| Literature DB >> 29497227 |
L Lakosi1, J Zsigrai2, A Kocsonya1, T C Nguyen1, H Ramebäck3,4, T Parsons-Moss5, N Gharibyan5, K Moody5.
Abstract
Low enriched uranium samples of unknown origin were analyzed by 16 laboratories in the context of a Collaborative Materials Exercise (CMX), organized by the Nuclear Forensics International Technical Working Group (ITWG). The purpose was to compare and prioritize nuclear forensic methods and techniques, and to evaluate attribution capabilities among participants. This paper gives a snapshot of the gamma spectrometric capabilities of the participating laboratories and summarizes the results achieved by gamma spectrometry.Entities:
Keywords: Gamma spectrometry; Intercomparison; Nondestructive assay; Nuclear forensics; Radiochronometry; Uranium isotopic composition
Year: 2018 PMID: 29497227 PMCID: PMC5820385 DOI: 10.1007/s10967-017-5667-2
Source DB: PubMed Journal: J Radioanal Nucl Chem ISSN: 0236-5731 Impact factor: 1.371
Average physical properties of the samples used in the exercise
| Sample ID | Physical form | Mass, g | Size, mm ± 0.1 | Approximate isotopic abundance, mass% | |||
|---|---|---|---|---|---|---|---|
| Diam. | Height | 234U | 235U | 238U | |||
| ES-1 | Powder | 2.9 | 0.025 | 2.9 | 97.1 | ||
| ES-2 | Pellet | 2.4 | 9.2 | 3.7 | 0.018 | 2.19 | 97.81 |
| ES-3 | Pellet | 2.5 | 9.1 | 3.7 | 0.025 | 2.9 | 97.1 |
HPGe detectors and software used by participants to determine the abundance of the major U isotopes
| Lab code | Detector | Software |
|---|---|---|
| Michelangelo | ORTEC Micro-Detective-HX, coaxial, diam. 50 mm, height 30 mm, electrically cooled, | Identify |
| Van Gogh | Canberra Falcon5000, broad-energy, diam. 61.8 mm, height 31.70 mm, electrically cooled | Manual, intrinsic efficiency calibration |
| Monet | Ortec GMX40P, coaxial, diam 63.0 mm, height 63.8 mm, electrically cooled | FRAM 5.2, (Param. set: ULEU-coax120-1010) |
| Rembrandt | Canberra GL0510, planar, diam. 24.8 mm, height 10 mm, active area 500 mm2 | MGAU V.3.2 |
| Renoir | Canberra GL0515R, planar diam. 25.2 mm, height 15 mm, active area 500 mm2 | MGAU V.4.2 |
| Picasso | For 24 h report: Canberra GL2020R, planar, diam. 50.5 mm, height 20 mm | MGAU V.4.2 |
| Buonarroti | Canberra GL0210R, planar, | MGAU V.4.2 |
| Pollock | Canberra BE3820, broad-energy Ge, diam. 70 mm, height 20 mm | U235 v1.51 (MGA ++) |
| Gauguin | Ortec, coaxial (averages from different detectors were used) | GAMANAL |
Available data of isotope abundances in mass %, measured by gamma spectrometry
| Lab code/software | ES-1 powder | ES-2 pellet | ES-3 pellet | ||||||
|---|---|---|---|---|---|---|---|---|---|
| 234U | 235U | 238U | 234U | 235U | 238U | 234U | 235U | 238U | |
| Michelang./Identify | _ | 1.93 ± 0.14 (Identify) 3.03 ± 0.18 (FRAM) | _ | _ | 1.50 ± 0.06 (Identify)2.14 ± 0.1 (FRAM) | _ | _ | 1.93 ± 0.32 (Identify) 3.41 ± 0.41 (FRAM) | _ |
| Van Gogh/manual | _ | 3.3 ± 0.5 | _ | _ | 2.6 ± 0.2 | _ | _ | 2.6 ± 0.5 | _ |
| Monet/(FRAM) | 0.023 ± 0.004 | 2.614 ± 0.140 | 97.34 ± 0.140 | 0.019 ± 0.006 | 2.290 ± 0.138 | 97.677 ± 0.140 | 0.016 ± 0.012 | 2.659 ± 0.32 | 97.308 ± 0.322 |
| Rembrandt/MGAU V3.2 | 0.028 ± 0.016 | 2.85 ± 0.16 | 97.12 ± 0.18 | 0.014 ± 0.012 | 2.11 ± 0.15 | 97.86 ± 0.15 | 0.016 ± 0.012 | 2.72 ± 0.15 | 97.25 ± 0.16 |
| Renoir/MGAU V4.2 | 0.023 ± 0.0024 | 2.868 ± 0.047 | 97.11 ± 0.0475 | 0.0154 ± 0.0032 | 2.202 ± 0.0424 | 97.78 ± 0.0430 | 0.0244 ± 0.0041 | 2.833 ± 0.059 | 97.143 ± 0.06 |
| Picasso/MGAU V4.2 | 0.026 ± 0.004 | 2.823 ± 0.039 | 97.151 ± 0.040 | 0.018 ± 0.004 | 2.138 ± 0.034 | 97.844 ± 0.035 | 0.025 ± 0.004 | 2.787 ± 0.041 | 97.188 ± 0.041 |
| Buonarroti/MGAU V4.2 | 0.025 ± 0.0014 | 2.851 ± 0.038 | 97.124 ± 0.040 | 0.0192 ± 0.0012 | 2.154 ± 0.030 | 97.827 ± 0.030 | 0.0250v± 0.0010 | 2.861 ± 0.036 | 97.114 ± 0.036 |
| Pollock/U235 V1.51 | 0.025 ± 0.010 | 2.91 ± 0.020 | 97.07 ± 0.020 | 0.019 ± 0.010 | 2.16 ± 0.020 | 97.83 ± 0.020 | 0.019 ± 0.010 | 2.86 ± 0.020 | 97.12 ± 0.020 |
| Gauguin/GAMANAL | 0.01565 ± 0.00072 | 2.752 ± 0.080 | 97.233 ± 0.080 | 0.00747 ± 0.00075 | 2.21 ± 0.12 | 97.78 ± 0.12 | 0.00916 ± 0.00092 | 2.99 ± 0.15 | 97.00 ± 0.15 |
| Community average MS without outliers | 0.02433 ± 0.00093 | 2.875 ± 0.040 | 97.076 ± 0.0082 | 0.01881 ± 0.00108 | 2.190 ± 0.040 | 97.789 ± 0.046 | 0.02441 ± 0.0083 | 2.875 ± 0.043 | 97.084 ± 0.084 |
Expanded uncertainties’ coverage factor k = 2. The community average of mass spectrometric results is calculated without the outliers, and its expanded standard deviation (k = 2) is given
Fig. 1Relative biases with respect to the community average mass spectrometric result. The bias of the mass ratio 234U/238U versus the bias of 235U/238 is shown for all three samples. Uncertainties are displayed with a coverage factor of k = 1
Fig. 2Spectrum of ES-3 (live time = 55,008 s) and of the background (live time = 65,122 s) taken by a well-type HPGe detector in a low-background iron chamber. Both spectra are normalized to 60,000 s live time (Picasso). No surplus of the peak area of 214Bi at 609 keV was observed above background. Abundance of 232U was evaluated from the net peak areas at 583, 860, and 2614 keV of 208Tl, and those at 238 and 727 keV of 212Pb and 212Bi, respectively