| Literature DB >> 29445716 |
A Tavakoli Golpaygani1, M M Movahedi2,3, M Reza4.
Abstract
INTRODUCTION: Nowadays, more than 10,000 different types of medical devices can be found in hospitals. This way, medical electrical equipment is being employed in a wide variety of fields in medical sciences with different physiological effects and measurements. Hospitals and medical centers must ensure that their critical medical devices are safe, accurate, reliable and operational at the required level of performance. Defibrillators are critical resuscitation devices. The use of reliable defibirillators has led to more effective treatments and improved patient safety through better control and management of complications during Cardiopulmonary Resuscitation (CPR).Entities:
Keywords: Defebrillator ; Electroshock ; Medical Electrical Equipment; Patient Safety ; Reliability ; Meterology
Year: 2017 PMID: 29445716 PMCID: PMC5809933
Source DB: PubMed Journal: J Biomed Phys Eng ISSN: 2251-7200
Figure1A simple circuit of a defibrillator
Figure2Different types of Waveforms
Performance parameters of defibrillators equipment
| Qualitative Tasks | Quantitative Tasks |
|---|---|
| Chassis/Housing | Accuracy of controls and instruments |
| Strain Reliefs | Charging time |
| Paddles/Electrodes | Internal electrical power source |
| Fitting/Connectors | SYNCHRONIZER |
| Controls/Switches | Device operation after battery Alarm |
| Indicators/Displays | Batteries |
| Alarm/Audible Signals | Energy waveform |
| Battery/Charger | Electrical Safety Tests |
A brief summary of evaluated technical result for each unit (units are frequent use, manual defibrillators)
| Items | U1 | U2 | U3 | U4 | U5 | U6 | U7 | U8 | U9 | U10 | U11 | U12 | U13 | U14 | U15 | U16 | U17 | U18 | U19 | U20 | |
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Energy Waveform | √ | √ | √ | √* | √* | √* | × | √ | √ | √ | × | √ | × | √ | × | √ | × | ||||
| CT | √ | √ | √ | × | √ | × | √ | × | √ | × | √ | √ | √ | × | √ | × | √ | × | √ | × | |
| SYNC. | √ | √ | √ | √ | √ | √ | √ | √ | √ | √ | √ | √ | √ | √ | √ | √ | √ | √ | √ | √ | |
| Battery | √ | √ | √ | × | √ | × | √ | × | √ | × | √ | √ | √ | × | √ | × | √ | × | √ | × | |
| DE | ** | * | * | ** | * | ** | * | ** | * | ** | * | * | * | ** | * | *** | ** | *** | * | *** | |
| ENRGY | E(A) | * | * | * | ** | * | ** | * | ** | * | ** | * | * | * | ** | * | ** | * | ** | * | *** |
| E(B) | * | * | * | ** | * | ** | * | ** | * | ** | * | * | * | ** | * | ** | * | ** | * | *** | |
| Patient Leakage Current | √ | √ | √ | × | √ | √ | √ | × | √ | √ | √ | √ | √ | × | √ | × | √ | × | √ | × | |
| Patient Auxillary Leakage Current | √ | √ | √ | × | √ | √ | √ | × | √ | × | √ | √ | √ | × | √ | × | √ | × | √ | × | |
| E(A): Accuracy of Delivered Energy in 50Ω load | E(B): Accuracy of Delivered Energy in different loads | ||||||||||||||||||||
| * 5-15% | ** 15-30% | *** Above 30% | |||||||||||||||||||
| CT:for Frequent Use, Manual Defibillators,Time Charging after 15 Sequential Discharge ≤ 15 s | Patient Auxillary Leakage Current | Patient Auxillary Leakage Current | DE: Delivered Energy after Discharge Alarm | ||||||||||||||||||
| Normal Condition | Fault Condition | ||||||||||||||||||||
| CF Type: 10 µA | CF Type: 50 µA | ||||||||||||||||||||
| BF Type: 100 µA | BF Type: 500 µA | ||||||||||||||||||||
| SYNC: Syncronization Operation | Patient Leakage Current | Patient Leakage Current | √ - In Standard Limitation | ||||||||||||||||||
| Normal Condition | Fault Condition | ||||||||||||||||||||
| NA: Not Applicable | NP: Not Performed | CF Type: 10 µA | CF Type: 50 µA | × - Out of Standard Limitation | |||||||||||||||||
| Bf Type: 100 µA | BF Type: 500 µA | ||||||||||||||||||||