MCNP5 evaluation of photoneutron production from the Alexandria University 15 MV Elekta Precise medical LINAC.

W M Abou-Taleb1, M H Hassan2, E A El Mallah3, S M Kotb4.   

Abstract

Photoneutron production, and the dose equivalent, in the head assembly of the 15 MV Elekta Precise medical linac; operating in the faculty of Medicine at Alexandria University were estimated with the MCNP5 code. Photoneutron spectra were calculated in air and inside a water phantom to different depths as a function of the radiation field sizes. The maximum neutron fluence is 3.346×10-9 n/cm2-e for a 30×30 cm2 field size to 2-4 cm-depth in the phantom. The dose equivalent due to fast neutron increases as the field size increases, being a maximum of 0.912 ± 0.05 mSv/Gy at depth between 2 and 4 cm in the water phantom for 40×40 cm2 field size. Photoneutron fluence and dose equivalent are larger to 100 cm from the isocenter than to 35 cm from the treatment room wall.
Copyright © 2018 Elsevier Ltd. All rights reserved.

Entities:  

Keywords:  Dose equivalent; Elekta Linac; MCNP5; Photoneutrons

Year:  2018        PMID: 29413836     DOI: 10.1016/j.apradiso.2018.01.036

Source DB:  PubMed          Journal:  Appl Radiat Isot        ISSN: 0969-8043            Impact factor:   1.513