Error analysis of mechanical system and wavelength calibration of monochromator.

Fudong Zhang1, Chen Chen1, Jie Liu1, Zhihong Wang1.   

Abstract

This study focuses on improving the accuracy of a grating monochromator on the basis of the grating diffraction equation in combination with an analysis of the mechanical transmission relationship between the grating, the sine bar, and the screw of the scanning mechanism. First, the relationship between the mechanical error in the monochromator with the sine drive and the wavelength error is analyzed. Second, a mathematical model of the wavelength error and mechanical error is developed, and an accurate wavelength calibration method based on the sine bar's length adjustment and error compensation is proposed. Based on the mathematical model and calibration method, experiments using a standard light source with known spectral lines and a pre-adjusted sine bar length are conducted. The model parameter equations are solved, and subsequent parameter optimization simulations are performed to determine the optimal length ratio. Lastly, the length of the sine bar is adjusted. The experimental results indicate that the wavelength accuracy is ±0.3 nm, which is better than the original accuracy of ±2.6 nm. The results confirm the validity of the error analysis of the mechanical system of the monochromator as well as the validity of the calibration method.

Entities:  

Year:  2018        PMID: 29495825     DOI: 10.1063/1.5006109

Source DB:  PubMed          Journal:  Rev Sci Instrum        ISSN: 0034-6748            Impact factor:   1.523


  1 in total

1.  Understanding the mechanical limitations of the performance of soft X-ray monochromators at MAX IV laboratory.

Authors:  Peter Sjöblom; Gabriela Todorescu; Samuli Urpelainen
Journal:  J Synchrotron Radiat       Date:  2020-02-19       Impact factor: 2.616

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.