BACKGROUND: Location of the affected bronchus of pleural air leaks is the most important step of trans-bronchoscopic bronchial occlusion for the treatment of intractable pneumothorax. The balloon occlusion test is the most commonly used technique, but has failed in some cases. The aim of the present study was: (1) to determine if endo-bronchial end-tidal CO2 (EtCO2) measurement can identify the affected bronchus that is the source of a persistent pleural air leak; and (2) to establish a methodology for endo-bronchial EtCO2 testing in locating affected bronchus in intractable pneumothorax.Entities:
Keywords: balloon occlusion; bronchoscope; end-tidal CO2; interventional pulmonology; pneumothorax
Mesh:
Substances:
Year: 2018 PMID: 29480071 PMCID: PMC5937148 DOI: 10.1177/1753465818756564
Source DB: PubMed Journal: Ther Adv Respir Dis ISSN: 1753-4658 Impact factor: 4.031
Figure 1.Instruments for the endo-bronchial EtCO2 test.
Figure 2.Example of the endo-bronchial EtCO2 test. The arrow points to the tip of the EtCO2 sampling catheter. Sampling sites: carina (A), main bronchus of left lobe (B), orifice of left upper lobe (C), orifice of left basal segments (D).
Power of endo-bronchial EtCO2 test and balloon detection.
| EtCO2 test | Balloon occlusion test ( | Sum (%)[ | |
|---|---|---|---|
| ( | Positive ( | Negative ( | |
| Positive | 8 | 9 | 17 (60.7) |
| Negative | 10 | 1 | 11 (39.2) |
| Sum (%) | 18 (64.3) | 10 (35.7) | 28 (100) |
There is a significant difference between positive rates of EtCO2 and balloon occlusion test (p = 0.041), tested by exact chi-square test. #Combination of the positive rate of the balloon occlusion test and the EtCO2 test was 96.4% (27/28).
EtCO2 values among different sites.
| Site of sampling | EtCO2 values ( | t-values ( | t-values ( |
|---|---|---|---|
| Affected bronchial | 34.27 ± 8.49 | – | −3.477 ( |
| Non-affected bronchial | 41.45 ± 7.49 | −3.337 ( | – |
| Main carina | 40.00 ± 8.37 | −2.102 ( | −0.652 ( |
| Main bronchus of affected lung | 40.35 ± 8.24 | −2.246 ( | −0.496 ( |
Figure 3.Bronchial occlusion results: according to the combination detection results of the balloon occlusion test with EtCO2 detection, 27 patients underwent bronchial occlusion by the use of bronchial spigot, autologous blood or both. The successful rate of occlusion was 85.7%.