| Literature DB >> 29440951 |
Cristian Tudor Buescu1, Cristian Mircea Dudescu2, Sergiu Caterev3, Dan Gheorghe Osvald Lucaciu1.
Abstract
BACKGROUND AND AIMS: Using free quadriceps tendon graft for anterior cruciate ligament (ACL) reconstruction has become more popular in the last years due to the low morbidity at the harvest site and biomechanical and structural properties similar to other types of graft (hamstring and patellar tendons). As long as the tibial tunnel fixation with interference screws is considered the standard fixation, femoral fixation is still an open subject. Even though the most common type of femoral tunnel fixation are the cortical suspension devices, the type of suture used for tying the graft to the loop is still evaluated and discussed. We aimed to evaluate the resistance and elongation of a double whip stitch used for tying a tendon to the loop of a cortical suspension device.Entities:
Keywords: anterior cruciate ligament reconstruction; double whip stitch; free quadriceps graft; porcine model
Year: 2018 PMID: 29440951 PMCID: PMC5808266 DOI: 10.15386/cjmed-816
Source DB: PubMed Journal: Clujul Med ISSN: 1222-2119
Figure 1Porcine flexor digitorum profundus with sesamoid bone preserved.
Figure 2Prepared tendon graft with Hi-Fi® suture.
Figure 3Construct fixed in testing device and measurements taken.
Figure 4Maximum load breaking of the suture wires.
Figure 5Elongation curves of samples according to tensile load and maximum load failure.
Maximum load, extension at maximum load and statistical analysis.
| Specimen Label | Maximum Load (N) | Extension at Maximum Load (mm) | |
|---|---|---|---|
| T_1 | 639.38 | 31.74853 | |
| T_2 | 529.34 | 38.33060 | |
| T_3 | 508.54 | 40.39202 | |
| T_4 | 358.93 | 41.18263 | |
| T_5 | 491.83 | 37.29122 | |
| T_6 | 554.70 | 48.60466 | |
| T_7 | 461.51 | 34.01135 | |
| T_8 | 596.61 | 38.30867 | |
| T_9 | 410.66 | 41.09939 | |
| T_10 | 505.37 | 44.50931 | |
| Coefficient of Variation | 16.38981 | 12.27300 | |
| Maximum | 639.38 | 48.60466 | |
| Mean | 505.68 | 39.54784 | |
| Median | 506.95 | 39.36131 | |
| Minimum | 358.93 | 31.74853 | |
| Range | 280.45 | 16.85613 | |
| Standard Deviation | 82.88078 | 4.85371 | |
| Mean + 1SD | 588.57 | 44.40154 | |
| Mean − 1SD | 422.80 | 34.69413 |
Initial and final lengths of tendon and total construct and elongation at maximum load.
| No. | Initial tend length (mm) | Final tend length (mm) | Initial total construct length (mm) | Final total construct length (mm) | Tendon elongation (mm) | Total construct elongation (mm) |
|---|---|---|---|---|---|---|
| 1 | 125 | 130 | 160 | 191 | 5 | 31 |
| 2 | 135 | 140 | 170 | 208 | 5 | 38 |
| 3 | 145 | 150 | 180 | 220 | 5 | 40 |
| 4 | 110 | 115 | 145 | 186 | 5 | 41 |
| 5 | 125 | 130 | 160 | 197 | 5 | 37 |
| 6 | 125 | 130 | 160 | 208 | 5 | 48 |
| 7 | 135 | 140 | 170 | 204 | 5 | 34 |
| 8 | 145 | 150 | 180 | 218 | 5 | 38 |
| 9 | 140 | 145 | 175 | 216 | 5 | 41 |
| 10 | 135 | 140 | 170 | 214 | 5 | 44 |