Biomechanical evaluation of pyrocarbon proximal interphalangeal joint arthroplasty: An in-vitro analysis.

A Completo1, A Nascimento2, A F Girão3, F Fonseca2.   

Abstract

BACKGROUND: Pyrocarbon proximal interphalangeal joint arthroplasty provided patients with excellent pain relief and joint motion, however, overall implant complications have been very variable, with some good outcomes at short-medium-term follow-up and some bad outcomes at longer-term follow-up. Implant loosening with migration, dislocation and implant fracture were the main reported clinical complications. The aim of the present work was to test the hypothesis that the magnitude proximal interphalangeal joint cyclic loads in daily hand functions generates stress-strain behaviour which may be associated with a risk of pyrocarbon component loosening in the long-term.
METHODS: This study was performed using synthetic proximal and middle phalanges to experimentally predict the cortex strain behaviour and implant stability considering different load conditions for both intact and implanted states. Finite element models were developed to assess the structural behaviour of cancellous-bone and pyrocarbon components, these models were validated against experimentally measured cortex strains.
FINDINGS: Cortex strains showed a significant increase at dorsal side and reduction at palmar side between intact and implanted states. Cancellous-bone adjacent to the condylar implant base components suffers a two to threefold strain increase, comparing with the intact condition.
INTERPRETATION: The use of pyrocarbon implant changes the biomechanical behaviour of the joint phalanges and is associated with a potential risk of support cancellous-bone suffer fatigue failure in mid to long term due to the strain increase for cyclic loads in the range of daily hand activities, this risk is more prominent than the risk of bone resorption due to strain-shielding effect.
Copyright © 2018 Elsevier Ltd. All rights reserved.

Entities:  

Keywords:  Experimental strains; Finite element model; Proximal interphalangeal joint; Stress-shielding

Mesh:

Substances:

Year:  2018        PMID: 29407860     DOI: 10.1016/j.clinbiomech.2018.01.005

Source DB:  PubMed          Journal:  Clin Biomech (Bristol, Avon)        ISSN: 0268-0033            Impact factor:   2.063


  1 in total

1.  Development and Biomechanical Evaluation of an Anatomical 3D Printing Modularized Proximal Inter-Phalangeal Joint Implant Based on the Computed Tomography Image Reconstructions.

Authors:  Yi-Chao Hunag; Chun-Ming Chang; Shao-Fu Huang; Chia-Heng Hong; Chun-Li Lin
Journal:  Int J Bioprint       Date:  2022-06-10
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.