Zheng He Wang1, Zhong Ping Yang1, Xi Jie Wang1, Yan Hui Dong1, Jun Ma1.
Abstract
To examine accuracy of body composition predicted by the Multi-Frequency Bioelectric Impedance Analysis (MF-BIA) compared with the Dual-energy X-ray Absorptiometry (DXA) in adults with obese. We measured body composition of 749 adults with obese both by the MF-BIA and DXA. The Lin's concordance correlation and the Bland-Altman plots were used to examine the consistency. The concordance correlation coefficient of %BF between the MF-BIA and DXA in men and women was 0.560, and 0.669, respectively. Compared with the DXA, the MF-BIA significantly underestimated %BF by 4.33% in men (P < 0.001), however overestimated %BF by 0.50% in women (P < 0.001). After corrected by the correction equations established in this study, the differences were significantly decreased. Therefore, the MF-BIA (TANITA MC-180) may need to be corrected in estimating body composition for adults with obese.Entities:
Mesh:
Year: 2018 PMID: 29409587 DOI: 10.3967/bes2018.008
Source DB: PubMed Journal: Biomed Environ Sci ISSN: 0895-3988 Impact factor: 3.118