B Derendinger1, M de Vos1, R R Nathavitharana2, T Dolby3, J A Simpson3, P D van Helden1, R M Warren1, G Theron4.
Abstract
The scale-up of rapid drug resistance testing for TB is a global priority. MTBDRplus is a WHO-endorsed multidrug-resistant (MDR)-TB PCR assay with suboptimal sensitivities and high indeterminate rates on smear-negative specimens. We hypothesised that widespread use of incorrect thermocycler ramp rate (speed of temperature change between cycles) impacts performance. A global sample of 72 laboratories was surveyed. We tested 107 sputa from Xpert MTB/RIF-positive patients and, separately, dilution series of bacilli, both at the manufacturer-recommended ramp rate (2.2 °C/s) and the most frequently reported incorrect ramp rate (4.0 °C/s). Mycobacterium tuberculosis-complex DNA (TUB-band)-detection, indeterminate results, accuracy, and inter-reader variability (dilution series only) were compared. 32 respondents did a median (IQR) of 41 (20-150) assays monthly. 78% used an incorrect ramp rate. On smear-negative sputa, 2.2 °C/s vs. 4.0 °C/s improved TUB-band positivity (42/55 vs. 32/55; p = 0.042) and indeterminate rates (1/42 vs. 5/32; p = 0.039). The actionable results (not TUB-negative or indeterminate; 41/55 vs. 28/55) hence improved by 21% (95% CI: 9-35%). Widespread use of incorrect ramp rate contributes to suboptimal MTBDRplus performance on smear-negative specimens and hence limits clinical utility. The number of diagnoses (and thus the number of smear-negative patients in whom DST is possible) will improve substantially after ramp rate correction.Entities:
Mesh:
Year: 2018 PMID: 29453405 PMCID: PMC5816666 DOI: 10.1038/s41598-018-21458-y
Source DB: PubMed Journal: Sci Rep ISSN: 2045-2322 Impact factor: 4.379
Figure 1Survey results showing breakdowns of (a) the geographical locations of laboratory survey respondents, (b) type of input material used (smear status and/or culture isolate) for MTBDRplus, (c) amount of MTBDRplus assays done per laboratory each month and (d) ramp rates used by respondents
Answers to survey questions stratified by continent.
| Country | MTBDR | MTBDR | MTBDR | MTBDR | MTBDR | MTBDR | Other LPAs | Thermocycler manufacturer | Thermocycler | Ramp rate (°C/s) | Customisable ramp rate | Pre-screening Xpert MTB/RIF |
|---|---|---|---|---|---|---|---|---|---|---|---|---|
|
| ||||||||||||
| South Africa | 900 | Diagnosis | Smear-positive, smear-negative, isolates | 50 | Diagnosis | Smear-positive, smear-negative, isolates | N/A | Applied Biosystems | SimpliAmp | 4.0 | Yes | Yes |
| Swaziland | 350 | Diagnosis, research | Smear-positive, isolates | 0 | Diagnosis, research | Smear-positive, isolates | N/A | Applied Biosystems | 2720 Thermal Cycler | 2.7 | No | Yes |
| South Africa | 245 | Diagnosis | Smear-positive, isolates | 0 | Diagnosis | Smear-positive, isolates | CM/AS | Applied Biosystems | ABI | 4.0 | Yes | Yes |
| Kenya | 240 | Diagnosis, research | Smear-positive, smear-negative | 0 | Research | Smear-positive | CM/AS | Applied Biosystems | GeneAmp 9700 | 2.3 | Yes | Yes |
| South Africa | 150 | Diagnosis, research | Isolates | 15 | Diagnosis | Isolates | N/A | Applied Biosystems | ABI | 6.0 | Yes | Yes |
| Ethiopia | 75 | Diagnosis, research | Smear-positive, isolates | 60 | Diagnosis, research | Smear-positive, smear-negative, isolates | CM/AS | Applied Biosystems | 2720 Thermal Cycler | 2.7 | No | Yes |
| South Africa | 50 | Diagnosis | Smear-positive, smear-negative, isolates | 45 | Diagnosis | Smear-positive, smear-negative, isolates | CM | Applied Biosystems | Proflex PCR system | 4.0 | Yes | No |
| Nigeria | 42 | Diagnosis, research | Smear-positive, smear-negative, isolates | 27 | Diagnosis, research | Smear-positive, smear-negative, isolates | CM/AS | Applied Biosystems | 2720 Thermal Cycler | 2.7 | No | Yes |
| South Africa | 40 | Diagnosis, research | Isolates | 0 | Diagnosis, research | Isolates | CM | Applied Diagnosis | Various | 4.0 | Yes | Yes |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| South Africa | 20 | Diagnosis, research | Smear-positive, smear-negative, isolates | 0 | Diagnosis, research | Smear-positive, smear-negative, isolates | CM | Bio-Rad | Various | 4.0 | No | No |
| Côte d’Ivoire | 14 | Diagnosis, research | Smear-positive, isolates | 4 | Diagnosis, research | Smear-positive, isolates | N/A | Applied Biosystems | 2720 Thermal Cycler | 2.7 | No | Yes |
|
| ||||||||||||
| Azerbaijan | 750 | Diagnosis, research | Smear-positive, isolates | 0 | Diagnosis, research | Smear-positive, isolates | N/A | Unknown | Unknown | 12.0 | No | Yes |
| Kyrgyzstan | 256 | Diagnosis | Smear-positive, smear-negative, isolates | 37 | Diagnosis | Smear-positive | N/A | Biometra | Tprofessional | 3.0 | Yes | Yes |
| Bangladesh | 150 | Diagnosis | Smear-positive, smear-negative, isolates | 120 | Diagnosis | Smear-positive, smear-negative, isolates | N/A | Applied Biosystems | 2720 Thermal Cycler | 2.7 | No | Yes |
| India | 150 | Diagnosis | Smear-positive, smear-negative, isolates | 0 | Diagnosis | Smear-positive, isolates | CM | Bio-Rad | Thermal Cycler T100 | 4.0 | No | No |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| Myanmar | 15 | Diagnosis | Smear-positive, smear-negative, isolates | 10 | Diagnosis | Smear-positive, smear-negative, isolates | N/A | Applied Biosystems | 2720 Thermal Cycler | 2.7 | No | Yes |
| Thailand | 5 | Diagnosis | Smear-positive, smear-negative | 0 | Research | Isolates | N/A | Applied Biosystems | 2720 Thermal Cycler | 2.7 | No | No |
|
| ||||||||||||
| Denmark | 90 | Diagnosis, research | Smear-positive, smear-negative, isolates | 5 | Diagnosis, research | Smear-positive, isolates | CM/AS | Applied Biosystems | SimpliAmp | 4.0 | Yes | No |
| Belarus | 75 | Diagnosis | Smear-positive, smear-negative, isolates | 80 | Diagnosis | Smear-positive, smear-negative, isolates | CM/AS | Applied Biosystems | 2720 Thermal Cycler | 2.7 | No | Yes |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| Belgium | 50 | Diagnosis, research | Smear-positive, isolates | 30 | Diagnosis, research | Smear-positive, isolates | Nipro | Biometra | Tprofessional | 3.0 | Yes | Yes |
| Belarus | 40 | Diagnosis | Smear-positive, isolates | 0 | Diagnosis | Smear-positive, isolates | CM | Applied Biosystems | 2720 Thermal Cycler | 2.7 | No | Yes |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| Denmark | 25 | Diagnosis, research | Smear-positive, smear-negative, isolates | 4 | Diagnosis, research | Smear-positive, smear-negative, isolates | CM/AS | Applied Biosystems | SimpliAmp | 4.0 | Yes | No |
| France | 25 | Diagnosis | Smear-positive, smear-negative, isolates | 17 | Diagnosis | Smear-positive, smear-negative, isolates | CM/AS, NTM-DR | Bio-Rad | Various | 4.0 | Yes | No |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| Spain | 10 | Diagnosis | Smear-positive, smear-negative, isolates | 2 | Diagnosis | Smear-positive, smear-negative, isolates | Inno-Lipa | Applied Biosystems | Various | 4.0 | Yes | No |
|
| ||||||||||||
| Brazil | 20 | Diagnosis, research | Smear-positive, smear-negative, isolates | 0 | Diagnosis, research | Smear-positive, smear-negative, isolates | MOTT ID | Bio-Rad | C1000 Touch | 5.0 | Yes | Yes |
|
| ||||||||||||
| 3987 | 632 | 16/32 respondents test smear-negative specimens and are hence likely affected by incorrect ramp rate usage | 7/32 used the correct ramp rate | 20/32 have a customisable ramp rate | 22/32 use Xpert as a pre-screen | |||||||
LPA - Line probe assay; HCV – HCV Genotype 2.0 Assay (LiPA); CM-Genotype Mycobacterium CM Ver 2.0; NTM-DR – GenoType NTM-DR Ver 1.0;
AS - Genotype Mycobacterium AS Ver 1.0; IQR - Interquartile Range; Xpert – Xpert MTB/RIF assay.
Survey questions are in the supplement. Text in bold refers to laboratories using the manufacturer-recommended ramp rate.
Performance of Genotype MTBDRplus at different ramp rates for the detection of M. tuberculosis-complex DNA (TUB-band), stratified by smear status, when done directly on sputa from Xpert MTB/RIF-positive patients. Genolysed extract from each specimen was tested at each ramp rate. TUB-band detection and the rate of indeterminates worsened in smear-negative specimens with use of the incorrect ramp rate.
| Smear microscopy status | Ramp rate (°C/s) | TUB-band positive (%) | Determinate (%) | Indeterminate (%) |
|---|---|---|---|---|
| Positive n = 52 | 4.0 | 52/52 (100) | 52/52 (100) | 0/52 (0) |
| 2.2* | 52/52 (100) | 52/52 (100) | 0/52 (0) | |
| Negative n = 55 | 4.0 | 32/55 (58) | 27/32 (84) | 5/32 (16) |
| 2.2* | 42/55 (76) (p = 0.042) | 41/42 (98) (p = 0.164) | 1/42 (2) (p = 0.039) |
P-values are for comparisons between ramp rates for smear-negative specimens.
*Manufacturer-recommended ramp rate.
Comparison of banding calls and diagnostic calls from two experienced readers of MTBDRplus done on serial dilutions of M. tuberculosis .
|
| ||
|---|---|---|
|
| ||
| Different band calls between readers for susceptible strain | 0/198* (0) | 2/198 (1) (p = 0.156) |
| Different band calls between readers for resistant strain | 1/198 (0.5) | 6/198 (3) (p = 0.057) |
|
| ||
| Susceptible strain correctly classified by both readers | 18/18 (100) | 16/18 (89) (p = 0.146) |
| Resistant strain correctly classified by both readers | 18/18 (100) | 16/18 (89) (p = 0.146) |
*22 bands per strip × three dilutions × three replicates = a total of 198 bands.
P-values are for within-row comparisons between different ramp rates.