Variable flip angle 3D ultrashort echo time (UTE) T1 mapping of mouse lung: A repeatability assessment.

Daniel F Alamidi1,2, Amir Smailagic3, Abdel W Bidar3, Nicole S Parker4, Marita Olsson3, Paul D Hockings5,6, Kerstin M Lagerstrand2, Lars E Olsson7.   

Abstract

BACKGROUND: Lung T1 is a potential translational biomarker of lung disease. The precision and repeatability of variable flip angle (VFA) T1 mapping using modern 3D ultrashort echo time (UTE) imaging of the whole lung needs to be established before it can be used to assess response to disease and therapy.
PURPOSE: To evaluate the feasibility of regional lung T1 quantification with VFA 3D-UTE and to investigate long- and short-term T1 repeatability in the lungs of naive mice. STUDY TYPE: Prospective preclinical animal study. POPULATION: Eight naive mice and phantoms. FIELD STRENGTH/SEQUENCE: 3D free-breathing radial UTE (8 μs) at 4.7T. ASSESSMENT: VFA 3D-UTE T1 calculations were validated against T1 values measured with inversion recovery (IR) in phantoms. Lung T1 and proton density (S0 ) measurements of whole lung and muscle were repeated five times over 1 month in free-breathing naive mice. Two consecutive T1 measurements were performed during one of the imaging sessions. STATISTICAL TESTS: Agreement in T1 between VFA 3D-UTE and IR in phantoms was assessed using Bland-Altman and Pearson 's correlation analysis. The T1 repeatability in mice was evaluated using coefficient of variation (CV), repeated-measures analysis of variance (ANOVA), and paired t-test.
RESULTS: Good T1 agreement between the VFA 3D-UTE and IR methods was found in phantoms. T1 in lung and muscle showed a 5% and 3% CV (1255 ± 63 msec and 1432 ± 42 msec, respectively, mean ± SD) with no changes in T1 or S0 over a month. Consecutive measurements resulted in an increase of 2% in both lung T1 and S0 . DATA
CONCLUSION: VFA 3D-UTE shows promise as a reliable T1 mapping method that enables full lung coverage, high signal-to-noise ratio (∼25), and spatial resolution (300 μm) in freely breathing animals. The precision of the VFA 3D-UTE method will enable better design and powering of studies. LEVEL OF EVIDENCE: 1 Technical Efficacy: Stage 2 J. Magn. Reson. Imaging 2018.
© 2018 International Society for Magnetic Resonance in Medicine.

Entities:  

Keywords:  3D radial; T1 mapping; biomarker; longitudinal relaxation time; lung imaging; mouse

Year:  2018        PMID: 29517831     DOI: 10.1002/jmri.25999

Source DB:  PubMed          Journal:  J Magn Reson Imaging        ISSN: 1053-1807            Impact factor:   4.813


  3 in total

1.  A pilot study of native T1-mapping for focal pulmonary lesions in 3.0 T magnetic resonance imaging: size estimation and differential diagnosis.

Authors:  Shuyi Yang; Fei Shan; Qinqin Yan; Jie Shen; Peiyan Ye; Zhiyong Zhang; Yuxin Shi; Rengyin Zhang
Journal:  J Thorac Dis       Date:  2020-05       Impact factor: 2.895

2.  Native T1-mapping and diffusion-weighted imaging (DWI) can be used to identify lung cancer pathological types and their correlation with Ki-67 expression.

Authors:  Guangzheng Li; Renjun Huang; Mo Zhu; Mingzhan Du; Jingfen Zhu; Zongqiong Sun; Kaili Liu; Yonggang Li
Journal:  J Thorac Dis       Date:  2022-02       Impact factor: 2.895

3.  Native T1 Mapping and Magnetization Transfer Imaging in Grading Bowel Fibrosis in Crohn's Disease: A Comparative Animal Study.

Authors:  Baolan Lu; Jinjiang Lin; Jinfang Du; Shaofu He; Qinghua Cao; Li Huang; Ren Mao; Canhui Sun; Ziping Li; Shiting Feng; Xuehua Li
Journal:  Biosensors (Basel)       Date:  2021-08-28
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.