Literature DB >> 29516955

Three-dimensional-printed individual template-guided 125I seed implantation for the cervical lymph node metastasis: A dosimetric and security study.

Yansong Liang1, Zeyang Wang1, Hongtao Zhang1, Zhen Gao1, Jinxin Zhao1, Aixia Sui1, Jing Zhao1, Zezhou Liu1, Juan Wang1.   

Abstract

OBJECTIVE: The objective of this study is to evaluate the dosimetric accuracy and pathway safety of the three-dimensional (3D)-printed individual template-guided 125I seed implantation for the treatment of cervical lymph node metastasis.
MATERIALS AND METHODS: A total of 15 consecutive patients with cervical lymph node metastasis were enrolled during September 2015-July 2017 (12 patients had a history of external beam radiotherapy, with the mean dose of 63 Gy), who undergone radioactive seed implantation guided by the 3D-printed individual template. The preplan was completed based on contrast-enhanced computed tomography images, and then, the 3D-printed individual template was printed according to the preplan. After the operation, the real-time dose verification was completed: the D90 was ranged from 60 to 113 Gy with median of 93 Gy and the number of seeds was ranged from 21 to 76 with median of 53. To observe the intraoperative complications and postoperative complications in 3 days, we collected and compared the deviation of the number of seeds, target volume, and dosimetric parameters (D90, V90, V100, and V150) between preplan and postoperative plan. RESULT: Intraoperatively, each template was observed to exactly fit and lock on the lamina, and the operation was successfully completed. No intraoperative complications and postoperative complications were observed in 3 days. There was no significant difference in P values between the two groups for all the parameters (P > 0.05).
CONCLUSION: The 3D-printed individual template-guided 125I seed implantation for the cervical lymph node metastasis has not only reduced the dosimetric differences between pre- and postplan but also lowered the difficulty of puncture, indicating that it was a safe and accurate guidance approach.

Entities:  

Keywords:  Head and neck neoplasms; individual template; radioactive seed implantation; three-dimensional printed

Mesh:

Substances:

Year:  2018        PMID: 29516955     DOI: 10.4103/jcrt.JCRT_619_17

Source DB:  PubMed          Journal:  J Cancer Res Ther        ISSN: 1998-4138            Impact factor:   1.805


  9 in total

1.  CT-MR Image Fusion for Post-Implant Dosimetry Analysis in Brain Tumor Seed Implantation- a Preliminary Study.

Authors:  Menglong Zhang; Cunkun Chu; Liyin Huang; Bijuan Hu
Journal:  Dis Markers       Date:  2022-05-17       Impact factor: 3.464

Review 2.  Three-Dimensional (3D) Printing in Cancer Therapy and Diagnostics: Current Status and Future Perspectives.

Authors:  Awaji Y Safhi
Journal:  Pharmaceuticals (Basel)       Date:  2022-05-27

3.  Significance of 125I radioactive seed implantation on growth differentiation factor and programmed death receptor-1 during treatment of oral cancer.

Authors:  Gang Xue; Yao Feng; Jia-Bin Li
Journal:  World J Clin Cases       Date:  2020-03-06       Impact factor: 1.337

4.  Computer-Assisted Design Template Guided Percutaneous Radiofrequency Thermocoagulation through Foramen Rotundum for Treatment of Isolated V2 Trigeminal Neuralgia: A Retrospective Case-Control Study.

Authors:  Ran Wang; Ying Han; Lijuan Lu
Journal:  Pain Res Manag       Date:  2019-03-03       Impact factor: 3.037

5.  The Accuracy of Individualized 3D-Printing Template-Assisted I125 Radioactive Seed Implantation for Recurrent/Metastatic Head and Neck Cancer.

Authors:  Bin Qiu; Yuliang Jiang; Zhe Ji; Haitao Sun; Jinghong Fan; Weiyan Li; Yuxia Shao; Ping Jiang; Junjie Wang
Journal:  Front Oncol       Date:  2021-03-31       Impact factor: 6.244

6.  Evaluation of the efficacy of CT-guided 3D template-assisted 125I seed implantation in the treatment of unresectable STS: a multicenter retrospective study.

Authors:  Guang Sheng Zhao; Song Liu; Jun Zhou; Ruo Yu Wang; Yong Chun Song; Hua Liu; Wen Cai Lv; Liang Yang; Chuang Li
Journal:  Sci Rep       Date:  2022-03-08       Impact factor: 4.379

7.  Evaluation of radioactive 125I seed implantation for the treatment of refractory malignant tumours based on a CT-guided 3D template-assisted technique: efficacy and safety.

Authors:  Guang Sheng Zhao; Song Liu; Liang Yang; Chuang Li; Ruo Yu Wang; Jun Zhou; Yue Wei Zhang
Journal:  BMC Cancer       Date:  2020-08-03       Impact factor: 4.430

8.  A novel three-dimensional template combined with MR-guided 125I brachytherapy for recurrent glioblastoma.

Authors:  Xiangmeng He; Ming Liu; Menglong Zhang; Roberto Blanco Sequeiros; Yujun Xu; Ligang Wang; Chao Liu; Qingwen Wang; Kai Zhang; Chengli Li
Journal:  Radiat Oncol       Date:  2020-06-08       Impact factor: 3.481

Review 9.  Dosimetry study of three-dimensional print template for 125I implantation therapy.

Authors:  Enli Chen; Yuwei Zhang; Hongtao Zhang; Chenfei Jia; Yansong Liang; Juan Wang
Journal:  Radiat Oncol       Date:  2021-06-24       Impact factor: 3.481

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.