test results and to identify any HBV transmission. A blood sample was collected in April 2017 and sent to the Division of Viral Hepatitis Laboratory Branch at CDC to confirm the HBV diagnosis. CDC testing found high HBV DNA levels (14,200,000 IU/mL), evidence of immunity (anti-HBs = 114 mIU/mL), and HBsAg positivity by one assay and negativity by another assay (Table). Sequencing the S gene of HBV DNA identified an sG145R surface antigen mutation. This mutation is associated with false-negative results, which explains the failure of multiple tests to identify the patient as being HBsAg-positive (). The epidemiologic investigation identified 45 recent dialysis contacts and 10 close family contacts who were at risk for infection. None of the contacts had prior evidence of HBV infection, all were screened by tests capable of detecting this mutant virus (using a suitable HBsAg assay or by HBV DNA), and all test results were negative, indicating no evidence of HBV transmission, despite the potential exposures to the HBV-infected patient. Family members without evidence of prior HBV vaccination were also advised to complete the HBV vaccination series. A subsequent survey of laboratories that reported HBsAg results to NDHHS in the previous year identified nine of 23 laboratories using tests that are not known to detect common HBsAg mutations. This included both local hospitals and large national reference laboratories. The prevalence of sG145R mutations and other HBsAg mutants associated with false-negative test results is not known. However, some studies suggest that mutant strains are found in 6%–12% of chronic HBV carriers (). In addition to issues with diagnostic detection, some mutants are also not recognized and neutralized by protective antibodies induced by current HBV vaccines and HBV immune globulin therapy (). This case highlights a unique challenge associated with detecting HBV infections when a surface antigen mutation is present. In recent years, some manufacturers have adapted their testing assays to better identify sG145R and other HBsAg mutations. It is important that laboratories use FDA-approved assays that have the ability to detect HBsAg mutants, which was not done in this case at Laboratory A. The clinical and public health community also must be aware of these testing limitations so that discordant results can be identified for correct diagnosis and care of HBV-infected persons and to minimize the spread of these mutant viruses. Abbreviation: NPHL = Nebraska Public Health Lab. * Deidentified commercial laboratory. † False-negative result.
TABLE

HBsAg lab results for the case patient by facility and testing instrument

Date collected Laboratory facilityTesting instrumentResult
November 11, 2010A*Advia Centaur XPTNegative
December 8, 2016A*Advia Centaur XPTNegative
December 9, 2016B*Advia Centaur XPPositive
December 14, 2016C*Advia Centaur XPPositive
January 5, 2017A*Advia Centaur XPTNegative
February 2, 2017A*Advia Centaur XPTNegative
March 2, 2017A*Advia Centaur XPTNegative
March 2, 2017C*Advia Centaur XPPositive
May 7, 2017CDCVitros EciNegative
May 7, 2017CDCAbbott ARCHITECTPositive
May 23, 2017NPHLAdvia Centaur XPPositive
July 25, 2017D*ETI-MAK-2 PLUSPositive
July 25, 2017E*Vitros 3600Negative

Abbreviation: NPHL = Nebraska Public Health Lab.

* Deidentified commercial laboratory.

† False-negative result.

  3 in total

1.  Variable capacity of 13 hepatitis B virus surface antigen assays for the detection of HBsAg mutants in blood samples.

Authors:  Annabelle Servant-Delmas; Mélanie Mercier-Darty; Thoai Duong Ly; Françoise Wind; Chakib Alloui; Camille Sureau; Syria Laperche
Journal:  J Clin Virol       Date:  2012-01-31       Impact factor: 3.168

2.  Hepatitis B virus genetic diversity.

Authors:  José M Echevarría; Ana Avellón
Journal:  J Med Virol       Date:  2006       Impact factor: 2.327

3.  Recommendations for preventing transmission of infections among chronic hemodialysis patients.

Authors: 
Journal:  MMWR Recomm Rep       Date:  2001-04-27
  3 in total
  4 in total

1.  An automated clinical mass spectrometric method for identification and quantification of variant and wild-type amyloid-β 1-40 and 1-42 peptides in CSF.

Authors:  Mari L DeMarco; Quyen Nguyen; Alice Fok; Ging-Yuek Robin Hsiung; J Grace van der Gugten
Journal:  Alzheimers Dement (Amst)       Date:  2020-06-30

2.  Molecular and serological characterization of hepatitis B vaccine breakthrough infections in serial samples from two plasma donors.

Authors:  Mary C Kuhns; Anne L McNamara; Vera Holzmayer; Gavin A Cloherty
Journal:  Virol J       Date:  2019-04-03       Impact factor: 4.099

3.  Hepatitis B Virus Mutant Infections in Hemodialysis Patients: A Case Series.

Authors:  Ibironke W Apata; Duc B Nguyen; Yury Khudyakov; Tonya Mixson-Hayden; Jon Rosenberg; Matt Zahn; Jane Greenko; Ernest Clement; Allison E Portney; Prathit A Kulkarni; Maura Comer; Eleanor Adams; Saleem Kamili; Priti R Patel; Anne C Moorman
Journal:  Kidney Med       Date:  2019-10-01

Review 4.  Proteoforms and their expanding role in laboratory medicine.

Authors:  Lauren M Forgrave; Meng Wang; David Yang; Mari L DeMarco
Journal:  Pract Lab Med       Date:  2021-11-27
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.