Literature DB >> 29436001

Effects of Pulsatile Blood Flow on Oxygenator Performance.

Lotte Schraven1, Andreas Kaesler1, Christian Flege2, Rüdger Kopp2, Thomas Schmitz-Rode1, Ulrich Steinseifer1, Jutta Arens1.   

Abstract

Extracorporeal membrane oxygenation (ECMO) is mainly used for the therapy of acute respiratory distress syndrome and chronic obstructive lung disease. In the last years, the development of these systems underwent huge steps in optimization, but there are still problems with thrombus formation, clogging, and thus insufficient gas exchange. One idea of ECMO optimization is a pulsatile blood flow through the oxygenator, but this is still a controversy discussion. Analyzing available publications, it was not possible to identify a general statement about the effect of pulsatile blood flow on the gas exchange performance. The variety of parameters and circuit components have such a high influence on the outcome that a direct comparison of the studies is difficult. For this reason, we performed a structured study to evaluate the effects of pulsatile blood flow on the gas exchange performance of oxygenator. In in vitro tests according to DIN EN ISO 7199, we tested a small oxygenator (0.25 m2 exchange surface, polymethylpentene fibers, 33 mL priming volume) with constant and pulsatile blood flow in comparison. Therefore, we varied the mean blood flow from 250 to 1200 mL/min, the amplitude of 0, 20, and 50%, and the frequency of 30, 60, and 90 bpm. The results demonstrate that the gas transfer for pulsatile and constant blood flow was similar (oxygen: 36-64 mLO2 /LBlood ; carbon dioxide 35-80 mLCO2 /LBlood ) for the same mean blood flow ranges. Over all, the results and analyses showed a statistically nonsignificant difference between pulsatile and nonpulsatile flow. Consequently, we conclude that the implementation of pulsatile blood flow has only a small to no effect on the gas exchange performance in an oxygenator. As the results were obtained using an oxygenator with a coiled fiber bundle, the test must be verified for a stacked fiber oxygenator.
© 2018 International Center for Artificial Organs and Transplantation and Wiley Periodicals, Inc.

Entities:  

Keywords:  -Extracorporeal lung assist; -Gas exchange; -Oxygenator; Pulsatile blood flow

Mesh:

Year:  2018        PMID: 29436001     DOI: 10.1111/aor.13088

Source DB:  PubMed          Journal:  Artif Organs        ISSN: 0160-564X            Impact factor:   3.094


  5 in total

1.  Early in vivo experience with the pediatric continuous-flow total artificial heart.

Authors:  Jamshid H Karimov; David J Horvath; Nicole Byram; Gengo Sunagawa; Barry D Kuban; Shengqiang Gao; Raymond Dessoffy; Kiyotaka Fukamachi
Journal:  J Heart Lung Transplant       Date:  2018-03-30       Impact factor: 10.247

2.  Low-Resistance, Concentric-Gated Pediatric Artificial Lung for End-Stage Lung Failure.

Authors:  Alex J Thompson; Skylar Buchan; Benjamin Carr; Clinton Poling; McKenzie Hayes; Uditha Piyumindri Fernando; Andreas Kaesler; Peter Schlanstein; Felix Hesselmann; Jutta Arens; Joseph A Potkay; Alvaro Rojas-PeÑa; Robert H Bartlett; Ronald B Hirschl
Journal:  ASAIO J       Date:  2020-04       Impact factor: 3.826

3.  Effects on the pulmonary hemodynamics and gas exchange with a speed modulated right ventricular assist rotary blood pump: a numerical study.

Authors:  Feng Huang; Zhe Gou; Yang Fu; Xiaodong Ruan
Journal:  Biomed Eng Online       Date:  2018-10-20       Impact factor: 2.819

4.  Suitable CO2 Solubility Models for Determination of the CO2 Removal Performance of Oxygenators.

Authors:  Benjamin Lukitsch; Paul Ecker; Martin Elenkov; Christoph Janeczek; Christian Jordan; Claus G Krenn; Roman Ullrich; Margit Gfoehler; Michael Harasek
Journal:  Bioengineering (Basel)       Date:  2021-03-02

5.  Extracorporeal Hyperoxygenation Therapy (EHT) for Carbon Monoxide Poisoning: In-Vitro Proof of Principle.

Authors:  Niklas B Steuer; Peter C Schlanstein; Anke Hannig; Stephan Sibirtsev; Andreas Jupke; Thomas Schmitz-Rode; Rüdger Kopp; Ulrich Steinseifer; Georg Wagner; Jutta Arens
Journal:  Membranes (Basel)       Date:  2021-12-31
  5 in total