Literature DB >> 29427296

Volumetric wireless coil based on periodically coupled split-loop resonators for clinical wrist imaging.

Alena V Shchelokova1, Cornelis A T van den Berg2, Dmitry A Dobrykh1, Stanislav B Glybovski1, Mikhail A Zubkov1, Ekaterina A Brui1, Dmitry S Dmitriev3, Alexander V Kozachenko1, Alexander Y Efimtcev1,4, Andrey V Sokolov1,4, Vladimir A Fokin1,4, Irina V Melchakova1, Pavel A Belov1.   

Abstract

PURPOSE: Design and characterization of a new inductively driven wireless coil (WLC) for wrist imaging at 1.5 T with high homogeneity operating due to focusing the B1 field of a birdcage body coil.
METHODS: The WLC design has been proposed based on a volumetric self-resonant periodic structure of inductively coupled split-loop resonators with structural capacitance. The WLC was optimized and studied regarding radiofrequency fields and interaction to the birdcage coil (BC) by electromagnetic simulations. The manufactured WLC was characterized by on-bench measurements and in vivo and phantom study in comparison to a standard cable-connected receive-only coil.
RESULTS: The WLC placed into BC gave the measured B1+ increase of the latter by 8.6 times for the same accepted power. The phantom and in vivo wrist imaging showed that the BC in receiving with the WLC inside reached equal or higher signal-to-noise ratio than the conventional clinical setup comprising the transmit-only BC and a commercial receive-only flex-coil and created no artifacts. Simulations and on-bench measurements proved safety in terms of specific absorption rate and reflected transmit power.
CONCLUSIONS: The results showed that the proposed WLC could be an alternative to standard cable-connected receive coils in clinical magnetic resonance imaging. As an example, with no cable connection, the WLC allowed wrist imaging on a 1.5 T clinical machine using a full-body BC for transmitting and receive with the desired signal-to-noise ratio, image quality, and safety.
© 2018 International Society for Magnetic Resonance in Medicine.

Keywords:  SNR enhancement; metasolenoid; split-loop resonators; transmit efficiency; wireless coil

Mesh:

Year:  2018        PMID: 29427296     DOI: 10.1002/mrm.27140

Source DB:  PubMed          Journal:  Magn Reson Med        ISSN: 0740-3194            Impact factor:   4.668


  2 in total

1.  Deep learning-based fully automatic segmentation of wrist cartilage in MR images.

Authors:  Ekaterina Brui; Aleksandr Y Efimtcev; Vladimir A Fokin; Remi Fernandez; Anatoliy G Levchuk; Augustin C Ogier; Alexey A Samsonov; Jean P Mattei; Irina V Melchakova; David Bendahan; Anna Andreychenko
Journal:  NMR Biomed       Date:  2020-05-11       Impact factor: 4.044

2.  Wireless coils based on resonant and nonresonant coupled-wire structure for small animal multinuclear imaging.

Authors:  Tania S Vergara Gomez; Marc Dubois; Stanislav Glybovski; Benoit Larrat; Julien de Rosny; Carsten Rockstuhl; Monique Bernard; Redha Abdeddaim; Stefan Enoch; Frank Kober
Journal:  NMR Biomed       Date:  2019-02-17       Impact factor: 4.044

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.