Literature DB >> 29475322

1.65 mm diameter forward-viewing confocal endomicroscopic catheter using a flip-chip bonded electrothermal MEMS fiber scanner.

Yeong-Hyeon Seo, Kyungmin Hwang, Ki-Hun Jeong.   

Abstract

We report a 1.65 mm diameter forward-viewing confocal endomicroscopic catheter using a flip-chip bonded electrothermal MEMS fiber scanner. Lissajous scanning was implemented by the electrothermal MEMS fiber scanner. The Lissajous scanned MEMS fiber scanner was precisely fabricated to facilitate flip-chip connection, and bonded with a printed circuit board. The scanner was successfully combined with a fiber-based confocal imaging system. A two-dimensional reflectance image of the metal pattern 'OPTICS' was successfully obtained with the scanner. The flip-chip bonded scanner minimizes electrical packaging dimensions. The inner diameter of the flip-chip bonded MEMS fiber scanner is 1.3 mm. The flip-chip bonded MEMS fiber scanner is fully packaged with a 1.65 mm diameter housing tube, 1 mm diameter GRIN lens, and a single mode optical fiber. The packaged confocal endomicroscopic catheter can provide a new breakthrough for diverse in-vivo endomicroscopic applications.

Entities:  

Year:  2018        PMID: 29475322     DOI: 10.1364/OE.26.004780

Source DB:  PubMed          Journal:  Opt Express        ISSN: 1094-4087            Impact factor:   3.894


  5 in total

1.  Lissajous Scanning Two-photon Endomicroscope for In vivo Tissue Imaging.

Authors:  Daniel Youngsuk Kim; Kyungmin Hwang; Jinhyo Ahn; Yeong-Hyeon Seo; Jae-Beom Kim; Soyoung Lee; Jin-Hui Yoon; Eunji Kong; Yong Jeong; Sangyong Jon; Pilhan Kim; Ki-Hun Jeong
Journal:  Sci Rep       Date:  2019-03-05       Impact factor: 4.379

2.  Endoscopic optical coherence tomography angiography using a forward imaging piezo scanner probe.

Authors:  Lara M Wurster; Ronak N Shah; Fabian Placzek; Simon Kretschmer; Michael Niederleithner; Laurin Ginner; Jason Ensher; Michael P Minneman; Erich E Hoover; Hans Zappe; Wolfgang Drexler; Rainer A Leitgeb; Çağlar Ataman
Journal:  J Biophotonics       Date:  2019-01-30       Impact factor: 3.207

Review 3.  Scanning and Actuation Techniques for Cantilever-Based Fiber Optic Endoscopic Scanners-A Review.

Authors:  Mandeep Kaur; Pierre M Lane; Carlo Menon
Journal:  Sensors (Basel)       Date:  2021-01-02       Impact factor: 3.576

Review 4.  MEMS Actuators for Optical Microendoscopy.

Authors:  Zhen Qiu; Wibool Piyawattanametha
Journal:  Micromachines (Basel)       Date:  2019-01-24       Impact factor: 2.891

Review 5.  Single-Pixel MEMS Imaging Systems.

Authors:  Guangcan Zhou; Zi Heng Lim; Yi Qi; Guangya Zhou
Journal:  Micromachines (Basel)       Date:  2020-02-20       Impact factor: 2.891

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.