Effects of Screw Configuration on the Preload Force of Implant-Abutment Screws.

Holger Zipprich, Florian Rathe, Sören Pinz, Luca Schlotmann, Hans-Christoph Lauer, Christoph Ratka.   

Abstract

PURPOSE: The aim of this study was to investigate the effects of tightening torque, screw head angle, and thread number on the preload force of abutment screws.
MATERIALS AND METHODS: The test specimens consisted of three self-manufactured components (ie, a thread sleeve serving as an implant analog, an abutment analog, and an abutment screw). The abutment screws were fabricated with metric M1.6 external threads. The thread number varied between one and seven threads. The screw head angles were produced in eight varying angles (30 to 180 degrees). A sensor unit simultaneously measured the preload force of the screw and the torsion moment inside the screw shank. The tightening of the screw with the torque wrench was performed in five steps (15 to 35 Ncm). The torque wrench was calibrated before each step.
RESULTS: Only the tightening torque and screw head angle affected the resulting preload force of the implant-abutment connection. The thread number had no effect. There was an approximately linear correlation between tightening torque and preload force.
CONCLUSION: The tightening torque and screw head angle were the only study parameters that affected the resulting preload force of the abutment screw. The results obtained from this experiment are valid only for a single torque condition. Further investigations are needed that analyze other parameters that affect preload force. Once these parameters are known, it will add value for a strong, but detachable connection between the implant and abutment. Short implants and flat-to-flat connections especially will benefit significantly from this knowledge.

Entities:  

Mesh:

Substances:

Year:  2018        PMID: 29534122     DOI: 10.11607/jomi.5837

Source DB:  PubMed          Journal:  Int J Oral Maxillofac Implants        ISSN: 0882-2786            Impact factor:   2.804


  4 in total

1.  Microgap and microleakage of a hybrid connection platform-switched implant system in the absence or presence of a silicone-based sealing agent.

Authors:  Pedro Ferrás Fernandes; Liliana Grenho; Maria Helena Fernandes; João Carlos Sampaio-Fernandes; Pedro Sousa Gomes
Journal:  Odontology       Date:  2021-08-27       Impact factor: 2.634

2.  Carfilzomib inhibits the proliferation and apoptosis of multiple myeloma cells by inhibiting STAT1/COX-2/iNOS signaling pathway.

Authors:  Shaolong He; Weiwei Tian; Jie Zhao; Rong Gong; Tao Wang; Liangming Ma
Journal:  Transl Cancer Res       Date:  2022-01       Impact factor: 1.241

3.  Influence of Anodizing Stages on the Preload Force of Implant-Abutment Screws and Their Benefits Regarding the Concept of Immediate Implant Placement-An In Vitro Study.

Authors:  Florian Rathe; Paul Weigl; Jan Wasiak; Christoph Ratka; Holger Zipprich
Journal:  Materials (Basel)       Date:  2022-01-20       Impact factor: 3.623

4.  Investigating the effect of abutment-implant connection type on abutment screw loosening in a dental implant system using finite element methods.

Authors:  Alireza Pournasrollah; Ramin Negahdari; Vahedeh Gharekhani; Ali Torab; Soheil Jannati Ataei
Journal:  J Dent Res Dent Clin Dent Prospects       Date:  2019
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.