Literature DB >> 29482670

Atrial septal defect closure with the new Cardia Ultrasept II™ device with interposed Goretex patch: Mexican experience - has the perforation of Ivalon's membrane been solved?

Roberto Mijangos-Vázquez1, Antonio J García-Montes2, Elena M Soto-López3, Verónica Guarner-Lans4, Carlos Zabal2.   

Abstract

OBJECTIVES: The objective of this study was to demonstrate the safety and feasibility of using the new Cardia Ultrasept II™ device with interposed Goretex patch referring to the perforation of polyvinyl alcohol membrane.
BACKGROUND: Great advances have been made in the development of devices for closure of atrial septal defect. The Cardia Ultrasept II™ with interposed Goretex patch is the modified last generation of Cardia devices, having the advantage of a super-low profile within the atria and an integral locking delivery-retrieval mechanism that ensures safe deployment. In addition, with the interposition of the Goretex, it has been possible to abolish perforation of Ivalon's membrane as a complication.Methods and resultsPatients with ostium secundum atrial septal defect with surrounding rims with a minimum length of 5 mm and who underwent atrial septal defect closure with the new Ultrasept II™ with Goretex patch were included from two paediatric cardiac centres. Primary end point was to determine perforation of the Goretex membrane at follow-up; secondary end point included right ventricular diastolic diameter. In total, 30 patients underwent atrial septal defect closure at a median age of 6 (1-29) years. At follow-up for 6 (range, 1-15) months, freedom from perforations was 100%. A continuous decrease in right ventricular diastolic diameter was found with an initial median of 30 (25-49) mm and after catheterisation of 27.5 (18-33) mm, p=0.01, and Z-score of 2.6 (1.7-3.6) versus 1.9 (1-2.9) after procedure, p=0.01.
CONCLUSIONS: The new modified generation of the Ultrasept II™ device with interposed Goretex patch is a good alternative to achieve atrial septal defect closure safely and feasibly with no membrane perforation at follow-up.

Entities:  

Keywords:  Atrial septal defect; CHD; Cardia Ultrasept II; Mexico

Mesh:

Substances:

Year:  2018        PMID: 29482670     DOI: 10.1017/S1047951118000100

Source DB:  PubMed          Journal:  Cardiol Young        ISSN: 1047-9511            Impact factor:   1.093


  2 in total

1.  Current outlook after 4 years since the beginning of interventional cardiology in congenital heart disease in Chiapas State

Authors:  Roberto Mijangos-Vázquez; Rogelio Hernández-Reyes
Journal:  Arch Cardiol Mex       Date:  2022-04-04

2.  A Rare Ultra-Long-Term Complication of Occluder Recanalization Due to Spontaneous Perforation of Polyvinyl Alcohol Membrane of Atrial Septal Defect Occluder: A Case Report and Review of the Literature.

Authors:  Kun Xiang; Huanwei Zhuang; Qin Wu; Mi Tang; Jinfu Yang; Chengming Fan
Journal:  Front Cardiovasc Med       Date:  2022-07-22