Application of low-tube current with iterative model reconstruction on Philips Brilliance iCT Elite FHD in the accuracy of spinal QCT using a European spine phantom.

Yan Wu1, Yaojun Jiang1, Xueli Han1, Mingyue Wang1, Jianbo Gao1.   

Abstract

BACKGROUND: To investigate the repeatability and accuracy of quantitative CT (QCT) measurement of bone mineral density (BMD) by low-mAs using iterative model reconstruction (IMR) technique based on phantom model.
METHODS: European spine phantom (ESP) was selected and measured on the Philips Brilliance iCT Elite FHD machine for 10 times. Data were transmitted to the QCT PRO workstation to measure BMD (mg/cm3) of the ESP (L1, L2, L3). Scanning method: the voltage of X-ray tube is 120 kV, the electric current of X-ray tube output in five respective groups A-E were: 20, 30, 40, 50 and 60 mAs. Reconstruction: all data were reconstructed using filtered back projection (FBP), IR levels of hybrid iterative reconstruction (iDose4, levels 1, 2, 3, 4, 5, 6 were used) and IMR (levels 1, 2, 3 were used). ROIs were placed in the middle of L1, L2 and L3 spine phantom in each group. CT values, noise and contrast-to-noise ratio (CNR) were measured and calculated. One-way analysis of variance (ANOVA) was used to compare BMD values of different mAs and different IMR.
RESULTS: Radiation dose [volume CT dose index (CTDIvol) and dose length product (DLP)] was positively correlated with tube current. In L1 with low BMD, different mAs in FBP showed P<0.05, indicating statistically significant BMD in ESP. In other iterative algorithms, different mAs under same iterative algorithms showed P>0.05, indicating no difference in BMD. And P>0.05 was observed among BMD of spine phantom in L1, L2 and L3 under same mAs joined with varied iterative reconstruction. The BMD in L1 varied greatly during FBP reconstruction, and less variation was observed in reconstruction of IMR [1] and IMR [2]. The BMD of L2 changed more during FBP reconstruction, where less was observed in IMR [2]. The BMD of L3 varied greatly during FBP reconstruction, and was less varied in all levels of iDose4 and reconstruction of IMR [2]. In addition, along with continuous mAs incensement, the CNRs in various algorithms continued to increase. Among them, CNR with the FBP algorithm is the lowest, and CNR of the IMR [3] algorithm is the highest.
CONCLUSIONS: Repeated measurements of BMD with QCT in the ESP multicenter showed that BMD changes in L1-L3 are the least varied at IMR [2] algorithm. It is recommended to scan at 120 kV with 20 mAs combined with IMR [2] algorithm. In this way, the BMD of spine by QCT could be accurately measured, while radiation dosage significantly reduced and imaging quality improved at the same time.

Entities:  

Keywords:  European spine phantom (ESP); Quantitative CT (QCT); X-ray computer; bone mineral density (BMD); iterative model reconstruction (IMR); radiation dosage

Year:  2018        PMID: 29541621      PMCID: PMC5835662          DOI: 10.21037/qims.2018.02.03

Source DB:  PubMed          Journal:  Quant Imaging Med Surg        ISSN: 2223-4306


  10 in total

Review 1.  Current methods and advances in bone densitometry.

Authors:  G Guglielmi; C C Gluer; S Majumdar; B A Blunt; H K Genant
Journal:  Eur Radiol       Date:  1995       Impact factor: 5.315

Review 2.  Quantitative computed tomography.

Authors:  Judith E Adams
Journal:  Eur J Radiol       Date:  2009-08-13       Impact factor: 3.528

3.  Clinical use of quantitative computed tomography and peripheral quantitative computed tomography in the management of osteoporosis in adults: the 2007 ISCD Official Positions.

Authors:  Klaus Engelke; Judith E Adams; Gabriele Armbrecht; Peter Augat; Cesar E Bogado; Mary L Bouxsein; Dieter Felsenberg; Masako Ito; Sven Prevrhal; Didier B Hans; E Michael Lewiecki
Journal:  J Clin Densitom       Date:  2008 Jan-Mar       Impact factor: 2.617

4.  Combination of a low-tube-voltage technique with hybrid iterative reconstruction (iDose) algorithm at coronary computed tomographic angiography.

Authors:  Yoshinori Funama; Katsuyuki Taguchi; Daisuke Utsunomiya; Seitaro Oda; Yumi Yanaga; Yasuyuki Yamashita; Kazuo Awai
Journal:  J Comput Assist Tomogr       Date:  2011 Jul-Aug       Impact factor: 1.826

5.  The European Spine Phantom--a tool for standardization and quality control in spinal bone mineral measurements by DXA and QCT.

Authors:  W A Kalender; D Felsenberg; H K Genant; M Fischer; J Dequeker; J Reeve
Journal:  Eur J Radiol       Date:  1995-07       Impact factor: 3.528

Review 6.  Quantitative computed tomography (QCT) of the lumbar spine and appendicular skeleton.

Authors:  C C Glüer; K Engelke; T F Lang; S Grampp; H K Genant
Journal:  Eur J Radiol       Date:  1995-09       Impact factor: 3.528

7.  Osteoporosis: review of the evidence for prevention, diagnosis and treatment and cost-effectiveness analysis. Executive summary.

Authors: 
Journal:  Osteoporos Int       Date:  1998       Impact factor: 4.507

8.  Assessment of a model-based, iterative reconstruction algorithm (MBIR) regarding image quality and dose reduction in liver computed tomography.

Authors:  Won Chang; Jeong Min Lee; Kyunghee Lee; Jeong Hee Yoon; Mi Hye Yu; Joon Koo Han; Byung Ihn Choi
Journal:  Invest Radiol       Date:  2013-08       Impact factor: 6.016

9.  Strategies for formulating appropriate MDCT techniques when imaging the chest, abdomen, and pelvis in pediatric patients.

Authors:  Dianna D Cody; Donna M Moxley; Kerry T Krugh; Jennifer C O'Daniel; Louis K Wagner; Farzin Eftekhari
Journal:  AJR Am J Roentgenol       Date:  2004-04       Impact factor: 3.959

10.  Comparison of low-dose with standard-dose multidetector CT in cervical spine trauma.

Authors:  T H Mulkens; P Marchal; S Daineffe; R Salgado; P Bellinck; B te Rijdt; B Kegelaers; J-L Termote
Journal:  AJNR Am J Neuroradiol       Date:  2007-09       Impact factor: 3.825

  10 in total
  5 in total

1.  A phantom study comparing low-dose CT physical image quality from five different CT scanners.

Authors:  Yali Li; Yaojun Jiang; Huilong Liu; Xi Yu; Sihui Chen; Duoshan Ma; Jianbo Gao; Yan Wu
Journal:  Quant Imaging Med Surg       Date:  2022-01

2.  The study protocol for the China Health Big Data (China Biobank) project.

Authors:  Yan Wu; Zhe Guo; Xiaoxia Fu; Jing Wu; Jianbo Gao; Qiang Zeng; Haihong Fu; Xiaoguang Cheng
Journal:  Quant Imaging Med Surg       Date:  2019-06

3.  Application of Dual-Energy Spectral Computed Tomography in Bone Mineral Density Measurement: Phantom and Clinical Research.

Authors:  Mingyue Wang; Yan Wu; Yue Zhou; Junqiang Dong; Shenshen Hu; Ping Hou; Jianbo Gao
Journal:  Int J Gen Med       Date:  2022-08-29

4.  Deep-learning image reconstruction for image quality evaluation and accurate bone mineral density measurement on quantitative CT: A phantom-patient study.

Authors:  Yali Li; Yaojun Jiang; Xi Yu; Binbin Ren; Chunyu Wang; Sihui Chen; Duoshan Ma; Danyang Su; Huilong Liu; Xiangyang Ren; Xiaopeng Yang; Jianbo Gao; Yan Wu
Journal:  Front Endocrinol (Lausanne)       Date:  2022-08-11       Impact factor: 6.055

5.  Opportunistic Screening Using Low-Dose CT and the Prevalence of Osteoporosis in China: A Nationwide, Multicenter Study.

Authors:  Xiaoguang Cheng; Kaiping Zhao; Xiaojuan Zha; Xia Du; Yongli Li; Shuang Chen; Yan Wu; Shaolin Li; Yong Lu; Yuqin Zhang; Xigang Xiao; YueHua Li; Xiao Ma; Xiangyang Gong; Wei Chen; Yingying Yang; Jun Jiao; Bairu Chen; Yinru Lv; Jianbo Gao; GuoBin Hong; Yaling Pan; Yan Yan; Huijuan Qi; Limei Ran; Jian Zhai; Ling Wang; Kai Li; Haihong Fu; Jing Wu; Shiwei Liu; Glen M Blake; Perry J Pickhardt; Yuanzheng Ma; Xiaoxia Fu; Shengyong Dong; Qiang Zeng; Zhiping Guo; Karen Hind; Klaus Engelke; Wei Tian
Journal:  J Bone Miner Res       Date:  2020-11-04       Impact factor: 6.741

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.