On the use of a novel Ferrous Xylenol-orange gelatin dosimeter for HDR brachytherapy commissioning and quality assurance testing.

Eleftherios P Pappas1, Vasiliki Peppa1, Costas J Hourdakis2, Pantelis Karaiskos1, Panagiotis Papagiannis3.   

Abstract

PURPOSE: To evaluate a commercially available Ferrous-Xylenol Orange-Gel (FXG) dosimeter (TrueView™) coupled with Optical-Computed Tomography (OCT) read out, for 3D dose verification in an Ir-192 superficial brachytherapy application.
METHODS: Two identical polyethylene containers filled with gel from the same batch were used. One was irradiated with an 18 MeV electron field to examine the dose-response linearity and obtain a calibration curve. A flap surface applicator was attached to the other to simulate treatment of a skin lesion. The dose distribution in the experimental set up was calculated with the TG-43 and the model based dose calculation (MBCA) algorithms of a commercial treatment planning system (TPS), as well as Monte Carlo (MC) simulation using the MCNP code. Measured and calculated dose distributions were spatially registered and compared.
RESULTS: Apart from a region close to the container's neck, where gel measurements exhibited an over-response relative to MC calculations (probably due to stray light perturbation), an excellent agreement was observed between measurements and simulations. More than 97% of points within the 10% isodose line (80 cGy) met the gamma index criteria established from uncertainty analysis (5%/2 mm). The corresponding passing rates for the comparison of experiment to calculations using the TG-43 and MBDCA options of the TPS were 57% and 92%, respectively.
CONCLUSION: TrueView™ is suitable for the quality assurance of demanding radiotherapy applications. Experimental results of this work confirm the advantage of the studied MBDCA over TG-43, expected from the improved account of scatter radiation in the treatment geometry.
Copyright © 2017 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.

Entities:  

Keywords:  Brachytherapy; FXG; Fricke; Gel dosimetry; HDR; MBDCA

Mesh:

Substances:

Year:  2017        PMID: 29472082     DOI: 10.1016/j.ejmp.2017.12.015

Source DB:  PubMed          Journal:  Phys Med        ISSN: 1120-1797            Impact factor:   2.685


  3 in total

Review 1.  A review of dosimetric impact of implementation of model-based dose calculation algorithms (MBDCAs) for HDR brachytherapy.

Authors:  Yousif A M Yousif; Alexander F I Osman; Mohammed A Halato
Journal:  Phys Eng Sci Med       Date:  2021-06-17

2.  Study of the Optimal Composition and Storage Conditions of the Fricke-XO-Pluronic F-127 Radiochromic Dosimeter.

Authors:  Michał Piotrowski; Piotr Maras; Sławomir Kadłubowski; Marek Kozicki
Journal:  Materials (Basel)       Date:  2022-01-27       Impact factor: 3.623

3.  Impact of Salt on Thermal Stability and Dose Response of the Fricke-XO-Pluronic F-127 3D Radiotherapy Dosimeter.

Authors:  Michał Piotrowski; Piotr Maras; Radosław Wach; Slawomir Kadlubowski; Marek Kozicki
Journal:  Materials (Basel)       Date:  2022-07-28       Impact factor: 3.748

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.