Development of an immunoaffinity chromatography and LC-MS/MS method for the determination of 6 zearalenones in animal feed.

Mi Jin Lee1, Ho Jin Kim1.   

Abstract

A novel and simple method for detecting 6 zearalenones in animal feed using liquid chromatography coupled with triple quadrupole mass spectrometry (LC-MS/MS) and immunoaffinity columns (IAC) was developed. The chromatographic peaks of the 6 zearalenones were successfully identified by comparing their retention times and mass spectrum with reference standards. The mobile phase was composed of mobile phase A (water) and B (0.5% formic acid in ACN). Method validation was performed with linearity, sensitivity, selectivity, accuracy and precision. The limits of detection (LODs) for the instrument used to study zearalenones ranged from 0.3 to 1.1 μg/kg, and the limits of quantification (LOQs) ranged from 1.0 to 2.2 μg/kg. Average recoveries of the 6 zearalenones ranged from 82.5% to 106.4%. Method replication resulted in intra-day and inter-day peak area variation of <3.8%. The developed method was specific and reliable and is suited for the routine analysis of zearalenones in animal feed.

Entities:  

Mesh:

Substances:

Year:  2018        PMID: 29505591      PMCID: PMC5837099          DOI: 10.1371/journal.pone.0193584

Source DB:  PubMed          Journal:  PLoS One        ISSN: 1932-6203            Impact factor:   3.240


Introduction

Zearalenone (ZON) is fungal toxin, a harmful substance for feed. The worldwide spotlight is shed upon its contamination of human food and animal feed. It is known to have critical influence not only on humans but also on livestock. ZON is fungal toxin that is produced by Fusarium graminearum, Fusarium culmorum, and other pathogens in corn, corn-derivatives, feeds, grains, and other food[1, 2]. ZON has α-zearalanol(α-ZAL), β-zearalanol(β-ZAL), α-zearalenol(α-ZOL) and β-zearalenol(β-ZOL), zearalanone(ZAN), which are isomer and metabolite, and these six types are called Zearalenone group (ZEN)[3]. ZON is metabolized mainly to a-ZOL and b-ZOL. Both metabolites are metabolized to a-ZAL and b-ZAL, respectively. ZAN is also metabolized y ZON, a-ZAL and b-ZAL, and the six isomers on matabolic pathway are correlated with each other. The chemical structures of ZON and ZON metabolite are similar to 17-estradiol (E2) and other natural estrogens (Fig 1). Therefore, competitive bonding to estrogen acceptors is conducted due to the action of environmental hormones. Relative estrogenicity of ZON and its metabolites was determined in comparison with E2 (92% for α-ZOL, 18% for α-ZAL, 3.5% for β-ZAL, 1% for ZON and 0.44% for β-ZOL). Regardless of its low acute toxicity, ZEN is considered hepatotoxic, hematotoxic, immunotoxic, genotoxic, teratogenic and carcinogenic in many mammals[4].
Fig 1

Structure of 17-estradiol, a similar chemical substance to ZEN.

Currently, there exists an analysis method for ZON in feeds, but there are no analytical studies on its isomers. ZON exists in trace quantities at the ppb level in feeds, thus there is a possibility that analytical error may lead to a change in outcome. For more accurate and precise analysis, it is necessary to develop an analysis method for the 6 isomers. Various methods have been reported for the characterization of ZEN, such as high performance liquid chromatography (HPLC)[2, 5], liquid chromatography (LC) coupled with mass spectrometry (MS)[6], LC coupled with MS/MS[7-10], and gas chromatography(GC) coupled with MS[11-14]. LC-MS and LC-MS/MS, which provide information about the molecular mass and structural features of components, are considered more useful than other methods for the separation, identification, and quantification of the characteristic ZEN compounds. In this study, ZEN in animal feed (α-ZAL, β-ZAL, α-ZOL, β-ZOL, ZAN and ZON) were analyzed simultaneously. Simultaneous identification was initially developed using the IAC and LC-MS/MS for the six types of ZEN, α-ZAL, β-ZAL, α-ZOL, β-ZOL, ZAN and ZON. A novel analytic method is therefore reported, which is suitable for detecting and quantifying the ZEN present in animal feed, while also aiding in the management of feed standards.

Experimental

Samples

The ZEN present in animal feed were targeted in this study. Animal feed was used after directly acquiring it from a local farm and old feed was used rather than feeds currently sold in the market. These samples were obtained from farms in five different cities. A total of 75 animal feed samples were used,. All dried samples were pulverized into fine powders (HMF-100; HANIL Electric Co., Seoul, Korea). The pulverizer was set at a maximum speed of 22,000 rpm to give fine powders ranging in size from 400 to 1000 μm. All samples were stored at 4°C.

Chemicals and reagents

HPLC grade acetonitrile and methanol were purchased from Merck (Darmstadt, Germany). Phosphate buffered saline(PBS) was purchased from Sigma (St. Louis, Mo, USA). ZEN(α-ZAL, β-ZAL, α-ZOL, β-ZOL, ZAN, ZON) were purchased from Sigma (St. Louis, Mo, USA). As purification IAC, immunoaffinity columns for Zearalenones (IAC-ZER) was purchased from Clover (Irvine, USA). All other organic chemicals and organic solvents were reagent grade or higher.

Standard preparation

The ZEN was dissolved with acetonitrile in a 10 mL volumetric flask to produce stock standard solutions. The concentrations of stock solutions ranged from 1.0 to 10 mg/mL. The appropriate volumes of each stock solution were mixed together, and then diluted serially to prepare the working standard solutions. All solutions were stored under refrigeration.

Sample preparation

The samples were prepared for extraction by IAC. Five grams of homogenized sample were place into a centrifuge tube and then mixed with 50 mL of solution (ACN:D.W, (80/20, [v/v])). The mixture was homogenized for 30 min, and sonicated (S450H; HUCOM system, Suwon, Korea) for 30 min at room temperature. It was then centrifuged at 3000 rpm for 5 min. The supernatant (10 mL) was mixed with 20 mL of PBS and filtered (Whatman, No. 4). 10 mL of the filtrate was passed through the IAC-ZER. 10 mL of water was passed through the IAC-ZER. The sample was eluted from the IAC-ZER by passing 2 mL of methanol, and all eluted samples were collected in a new test tube. The sample eluate was evaporated under a gentle stream of nitrogen at 50°C and the residue was dissolved in 500 μL of solution (40% ACN) and filtered through a 0.22 μm disposable filter prior to chemical analysis. All measurements were performed in triplicate.

HPLC-MS/MS analysis

HPLC analysis was performed on an Agilent 1200 HPLC system coupled with an Agilent 6460 triple quadrupole mass spectrometer (Agilent Technologies, Diegem, Belgium) in positive ion mode ([M+H]+) that was equipped with an Agilent 1260 series HPLC system. The chromatographic column used was a UK-C18 column (150 x 3 mm, 3 μm) (Tokyo, Japan) at 40°C. The mobile phase was filtered through a 0.45 μm membrane filter (Millipore, Milford, MA, USA) and degassed under vacuum. The mobile phase was composed of mobile phase A (water) and B (0.5% formic acid in ACN) with the following gradient elution: 0 min, 41% B; 0–5 min, 41% B; 5–35 min, 42.2% B; 35–40 min, 42.2% B; 40–40.1 min, 41% B; 40.1–45 min, 41% B. The sample injection volume was 10 μL, and the flow rate was set at 0.5 mL/min. The optimized electrospray ionization conditions were as follows: gas temperature, 150°C; gas flow, 9 L/min; sheath gas temperature, 350°C; sheath gas flow, 11 L/min; capillary voltage, 3000 V. Ionization condition and multiple reaction monitoring(MRM) for quantitative analysis are as illustrated in the Table 1
Table 1

Multiple reaction monitoring (MRM) conditions for the six Zearalenone group (ZEN) compounds by LC-MS/MS.

Compound NameQ1 (m/z)Q3 (m/z)FragmentorCollision EnergyCell Accelerator VoltagePolarity
α-zearalanol (α-ZAL)323.2287.450107positive
305.25007positive
β-zearalanol (β-ZAL)323.2305.16007positive
287.160107positive
α-zearalenol (α-ZOL)321.9303.15057positive
285.25057positive
β-zearalenol (β-ZOL)321.2303.26007positive
285.16057positive
Zearalanone (ZAN)321.2303.26057positive
285.160107positive
Zearalenone (ZON)319.2301.16007positive
283.16057positive

Method validation

Method validation was performed according to the guidelines set by the International Conference on Harmonization (ICH, 2005)[15] and the International Union of Pure and Applied Chemistry (IUPAC, 2002)[16]. The method was validated for linearity, sensitivity, selectivity, accuracy and precision.

Results and discussion

The LC-MS/MS method had been developed by using IAC was validated to verify that its performance was compatible with the required performance for routine ZEN analysis in animal feeds. A few performance characteristics were measured, including selectivity, linearity, sensitivity, accuracy, and precision[17, 18]. The selectivity was determined without interfering on the chromatographic window. The chromatograms of the ZEN indicate the separation of all six compounds in <35 min successful in high good resolution (1.32–18.22) and asymmetry (0.92–1.50) as shown in Fig 2 and Tables 2–4, thereby, it displays the satisfactory selectivity of this LC-MS/MS system.
Fig 2

LC-MS/MS chromatograms of in a standard mixture of 6 zearalenone metabolites.

Standard mixture of the six types of zearalenone group was put to 2 μL injection. (1) β-zearalanol(β-ZAL), (2) β-zearalenol(β-ZOL), (3) α-zearalanol(α-ZAL), (4) α-zearalenol(α-ZOL) (5) zearalanone(ZAN), (6) zearalenone(ZON).

Table 2

Validation parameters of the developed LC-MS/MS method.

CompoundRTaslopeinterceptr2bLODc (μg/kg)LOQd (μg/kg)RseSymmf
α-zearalanol (α-ZAL)20.0126070-200820.99940.41.3-0.92
β-zearalanol (β-ZAL)14.3036473-11641390.99990.31.01.550.97
α-zearalenol (α-ZOL)21.9650152-17248190.99980.62.28.770.99
β-zearalenol (β-ZOL)15.0311336-4165810.99990.62.12.900.93
Zearalanone (ZAN)32.2312803-4824540.99970.62.018.221.21
Zearalenone (ZON)33.5416172-4409450.99981.13.11.321.50

a Retention time (min).

b Coefficients of correlation

c Limit of detection, the lowest analyte concentration that produces a response detectable above the noise level of the system.

d Limit of quantification, the lowest level of analyte that can be accurately and precisely measured.

e Resolution.

f Symmetry.

Table 4

Recovery of Zearalenone group(ZEN) in animal feeds.

CompoundSpike levelaverageStandard deviation%RSD
2 μg/kg20 μg/kg
α-ZAL96.8898.2998.3498.7397.3299.7398.220.930.95
β-ZAL105.08103.12104.06102.35105.13104.52104.041.020.98
α-ZOL88.3089.5789.1389.4189.9688.2789.110.630.71
β-ZOL82.6880.3185.1882.1883.7581.1582.541.611.95
ZAN94.9995.4194.0793.7394.4095.6794.710.700.74
ZON107.11106.08105.85107.29106.33106.01106.440.550.52

LC-MS/MS chromatograms of in a standard mixture of 6 zearalenone metabolites.

Standard mixture of the six types of zearalenone group was put to 2 μL injection. (1) β-zearalanol(β-ZAL), (2) β-zearalenol(β-ZOL), (3) α-zearalanol(α-ZAL), (4) α-zearalenol(α-ZOL) (5) zearalanone(ZAN), (6) zearalenone(ZON). a Retention time (min). b Coefficients of correlation c Limit of detection, the lowest analyte concentration that produces a response detectable above the noise level of the system. d Limit of quantification, the lowest level of analyte that can be accurately and precisely measured. e Resolution. f Symmetry. The linearity was evaluated by building external calibration curves for each compound using the ZEN-containing working solutions. The calibration curves were obtained by plotting the analyte peak area with respect to its seven differed concentrations. The respective concentration of the mixed standard solution was injected in triplicate, and then, the regression parameters were calculated. The results are shown in Table 2. The correlation coefficients (r2>0.999) were obtained for all compounds studied. These results demonstrate that an external standard calibration can be applied for quantitative purposes. The sensitivity of the developed method was evaluated by determining the limits of detection (LOD) and quantification (LOQ) values. These values were calculated based on the response and slope of each regression equation at signal-to-noise ratios (S/N) of 3:1 and 10:1 respectively under the described chromatographic conditions. The LOD values were ranged across from 0.3 to 1.1 μg/kg for the different components, while the LOQ values ranged from 1.0 to 3.1 μg/kg. Detailed data are displayed in Table 2. The method precision was then determined by measuring the intra- and inter-day precision. For the intra-day precision, six replicates of the mixed standard solutions were analyzed within a day, while for inter-day precision, the solutions were examined in triplicate for three consecutive days. The precision was expressed by the percentage of the relative standard deviation (%RSD). The overall %RSD values for the intra-days were < 3.8%, while the inter-day values were < 3.2% (see Table 3).
Table 3

Intra- and inter- day precision and accuracy data of Zearalenone group(ZEN) determination (n: Independent replicates).

Nominal (conc, μg/kg)Intra-day (n = 6)Inter-day (n = 9)
Mean measured concentration (μg/kg)RSD (%)Mean measured concentration (μg/kg)RSD (%)
α-ZAL21.922.51.843.2
2019.4019.16
β-ZAL21.883.81.902.7
2019.0019.14
α-ZOL21.912.21.901.8
2019.3519.44
β-ZOL21.883.21.902.4
2019.3119.33
ZAN21.911.81.912.3
2019.5119.20
ZON21.931.41.931.8
2019.6619.52
The accuracy was evaluated by adding the mixed standard solutions at two different concentrations (high: 20.0 μg/kg and low: 2.0 μg/kg) to the feed supplied from the Association of American Feed Control Official (AFFCO, 201625-Swine Feed, Medicated). The mixture was filtered and extracted using the developed IAC method. All tests were performed in triplicate. The results showed that the recovery rates were 82.5% -106.4% and also indicated that the accuracy level of the method was high. The details of the data are shown in Table 4. Based on the above validation data, it was concluded that the proposed method provides excellent linearity, sensitivity, selectivity, accuracy, and precision for the simultaneous analysis of ZEN.

Application of the developed method

ZEN was detected in all 75 animal feeds acquired from local farms that were analyzed in this study (Table 5). Each sample was analyzed in triplicate. Identification of the six compounds was by comparison of their retention times and mass spectrum with those of the standards and the pure compounds (Tables 1–4). ZEN was detected in 5 different types of animal feeds: dog, chicken, duck, pig and cattle, and ZEN was found in all of them. The reason for this lies in the fact that ZEN is expected to convert into and remain as other isomers due to oxidation and reduction in contamination feeds. Additionally, all ZEN was detected. One sample was detected from cattle, two samples were detected from chicken and duck, and three or four samples were detected from dog and pig. From the dog feed, β-ZAL and ZAN were each detected in the quantity of 2.40–18.73 μg/kg and 20.13–78.09 μg/kg. From chicken feed, α-ZAL was extracted in the quantity of 13.67–19.10 μg/kg and β-ZOL was detected lower than LOD. From duck feed, α-ZOL (4.19 μg/kg) and ZON (39.08–47.61 μg/kg) were extracted. For pig feed from which four samples were extracted, α-ZAL, β-ZAL, α-ZOL, ZAN and ZON were extracted. Detailed results are illustrated in Table 5.
Table 5

Determination of the contents (mg/kg) of 6 compounds in feed using the proposed method.

Feed typeNo.*No. of DetectedConcentration(μg/kg)
α-zearalanol (α-ZAL)β-zearalanol (β-ZAL)α-zearalenol (α-ZOL)β-zearalenol (β-ZOL)Zearalanone (ZAN)Zearalenone (ZON)
Dog163-*2.40~18.73--20.13~78.09-
Chicken13213.67~19.10--N.D.*--
Duck152--4.19--39.08~47.61
Pig1742.31~2.483.11N.D.-4.71~6.69124.78
Cattle141-2.45---N.D.

*No.: Number of total tested sample

-: Not Detected; N.D.: Lower than LOD

*No.: Number of total tested sample -: Not Detected; N.D.: Lower than LOD A simple qualitative and quantitative method for simultaneous determination of ZEN compounds from feeds was successfully developed and validated using LC-MS detection. The proposed method showed appropriate accuracy and precision, and it was successfully used for analyzing different types of feeds. The satisfactory results demonstrated that the LC-MS method provides a good alternative for routine analysis owing to its simplicity, specificity and sensitivity and to its potential to be applied as a reliable quality evaluation method for animal feed.
  13 in total

1.  Guidelines from the International Conference on Harmonisation (ICH).

Authors:  Sarah K Branch
Journal:  J Pharm Biomed Anal       Date:  2005-08-10       Impact factor: 3.935

2.  HPLC and UPLC methods for the determination of zearalenone in noodles, cereal snacks and infant formula.

Authors:  Hyun Ee Ok; Sung-Wook Choi; Meehye Kim; Hyang Sook Chun
Journal:  Food Chem       Date:  2014-05-09       Impact factor: 7.514

3.  Validation of a UHPLC-MS/MS method for quantification of zearalenone, α-zearalenol, β-zearalenol, α-zearalanol, β-zearalanol and zearalanone in human urine.

Authors:  H Belhassen; I Jiménez-Díaz; R Ghali; H Ghorbel; J M Molina-Molina; N Olea; A Hedili
Journal:  J Chromatogr B Analyt Technol Biomed Life Sci       Date:  2014-05-19       Impact factor: 3.205

4.  Assessment of estrogenic and anti-androgenic activities of the mycotoxin zearalenone and its metabolites using in vitro receptor-specific bioassays.

Authors:  José-Manuel Molina-Molina; Macarena Real; Inmaculada Jimenez-Diaz; Hidaya Belhassen; Abderazzak Hedhili; Pablo Torné; Mariana F Fernández; Nicolás Olea
Journal:  Food Chem Toxicol       Date:  2014-12       Impact factor: 6.023

5.  Direct analysis of several Fusarium mycotoxins in cereals by capillary gas chromatography-mass spectrometry.

Authors:  Y Onji; Y Aoki; N Tani; K Umebayashi; Y Kitada; Y Dohi
Journal:  J Chromatogr A       Date:  1998-07-31       Impact factor: 4.759

6.  Development of an immunoaffinity chromatography and HPLC-UV method for determination of 16 sulfonamides in feed.

Authors:  Ho Jin Kim; Min Hee Jeong; Hye Jin Park; Won Chan Kim; Jang Eok Kim
Journal:  Food Chem       Date:  2015-10-09       Impact factor: 7.514

7.  Simultaneous determination of zearalenone and its derivatives in edible vegetable oil by gel permeation chromatography and gas chromatography-triple quadrupole mass spectrometry.

Authors:  Mingrong Qian; Hu Zhang; Liqin Wu; Nuo Jin; Jianmei Wang; Kezhi Jiang
Journal:  Food Chem       Date:  2014-06-05       Impact factor: 7.514

Review 8.  Mycotoxins in rice.

Authors:  Kenji Tanaka; Yuki Sago; Yazhi Zheng; Hiroyuki Nakagawa; Masayo Kushiro
Journal:  Int J Food Microbiol       Date:  2007-08-09       Impact factor: 5.277

9.  Development of a liquid-chromatography tandem mass spectrometry and ultra-high-performance liquid chromatography high-resolution mass spectrometry method for the quantitative determination of zearalenone and its major metabolites in chicken and pig plasma.

Authors:  S De Baere; A Osselaere; M Devreese; L Vanhaecke; P De Backer; S Croubels
Journal:  Anal Chim Acta       Date:  2012-10-24       Impact factor: 6.558

10.  Study of the chemical derivatization of zearalenone and its metabolites for gas chromatography-mass spectrometry analysis of environmental samples.

Authors:  Said Kinani; Stéphane Bouchonnet; Sophie Bourcier; Jean-Marc Porcher; Sélim Aït-Aïssa
Journal:  J Chromatogr A       Date:  2008-03-13       Impact factor: 4.759

View more
  6 in total

1.  Practical Application of Urinary Zearalenone Monitoring System for Feed Hygiene Management of a Japanese Black Cattle Breeding Herd-The Relationship between Monthly Anti-Müllerian Hormone and Serum Amyloid A Concentrations.

Authors:  Oky Setyo Widodo; Makoto Etoh; Emiko Kokushi; Seiichi Uno; Osamu Yamato; Dhidhi Pambudi; Hiroaki Okawa; Masayasu Taniguchi; Mirni Lamid; Mitsuhiro Takagi
Journal:  Toxins (Basel)       Date:  2022-02-16       Impact factor: 4.546

Review 2.  Occurrence, Impact on Agriculture, Human Health, and Management Strategies of Zearalenone in Food and Feed: A Review.

Authors:  Dipendra Kumar Mahato; Sheetal Devi; Shikha Pandhi; Bharti Sharma; Kamlesh Kumar Maurya; Sadhna Mishra; Kajal Dhawan; Raman Selvakumar; Madhu Kamle; Awdhesh Kumar Mishra; Pradeep Kumar
Journal:  Toxins (Basel)       Date:  2021-01-26       Impact factor: 4.546

3.  Immunoaffinity Cleanup and Isotope Dilution-Based Liquid Chromatography Tandem Mass Spectrometry for the Determination of Six Major Mycotoxins in Feed and Feedstuff.

Authors:  Ying Liu; Yongpeng Jin; Qi Guo; Xiong Wang; Sunlin Luo; Wenjun Yang; Juntao Li; Yiqiang Chen
Journal:  Toxins (Basel)       Date:  2022-09-11       Impact factor: 5.075

Review 4.  Zearalenone (ZEN) in Livestock and Poultry: Dose, Toxicokinetics, Toxicity and Estrogenicity.

Authors:  Jundi Liu; Todd Applegate
Journal:  Toxins (Basel)       Date:  2020-06-07       Impact factor: 4.546

Review 5.  Pre-Concentration and Analysis of Mycotoxins in Food Samples by Capillary Electrophoresis.

Authors:  Raffaella Colombo; Adele Papetti
Journal:  Molecules       Date:  2020-07-29       Impact factor: 4.411

Review 6.  Isoflavones in Animals: Metabolism and Effects in Livestock and Occurrence in Feed.

Authors:  Dino Grgic; Elisabeth Varga; Barbara Novak; Anneliese Müller; Doris Marko
Journal:  Toxins (Basel)       Date:  2021-11-24       Impact factor: 4.546

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.