Literature DB >> 29432140

Noise Exposure From High-Flow Nasal Cannula Oxygen Therapy: A Bench Study on Noise Reduction.

Takamitsu Kubo1, Hiroaki Nakajima2, Ryo Shimoda2, Tatsuya Seo2, Yurie Kanno2, Toshikazu Kondo2, Sunao Tamai3.   

Abstract

BACKGROUND: High-flow nasal cannula (HFNC) oxygen therapy produces noise at a level such that patients often complain. However, the noise level has not been measured digitally.
METHODS: We evaluated 3 types of HFNCs without filters and 2 types with filters attached for noise reduction. Optiflow (with and without a filter), MaxVenturi (with and without a filter) and AIRVO2 (without a filter only) were positioned at the center of a hospital room. We measured the noise levels at the distance of 1 m from the equipment at various total flows (30, 40, 50, 60 L/min) and FIO2 (0.40, 0.60, and 0.90).
RESULTS: Noise levels were increased with the AIRVO2 and MaxVenturi when total flow and FIO2 were increased. Noise levels decreased with the MaxVenturi when a filter was used. The noise level did not change with the Optiflow when total flow and FIO2 were increased. The noise level decreased in the groups with AIRVO2 and Optiflow compared with MaxVenturi without a filter.
CONCLUSIONS: The findings in this study show that the noise level of HFNC/Venturi could be reduced by attaching an intake filter. However, the noise level of HFNC/blender and HFNC/turbine decreased in comparison with HFNC/Venturi without an intake filter.
Copyright © 2018 by Daedalus Enterprises.

Entities:  

Keywords:  airvo2; hfnc; high flow nasal cannula; maxventuri; noise; noise level; optiflow

Mesh:

Substances:

Year:  2018        PMID: 29432140     DOI: 10.4187/respcare.05668

Source DB:  PubMed          Journal:  Respir Care        ISSN: 0020-1324            Impact factor:   2.258


  2 in total

Review 1.  Conventional versus high-flow oxygen therapy in dogs with lower airway injury.

Authors:  Meera Ramesh; Elizabeth Thomovsky; Paula Johnson
Journal:  Can J Vet Res       Date:  2021-10       Impact factor: 1.310

2.  The comfort assessment in healthy adults during constant-flow mode in noninvasive ventilator.

Authors:  Juanjuan Yao; Wei Li; Mengmeng Peng; Kaixin He; Dedong Ma; Hongxiu Lu
Journal:  Clin Respir J       Date:  2021-12-06       Impact factor: 1.761

  2 in total