Literature DB >> 29449354

High-Resolution Infrared Thermography of Esophageal Temperature During Radiofrequency Ablation of Atrial Fibrillation.

Matthew G Daly1, Iain Melton2, Graham Roper2, Gary Lim2, Ian G Crozier2.   

Abstract

BACKGROUND: Catheter ablation for atrial fibrillation has potential to cause esophageal thermal injury. Esophageal temperature monitoring during ablation is commonly used; however, it has not eliminated thermal injuries, possibly because conventional sensors have poor spatial sampling and response characteristics. To enhance understanding of temperature dynamics that may underlie esophageal injury, we tested a high-resolution, intrabody, infrared thermography catheter to continuously image esophageal temperatures during ablation. METHODS AND
RESULTS: Atrial fibrillation ablation patients were instrumented with a flexible, 9F infrared temperature catheter inserted nasally (n=8) or orally (n=8) into the esophagus adjacent to the left atrium. Ablation was performed while the infrared catheter continuously recorded surface temperatures from 7680 points per second circumferentially over a 6-cm length of esophagus. Physicians were blinded to temperature data. Endoscopy was performed within 24 hours to document esophageal injury. Thermal imaging showed that most patients (10/16) experienced ≥1 events where peak esophageal temperature was >40°C. Three patients experienced temperatures >50°C; and 1 experienced >60°C. Analysis of temperature data for each subject's maximum thermal event revealed high gradients (2.3±1.4°C/mm) and rates of change (1.5±1.3°C/s) with an average length of esophageal involvement of 11.0±5.4 mm. Endoscopy identified 3 distinct thermal lesions, all in patients with temperatures >50°C; all resolved within 2 weeks.
CONCLUSIONS: Infrared thermography provided dynamic, high-resolution mapping of esophageal temperatures during cardiac ablation. Esophageal thermal injury occurred with temperatures >50°C and was associated with large spatiotemporal gradients. Additional studies are warranted to determine the relationships between thermal parameters and esophageal injury.
© 2018 American Heart Association, Inc.

Entities:  

Keywords:  atrial fibrillation; catheter ablation; endoscopy; temperature; thermography

Mesh:

Year:  2018        PMID: 29449354     DOI: 10.1161/CIRCEP.117.005667

Source DB:  PubMed          Journal:  Circ Arrhythm Electrophysiol        ISSN: 1941-3084


  5 in total

1.  Preventing esophageal complications from atrial fibrillation ablation: A review.

Authors:  Lisa W M Leung; Zaki Akhtar; Mary N Sheppard; John Louis-Auguste; Jamal Hayat; Mark M Gallagher
Journal:  Heart Rhythm O2       Date:  2021-09-22

Review 2.  Prevention and Treatment of Atrioesophageal Fistula Related to Catheter Ablation for Atrial Fibrillation.

Authors:  George M Bodziock; Caleb A Norton; Jay A Montgomery
Journal:  J Innov Card Rhythm Manag       Date:  2019-05-15

3.  Double-parabolic-reflectors acoustic waveguides for high-power medical ultrasound.

Authors:  Kang Chen; Takasuke Irie; Takashi Iijima; Takeshi Morita
Journal:  Sci Rep       Date:  2019-12-06       Impact factor: 4.379

4.  The AF-FICIENT magnetic resonance imaging and endoscopy safety substudy: A visually guided radiofrequency balloon ablation catheter for pulmonary vein isolation.

Authors:  Matthew Daly; Michael Hurrell; Iain Melton; Gary Lim; Audrius Aidietis; Gediminas Rackauskas; Amin Al-Ahmad; Ian Crozier
Journal:  Heart Rhythm O2       Date:  2021-12-10

Review 5.  Protecting Against Collateral Damage to Non-cardiac Structures During Endocardial Ablation for Persistent Atrial Fibrillation.

Authors:  Lisa Wm Leung; Zaki Akhtar; Jamal Hayat; Mark M Gallagher
Journal:  Arrhythm Electrophysiol Rev       Date:  2022-04
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.