| Literature DB >> 29524294 |
Ileana Iftimia1,2, Per H Halvorsen1.
Abstract
PURPOSE: The aim of this study was to develop Quality Assurance procedures for the BrainLab ExacTrac (ET) imaging system following the TG 142 recommendations for planar kV imaging systems.Entities:
Keywords: ExacTrac; quality assurance
Mesh:
Year: 2018 PMID: 29524294 PMCID: PMC5978564 DOI: 10.1002/acm2.12301
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
Figure 1(a) Photo with the holder, phantom, and paper template used for the ET MQA testing; photo with the MQA setup for tube #2, showing laser intersection on the phantom (b) face and (c) back.
Figure 2(a–c) Schematic diagrams with the holder‐phantom combination showing the distances used to estimate the couch vertical and distance from the inside bottom plate of the holder and the bottom front plane of the phantom (see text for details and notations).
Figure 3Photo showing the setup for the ET AQA testing.
Results for tube 1 (average over 24 sets)
| Image 1 | Image 2 | ||||||||
|---|---|---|---|---|---|---|---|---|---|
| Noise both img. | F50 | F40 | F30 | CNR | F50 | F40 | F30 | CNR | |
| Average | 1.25 | 1.07 | 1.19 | 1.31 | 70.3 | 1.07 | 1.19 | 1.31 | 70.4 |
| %Diff. min vs average | −16 | −0.9 | −0.7 | −0.7 | −4.7 | −1.1 | −0.8 | −0.6 | −4.7 |
| %Diff. max vs average | 12 | 3.2 | 1.7 | 0.6 | 3.0 | 3.5 | 1.7 | 0.4 | 3.2 |
Baseline values and results for both tubes (average over 24 sets)
| Noise | F50 | F40 | F30 | CNR | |
|---|---|---|---|---|---|
| Baseline values (five initial sets, for each tube) | 1.32 | 1.06 | 1.18 | 1.30 | 67.30 |
| Average (both tubes; both img., 24 sets) | 1.29 | 1.07 | 1.19 | 1.31 | 66.90 |
| %diff. av. vs baseline | −2.0 | 0.5 | 0.6 | 0.7 | −0.6 |
| Refined tolerances (see text for details) | ±20% warning; ±25% fail | ±5% warning; ±10% fail | ±5% warning; ±10% fail | ±5% warning; ±10% fail | ±15% warning; ±20% fail |
Image quality parameters (average over five tests) for the 100 kV/1 mAs technique
| Noise | F50 | F40 | F30 | CNR | |
|---|---|---|---|---|---|
| Baseline values (five initial sets, for each tube for 80 kV/2mAs) | 1.32 | 1.06 | 1.18 | 1.30 | 67.30 |
| Average (both tubes; both img., five sets) for 100 kV/1mAs | 1.47 | 1.06 | 1.19 | 1.31 | 57.50 |
| %Diff. av. vs baseline at 80 kV/2 mAs | <11.5 | <0.5 | <1.0 | <1.0 | <−15.0 |
Figure 4Phantom image imported in ARIA with dimensions displayed.
AQA results for the ET system version 6.2.0 (kV and dose)
| Protocol | Technique kV/mAs | Meas. kV tube 1 | %Diff. from nominal value | Meas. kV tube 2 | %diff. from nominal value | Total dose (mGy) | %diff. from prior AQA |
|---|---|---|---|---|---|---|---|
| Abdomen high | 145/25 | 146.0 | 0.7 | 143.0 | −1.4 | 1.353 | 1.7 |
| Cranial high | 100/12.5 | 100.1 | 0.1 | 99.8 | −0.2 | 0.323 | 2.4 |
| Cranial low | 80/6.3 | 80.8 | 1.0 | 80.3 | 0.4 | 0.099 | 1.7 |
| Head and neck high | 100/10 | 100.4 | 0.4 | 99.3 | −0.7 | 0.258 | 2.0 |
| Pelvis high | 130/25 | 132.1 | 1.6 | 129.4 | −0.5 | 1.104 | 1.8 |
| Thorax medium | 120/25 | 121.9 | 1.6 | 120.1 | 0.1 | 0.952 | 2.5 |