Lower airway obstruction due to a massive clot resulting from late bleeding following mini-tracheostomy tube insertion and subsequent clot removal and re-intubation.

Hiroshi Inoue1, Jun Ito1, Hiroaki Uchida1, Mariko Morita2, Takahiko Masuda3, Kazuhiro Yamaya3, Masaki Hata3, Shigeaki Kato4.   

Abstract

BACKGROUND: Easier to perform than the conventional procedure, mini-tracheostomy (MT) is widely used in the operating room or intensive care unit to remove sputum or other obstructions of the upper airway. This option, however, does carry the risk of various complications, including malposition, disposition, bleeding, and subcutaneous emphysema. Here, we report a case of endotracheal tube obstruction due to a massive clot resulting from late bleeding around the insertion site of an MT tube. This necessitated removal of the endotracheal tube together with the clot followed tube re-introduction. CASE
PRESENTATION: The patient was an 85-year-old man in whom an MT tube had been inserted 6 days earlier following aortic replacement surgery. On re-admittance to our intensive care unit, large amounts of hemosputum and clotting were observed around the insertion site of the tube. The MT tube was subsequently removed and tracheal intubation performed. Ventilation via the endotracheal tube proved impossible, however, and cardiac arrest ensued. Fiberoptic bronchoscopy revealed that the endotracheal tube was completely obstructed by a massive clot. Therefore, we immediately pushed the clot toward the right main bronchus to secure ventilation via the left lung. After many attempts to remove the massive clot, including suction and grasping with basket forceps, it was successfully dislodged by replacing the endotracheal tube with a new one while maintaining oxygenation by one-lung ventilation. Any small fragments of the clot that still remained were then removed by suction under fiberoptic bronchoscopy.
CONCLUSIONS: Here, we report a case of endotracheal tube obstruction due to a clot derived from very late (6 days) bleeding after insertion of an MT tube. The patient was successfully rescued by replacing the clot-bearing endotracheal tube with a new one. This experience suggests that the intensive care physician should be aware of the potential risk of clot retention in endotracheal tubes after the elapse of several days.

Entities:  

Keywords:  Clot; Endotracheal tube obstruction; Extubation; Fiberoptic bronchoscopy; Lower airway obstruction; Mini-tracheostomy

Year:  2017        PMID: 29457060      PMCID: PMC5804594          DOI: 10.1186/s40981-017-0087-4

Source DB:  PubMed          Journal:  JA Clin Rep        ISSN: 2363-9024


Background

Mini-tracheostomy (MT) was first reported in 1984 and is applied in cases of sputum retention [1] or acute airway obstruction [2]. It is usually performed in an operating room or intensive care unit (ICU). Although considered less invasive than an open tracheostomy, an MT carries the risk of complications, including malposition, disposition, bleeding, subcutaneous emphysema, and delayed healing [3-5]. Meanwhile, airway trouble is a major cause of death in the ICU [6]. Here, we report a case of endotracheal tube (ETT) obstruction due to a clot at the insertion site of an MT tube due to very late bleeding.

Case presentation

An 85-year-old man who had undergone ascending aortic replacement developed large amounts of sputum due to postoperative bronchopneumonia. At 16 days postoperatively, an MT with a Portex Mini-Trach® 2 Kit (Smith Medical International: Ashford, Kent, U.K.) was uneventfully performed to alleviate sputum retention, although he was still on warfarin for new onset of sustained arterial fibrillation and arterial flutter. At that time, he was placed on a prothrombin time-international normalization ratio of between 2 and 3. At 22 days, the amount of sputum increased and was, for the first time, accompanied by hemosputum. Fiberoptic bronchoscopy (FOB) revealed a large clot due to bleeding around the insertion site of the MT tube, but there was no bleeding from the bronchial tree. His respiratory condition gradually deteriorated, and his oxygen saturation decreased. After removing the MT tube to avoid interference, tracheal intubation with the ETT (internal diameter of 8.0 mm) was performed. However, immediately after this procedure, the patient’s respiratory condition deteriorated catastrophically and his lung could not be ventilated with a self-expandable respiratory bag. Cardiac arrest soon followed, and cardiopulmonary resuscitation was performed. Direct laryngoscopy reconfirmed that the ETT was securely positioned in the trachea. An FOB examination performed during resuscitation revealed a massive blood clot completely obstructing the ETT. The FOB was used to push the clot away from the ETT toward the right main bronchus. Thereafter, his left lung was successfully ventilated and spontaneous circulation resumed. Any attempts to remove the clot by drawing it out through the ETT consequently failed because it was too big or too solid. Fortunately, oxygenation was preserved via one-lung ventilation during these procedures. Finally, it was decided to replace the ETT after removing the original with the clot still just inside. The clot was subsequently secured with the FOB and basket forceps and pulled into the ETT, after which, both were removed and a new ETT successfully inserted (Fig. 1). Thereafter, other smaller clots and blood were removed with the FOB and absence of bleeding from the bronchial tree confirmed. Open tracheostomy was performed 14 days later. The patient was weaned from mechanical ventilation at 48 days after this event and discharged in satisfactory condition with no psychiatric or neurological damage.
Fig. 1

Removed clot and endotracheal tube. The removed clot lodged within the endotracheal tube (black arrow); the remaining clot (dotted arrow) removed by suction under fiberoptic bronchoscopy. ETT endotracheal tube

Removed clot and endotracheal tube. The removed clot lodged within the endotracheal tube (black arrow); the remaining clot (dotted arrow) removed by suction under fiberoptic bronchoscopy. ETT endotracheal tube

Discussion

Mini-tracheostomy is performed widely to remove accumulated sputum [1] and acute obstructions of the airway [2]. There is no clear evidence, however, suggesting the superiority of MT over other sputum clearance methods such as medical or physiotherapies. In a review of the management of sputum retention, Beach et al. noted that an MT might be useful along with physiotherapy in adults following thoracic surgery. Its utility in other situations, including an acute care setting, remains unclear [7]. Although an MT is considered less invasive and more easily performable than conventional tracheostomy, it carries the risk of a number of complications [3-5]. Catastrophic airway obstruction due to clotting has been reported, but in most cases, this has occurred during or immediately after insertion of an MT tube [4, 5]. Late bleeding after MT tube insertion can therefore be considered a rare occurrence: indeed, in one case, the latest bleeding time was reported as occurring at approximately 38 h after insertion of an MT tube [8]. In the present case, bleeding appeared to occur at 6 days after insertion, and meanwhile, the patient had not complained of hemosputum. Fiberoptic bronchoscopy revealed a massive clot in the tracheal wall around the insertion site of the MT tube, but not in the lower bronchi (Fig. 2). It was assumed that this clot, which became detached upon tracheal intubation, was obstructing ventilation in the trachea. The etiology of this late bleeding at the insertion site in the present case is unknown at this stage, but we cannot exclude the possibility that it was due to the anti-coagulant effect of the warfarin. Mini-tracheostomy tube insertion is not contraindicated in anticoagulated patients in Japan. However, the present case suggests that MT tube insertion carries the risk of abnormal bleeding at the insertion site in such patients. This indicates the need to be aware of the potential of such an event when performing MT tube insertion in patients receiving anticoagulants.
Fig. 2

Schematic of the mechanism of occlusion of the endotracheal tube. Suspected three-stage mechanism of airway obstruction. a Pre-intubation condition of the airway containing a clot around the insertion site of a mini-tracheostomy tube. b The clot falling into the trachea when the mini-tracheostomy tube was removed. c Lower airway obstruction caused by a clot. MT mini-tracheostomy, ETT endotracheal tube

A number of methods are used to remove clots from the lower airway, including suction, forceps under FOB, a rigid bronchoscope, an arterial embolectomy catheter, and fibrinolytic drugs to the airway. Bodenham reported the development of a unique suction technique, which involved combining the ETT with a Y-connecter as a suction catheter [9]. If unsuccessful, a number of other options are available. These include rigid bronchoscopy, an arterial embolectomy catheter (i.e., number 6 Fogarty catheter inflated with 1.5 ml of air) [10], and topical thrombolysis. This latter option involves the application of 3000 to 12,000 units of streptokinase [11, 12] or 2500 units of urokinase [13] mixed with normal saline under visualization of the clot via FOB. Replacement of a clot-bearing ETT with a new one offers another option but can only be considered if the condition of the patient allows. Only patients in whom respiratory and/or circulatory conditions are not compromised may be eligible for ETT tube replacement. These techniques are summarized in Table 1.
Table 1

Various previously described methods of removing airway clots

Tracheal suction
Suction with flexible FOB
Basket forceps under FOB
Rigid bronchoscope with suction and/or forceps
Arterial embolectomy passing catheter as far as the distal aspect of the clot
Topical fibrinolytic drugs administered to the airway
Removal of the endotracheal tube and clot

FOB fiberoptic bronchoscope

Various previously described methods of removing airway clots FOB fiberoptic bronchoscope

Conclusions

Here, we reported a rare case of ETT obstruction due to very late bleeding caused by the insertion of an MT tube. Lower airway obstruction is a potentially fatal complication, and the intensive care physician doctor must be aware of the risk of airway troubles due to clot retention in the tubes. Schematic of the mechanism of occlusion of the endotracheal tube. Suspected three-stage mechanism of airway obstruction. a Pre-intubation condition of the airway containing a clot around the insertion site of a mini-tracheostomy tube. b The clot falling into the trachea when the mini-tracheostomy tube was removed. c Lower airway obstruction caused by a clot. MT mini-tracheostomy, ETT endotracheal tube
  13 in total

1.  Removal of obstructing blood clot from the lower airway: an alternative suction technique.

Authors:  A R Bodenham
Journal:  Anaesthesia       Date:  2002-01       Impact factor: 6.955

2.  Minitracheostomy--a cautionary tale of late bleeding.

Authors:  R K Choong; T Solano
Journal:  Anaesth Intensive Care       Date:  2001-10       Impact factor: 1.669

3.  Endobronchial thrombolysis with streptokinase for airway obstruction due to blood clots.

Authors:  Z Vajo; J M Parish
Journal:  Mayo Clin Proc       Date:  1996-06       Impact factor: 7.616

4.  Endobronchial streptokinase to dissolve a right mainstem clot.

Authors:  D B Thomson
Journal:  Chest       Date:  1986-06       Impact factor: 9.410

Review 5.  The efficacy of minitracheostomy for the management of sputum retention: a systematic review.

Authors:  Lisa Beach; Linda Denehy; Annemarie Lee
Journal:  Physiotherapy       Date:  2013-06-06       Impact factor: 3.358

6.  Clinical practice and risk factors for immediate complications of endotracheal intubation in the intensive care unit: a prospective, multiple-center study.

Authors:  Samir Jaber; Jibba Amraoui; Jean-Yves Lefrant; Charles Arich; Robert Cohendy; Liliane Landreau; Yves Calvet; Xavier Capdevila; Aba Mahamat; Jean-Jacques Eledjam
Journal:  Crit Care Med       Date:  2006-09       Impact factor: 7.598

7.  Minitracheotomy--a life-threatening complication.

Authors:  A K Daborn; M N Harris
Journal:  Anaesthesia       Date:  1989-10       Impact factor: 6.955

8.  Treatment of sputum retention by minitracheotomy.

Authors:  H R Matthews; R B Hopkinson
Journal:  Br J Surg       Date:  1984-02       Impact factor: 6.939

9.  Emergency airway clot removal in acute hemorrhagic respiratory failure.

Authors:  R P Allen; A D Siefkin
Journal:  Crit Care Med       Date:  1987-10       Impact factor: 7.598

10.  Is minitracheotomy a simple and safe procedure? A prospective investigation in the intensive care unit.

Authors:  J Pedersen; B A Schurizek; N C Melsen; B Juhl
Journal:  Intensive Care Med       Date:  1991       Impact factor: 17.440

View more
  1 in total

1.  Retrieval of a tracheobronchial blood clot with a Yankauer suction catheter in complete airway obstruction.

Authors:  Sharad Kaushik; Gaurav Jain; Namrata Gupta; Lokesh Kumar Saini; Girish Sindhwani
Journal:  Acute Crit Care       Date:  2021-01-13
  1 in total