Literature DB >> 29541529

Radio Frequency Immunity Testing of Hearing Aids.

Stephen D Julstrom1, Linda K Kozma-Spytek2, Brian B Beard3.   

Abstract

For many Equipment Under Test (EUT), such as the hearing aids examined in this study, the desired RF immunity measurement result is that which would be measured in the most sensitive EUT orientation relative to an applied RF field. This is generally approximated from measurements at a number of predetermined orientations within a GTEM cell. This paper presents new 6 and 12-orientation "maximal sum" methods of small EUT immunity measurement, which may be considered extensions to present sorted three-input vector summation techniques. Experimental results for the new methods approached the established reference goal more consistently than did other approaches examined employing a comparable number of contributing measurements.

Keywords:  Electromagnetic compatibility testing; GTEM; RF immunity testing; hearing aid interference

Year:  2013        PMID: 29541529      PMCID: PMC5846684          DOI: 10.1109/MEMC.2013.6550935

Source DB:  PubMed          Journal:  IEEE Electromagn Compat Mag        ISSN: 2162-2264


  3 in total

1.  Digital wireless telephones and hearing aids: new challenges for audiology.

Authors:  H Levitt; J Harkins
Journal:  J Am Acad Audiol       Date:  2001-06       Impact factor: 1.664

2.  Telecoil-mode hearing aid compatibility performance requirements for wireless and cordless handsets: magnetic signal-to-noise.

Authors:  Stephen Julstrom; Linda Kozma-Spytek; Scott Isabelle
Journal:  J Am Acad Audiol       Date:  2011-09       Impact factor: 1.664

3.  An evaluation of digital cellular handsets by hearing aid users.

Authors:  Linda Kozma-Spytek; Judith Harkins
Journal:  J Rehabil Res Dev       Date:  2005 Jul-Aug
  3 in total