A step towards standardization: A method for end-point titer determination by fluorescence index of an automated microscope. End-point titer determination by fluorescence index.

Teresa Carbone1, Michele Gilio2, Maria Carmela Padula2, Giuseppina Tramontano2, Salvatore D'Angelo2, Vito Pafundi3.   

Abstract

OBJECTIVES: Indirect Immunofluorescence (IIF) is widely considered the Gold Standard for Antinuclear Antibody (ANA) screening. However, the high inter-reader variability remains the major disadvantage associated with ANA testing and the main reason for the increasing demand of the computer-aided immunofluorescence microscope. Previous studies proposed the quantification of the fluorescence intensity as an alternative for the classical end-point titer evaluation. However, the different distribution of bright/dark light linked to the nature of the self-antigen and its location in the cells result in different mean fluorescence intensities. The aim of the present study was to correlate Fluorescence Index (F.I.) with end-point titers for each well-defined ANA pattern.
METHODS: Routine serum samples were screened for ANA testing on HEp-2000 cells using Immuno Concepts Image Navigator System, and positive samples were serially diluted to assign the end-point titer. A comparison between F.I. and end-point titers related to 10 different staining patterns was made.
RESULTS: According to our analysis, good technical performance of F.I. (97% sensitivity and 94% specificity) was found. A significant correlation between quantitative reading of F.I. and end-point titer groups was observed using Spearman's test and regression analysis. A conversion scale of F.I. in end-point titers for each recognized ANA-pattern was obtained.
CONCLUSIONS: The Image Navigator offers the opportunity to improve worldwide harmonization of ANA test results. In particular, digital F.I. allows quantifying ANA titers by using just one sample dilution. It could represent a valuable support for the routine laboratory and an effective tool to reduce inter- and intra-laboratory variability.
Copyright © 2018. Published by Elsevier B.V.

Keywords:  Antinuclear antibodies; Automation; Indirect immunofluorescence; Standardization

Mesh:

Substances:

Year:  2018        PMID: 29470974     DOI: 10.1016/j.jim.2018.02.014

Source DB:  PubMed          Journal:  J Immunol Methods        ISSN: 0022-1759            Impact factor:   2.303


  1 in total

Review 1.  Standardization and Quality Assessment Under the Perspective of Automated Computer-Assisted HEp-2 Immunofluorescence Assay Systems.

Authors:  Luigi Cinquanta; Nicola Bizzaro; Giampaola Pesce
Journal:  Front Immunol       Date:  2021-02-25       Impact factor: 7.561

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.